Coordinate Measuring Machine
Closed Solicitation Posted
- Solicitation number
- FA812521Q0029
- Agency
- Air Force Sustainment Center Air Force Materiel Command, Department of Defense
- Responses due
- Set-aside
- Total Small Business
Opportunity facts
- NAICS code
- 334513 Instruments and Related Products Manufacturing for Measuring, Displaying, and Controlling Industrial Process Variables
- Place of performance
- Tinker Afb Ok 73145 USA
Notice details come from SAM.gov. Updated .
Notice text
4 versions
Update #4 · Latest ·
FA812521Q0029-01
(1) Coordinate Measuring Machine
The requested descriptions/specifications are outlined below.
- A bridge style CMM, able to measure variety of parts to an accuracy of 1.5 + L/333 microns or better. Length measuring uncertainties guaranteed from 18 - 22°C.
- Minimum table measurement range of x=500mm, y=700mm, z=500mm.
- Reference/Probe Qualification sphere.
- Automatic Stylus Changer.
- Stylus Kit with multiple probes from 3mm diameter to 8 mm diameter.
- Option to add optical sensors.
- 2.5 degree articulating head
- Scanning Probe
- Option to add optical sensors.
- Ability to program directly from CAD models
- Ability to create parametric programs
- Software must be compatible with Windows 10 Operating System
- Jog Box with single joystick or an operator panel with dual joysticks
- Air Dryer/Oil removal that supports a minimum of 10 CFM.
- Total footprint allocated for the installation of the CMM shall not exceed 72” by 72”.
- Room ceiling height is 102 inches. The ceiling is a drop ceiling, but the Government does not wish to raise the ceiling to accommodate a new CMM. However, if required, the Government will raise the ceiling tile directly over the final placement of the CMM. The ceiling tile directly over the CMM can be raised a height of 18” for a total ceiling height of 120”. The access doorway for the room is 71 inches wide by 83 inches tall.
- System shall operate on 120 VAC, 1 phase, 60 hertz with amperage not to exceed 20 amps.
- Low pressure air at 150 psig not to exceed 20 cfm.
- The contractor shall deliver and perform on-site installation of the CMM at 3538 C. Avenue, Tinker AFB, OK 73145 in Building 200, Room 118. The contractor will supply all transportation, rigging, forklift/crane, and labor necessary for offloading, transport and installation at permanent location.
- Operator Training for a minimum of 2 attendees (may be onsite, online or at contractor’s facility or combination thereof).
- Basic training - basic understanding of software and measurement capabilities of CMM and operating CMM for first time users.
- Advanced training - cover advance concepts beyond the basic training, such as importing CAD files for parts measurement, learning scanning concepts for analog scanning probes.
- Contractor must meet all requirements outlined in the Statement of Work (SOW).
Update #3 ·
See attachment:
Coordinat Measuring Machine Updates_Timelines_ Deadlines
Update #2 ·
Please Note:
A site visit will be held April 14, 2021 (Due to COVID-19 restrictions, this visit will be limited to 10 people).
Government Answers to Industry Questions are attached.
The RFQ is extended to April 28, 2021.
Requirment remains the same and unchanged:
(1) Coordinate Measuring Machine
The requested software descriptions/specifications are outlined below.
- A bridge style CMM, able to measure variety of parts to an accuracy of 1.5 + L/333 microns or better. Length measuring uncertainties guaranteed from 18 - 22°C.
- Minimum table measurement range of x=500mm, y=700mm, z=500mm.
- Reference/Probe Qualification sphere.
- Automatic Stylus Changer.
- Stylus Kit with multiple probes from 3mm diameter to 8 mm diameter.
- Option to add optical sensors.
- Air Dryer/Oil removal that supports a minimum of 10 CFM.
- Total footprint allocated for the installation of the CMM shall not exceed 72” by 72”.
- Room ceiling height is 102 inches. The ceiling is a drop ceiling, but the Government does not wish to raise the ceiling to accommodate a new CMM. However, if required, the Government will raise the ceiling tile directly over the final placement of the CMM. The ceiling tile directly over the CMM can be raised a height of 18” for a total ceiling height of 120”. The access doorway for the room is 71 inches wide by 83 inches tall.
- System shall operate on 120 VAC, 1 phase, 60 hertz with amperage not to exceed 20 amps.
- Low pressure air at 150 psig not to exceed 20 cfm.
- The contractor shall deliver and perform on-site installation of the CMM at 3538 C. Avenue, Tinker AFB, OK 73145 in Building 200, Room 118. The contractor will supply all transportation, rigging, forklift/crane, and labor necessary for offloading, transport and installation at permanent location.
- Operator Training for a minimum of 2 attendees (may be onsite, online or at contractor’s facility or combination thereof).
- Basic training - basic understanding of software and measurement capabilities of CMM and operating CMM for first time users.
- Advanced training - cover advance concepts beyond the basic training, such as importing CAD files for parts measurement, learning scanning concepts for analog scanning probes.
- Contractor must meet all requirements outlined in the Statement of Work (SOW).
Update #1 ·
(1) Coordinate Measuring Machine
The requested software descriptions/specifications are outlined below.
- A bridge style CMM, able to measure variety of parts to an accuracy of 1.5 + L/333 microns or better. Length measuring uncertainties guaranteed from 18 - 22°C.
- Minimum table measurement range of x=500mm, y=700mm, z=500mm.
- Reference/Probe Qualification sphere.
- Automatic Stylus Changer.
- Stylus Kit with multiple probes from 3mm diameter to 8 mm diameter.
- Option to add optical sensors.
- Air Dryer/Oil removal that supports a minimum of 10 CFM.
- Total footprint allocated for the installation of the CMM shall not exceed 72” by 72”.
- Room ceiling height is 102 inches. The ceiling is a drop ceiling, but the Government does not wish to raise the ceiling to accommodate a new CMM. However, if required, the Government will raise the ceiling tile directly over the final placement of the CMM. The ceiling tile directly over the CMM can be raised a height of 18” for a total ceiling height of 120”. The access doorway for the room is 71 inches wide by 83 inches tall.
- System shall operate on 120 VAC, 1 phase, 60 hertz with amperage not to exceed 20 amps.
- Low pressure air at 150 psig not to exceed 20 cfm.
- The contractor shall deliver and perform on-site installation of the CMM at 3538 C. Avenue, Tinker AFB, OK 73145 in Building 200, Room 118. The contractor will supply all transportation, rigging, forklift/crane, and labor necessary for offloading, transport and installation at permanent location.
- Operator Training for a minimum of 2 attendees (may be onsite, online or at contractor’s facility or combination thereof).
- Basic training - basic understanding of software and measurement capabilities of CMM and operating CMM for first time users.
- Advanced training - cover advance concepts beyond the basic training, such as importing CAD files for parts measurement, learning scanning concepts for analog scanning probes.
- Contractor must meet all requirements outlined in the Statement of Work (SOW).
Attachments
Show all 12
Notice history
| Notice | Type | Posted |
|---|---|---|
| Coordinate Measuring Machine | Solicitation | |
| Coordinate Measuring Machines | Pre-Solicitation |
On GovTribe
Work this opportunity on GovTribe
- Track it in your pipeline
- Find teaming partners
- Similar opportunities
- Ask GovTribe AI about this opportunity