High-speed nuclear data acquisition and processing system

Closed Pre-Solicitation Posted

A newer solicitation was posted. See the latest solicitation from .

Solicitation number
FA8601-AFITENP-NuclearData
Agency
AFLCMC Wright Patterson AFB Lifecycle Management Center, Department of Defense
Responses due
Set-aside
No set-aside

Opportunity facts

NAICS code
334516 Analytical Laboratory Instrument Manufacturing
PSC
6625 Electrical And Electronic Properties Measuring And Testing Instruments
Place of performance
Wright Patterson AFB, Ohio 45433, United States
Points of contact

Notice details come from SAM.gov. Updated .

About this opportunity

Paragraph 1:

The Department of the Air Force Materiel Command Lifecycle Management Center at Wright Patterson Air Force Base is seeking to procure a High-Speed Nuclear Data Acquisition and Processing System. The key requirements include simultaneous sampling at 500 MS/s on 40 channels with 14-bit ADC resolution, trigger stamp resolution better than 100 ps, memory buffer of 5 MS/channel, data transfer rate of 80 MB/s to 50 TB of data storage, and online digital pulse processing capabilities. The system must interface with AFIT's recently procured dual, double-sided strip detector. The period of performance is 12 weeks after contract award, with delivery to WPAFB.

Paragraph 2:

This is a pre-solicitation notice and does not indicate any set-aside designations. There is no information provided about potential incumbents or current contractors. The budget or award value range is not specified. The key numerical details include the requirement for 40 channels of simultaneous data acquisition, processing, and transfer to data storage.

Notice text

High-Speed Nuclear Data Acquisition and Processing System

Document Date:  20 Mar 2024

1. Scope: The measurement of short-lived nuclear events requires instrumentation with fast acquisition rates. Further, as multi-detector systems are often employed in these experiments to improve counting efficiency and/or measure additional event characteristics, there is a need for simultaneous measurement across several data acquisition channels. Real-time monitoring of experiment status is facilitated by the associated pulse processing software.

1.1 Background:  AFIT’s current nuclear digital data acquisition capability is not at the state-of-the-art in digitization rate, computational speed, or storage capacity. Whereas the digitization frequency of modern data acquisition equipment is about 100 GHz, AFIT’s current capability is 5 GHz at best, and usually slower. As a result, nuclear events of 100 picoseconds or less cannot be measured by students and faculty in our Nuclear Engineering program.

An appropriate High-Speed Data Acquisition and Processing System for Nuclear Measurements includes several subsystems: Digital Pulse Acquisition, Digital Pulse Processing and Transduction, Fast Optical Download, and Data Storage. These integrated hardware components are typically accompanied by control modules for instrument set-up and data acquisition, and for post-collection data analysis.

In one intended research application, the data acquisition system will interface with the recently procured dual, double-sided strip detector (dDSSD), which outputs 40 channels of data (from two 10x10 imaging arrays). The limiting factor in image acquisition rate is the speed of the multichannel acquisition hardware and the transfer of data from the acquisition buffer into the data storage system. The currently available (analog) acquisition hardware severely limits data collection and, consequently, the range of applications in which the dDSSD would be considered an acceptable imager choice. Here, the data download capability is also key and fast throughput would allow efficient generation of rich datasets amenable to Artificial Intelligence interrogation s. In another application, the fast data acquisition system would allow coincidence detection of multiple nuclear events with the necessary timing resolution to investigate material properties with much greater accuracy.

2. Applicable Documents

N/A

3. Requirements/Salient Characteristics:

  • Ability to sample waveforms from charge-sensitive preamp of radiation detector, with unipoloar Vpeak ≤ 2 V
  • Simultaneous sampling at ≥ 500 MS/s on ≥ 40 channels, with all channels time-synched
  • ≥ 14-bit analog-to-digital conversion (ADC) resolution
  • Internal ADC sampling clock with ≥ 50 MHz timing resolution; capability to utilize an external clock signal
  • Individual self-trigger with channel over/under threshold, and trigger stamp resolution of better than 100 ps
  • Memory buffer ≥ 5 MS/channel
  • Communication interface that enables data in all active channels to simultaneously transfer data to data storage at a rate ≥ 80 MB/s, with included hardware for data storage (≥ 50 TB)
  • Online digital pulse processing with ability to customize gating, pulse shaping parameters and pulse discrimination (over/under threshold, coincidence, veto, etc.)
  • Compatible DAQ software that allows independent control of each active channel
  • All necessary cables and interfaces to implement ≥ 40 channels of simultaneous data acquisition, data processing, and transfer to data storage

4. Delivery/Period of Performance: Equipment will be shipped to WPAFB.

PoP 12 week after contract award.

4.1 Place of Performance:

Air Force Institute of Technology (AFIT/EN)

2950 Hobson Way

Wright-Patterson AFB, OH 45433-7765

Attachments

Files attached to this notice, newest first
File Type Posted
08 Statement of Work Fast Nuclear DAQ System 230821.docx DOCX document

Notice history

Notices posted for this opportunity, newest first
Notice Type Posted
High-speed nuclear data acquisition and processing system Latest solicitation Solicitation
High-speed nuclear data acquisition and processing system This notice · Original Pre-Solicitation

On GovTribe

Work this opportunity on GovTribe

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity