Attach (5) Coordinate Measuring Machine Technical Proposal Evaluation.docx
DOCX document 26 KB Posted
- Attached to
- Coordinate Measuring Machine Federal contract opportunity
- Solicitation number
- FA812521Q0029
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Attach (4) Coordinate Measuring Machine Item Description Amendment 01.docx | DOCX document | |
| Coordinate Measuring Machine_Combined Synopsis Solicitation (CSS) FA812521Q0029_Amend 01.doc | DOC document | |
| Attach (2) Coordinate Measuring Machine Statment of Work (SOW) Amendment 01.pdf | ||
| Attach (3) CMM Questions and Answers FA812521Q0029 01.docx | DOCX document | |
| Attach (1) Coordinate Measuring Machine Update dated April 28 2021.docx | DOCX document | |
| Coordinate Measuring Machine Updates_Timelines_Deadlines.docx | DOCX document | |
| Notification of Site Visit 14 April 21 and RFQ Extension 28 April 21.docx | DOCX document | |
| Questions and Answers_ Coordinate Measuring Machines FA812521Q0029.docx | DOCX document | |
| Coordinate Measuring Machine SOW.docx | DOCX document | |
| Coordinate Measuring Machine Item Description.docx | DOCX document | |
| Coordinate Measuring Machine Solicitation No. FA812521Q0029.pdf |
Show all 11
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Technical Proposal Evaluation – Critical Factor List Project: _C21-006 Coordinate Measuring Machine ____________ Proposal Number: ______________________________ Contractor: ____________________________________ Evaluator: __Blake Grimwood ______________ _______ Date: _________________________________
| Item |
| Acceptable |
| Non-Acceptable |
| Remarks |
A bridge style CMM, able to measure variety of parts to an accuracy of 1.5 + L/333 microns or better. Length measuring uncertainties guaranteed from 18 - 22°C.
Minimum table measurement range of x=500mm, y=700mm, z=500mm.
Reference/Probe Qualification sphere.
Automatic Stylus Changer.
Stylus Kit with multiple probes from 3mm diameter to 8 mm diameter.
2.5 degree articulating head
Scanning probe
Option to add optical sensors.
Ability to program directly from CAD models
Ability to create parametric programs
Software must be compatible with Windows 10 Operating System
Jog Box with single joystick or an operator panel with dual joysticks
Air Dryer/Oil removal that supports a minimum of 10 CFM.
Uninterruptable Power Supply & Power Conditioner
Total footprint allocated for the installation of the CMM shall not exceed 72” by 72”.
Room ceiling height is 102 inches. The ceiling is a drop ceiling, but the Government does not wish to raise the ceiling to accommodate a new CMM. However, if required, the Government will raise the ceiling tile directly over the final placement of the CMM. The ceiling tile directly over the CMM can be raised a height of 18” for a total ceiling height of 120”. The access doorway for the room is 71 inches wide by 83 inches tall.
System shall operate on 120 VAC, 1 phase, 60 hertz with amperage not to exceed 20 amps.
Low pressure air at 150 psig not to exceed 20 cfm.
The contractor shall deliver and perform on-site installation of the CMM at 3538 C. Avenue, Tinker AFB, OK 73145 in Building 200, Room 118. The contractor will supply all transportation, rigging, forklift/crane, and labor necessary for offloading, transport and installation at permanent location.
Operator Training for a minimum of 2 attendees (may be onsite, online or at contractor’s facility or combination thereof).
Basic training - basic understanding of software and measurement capabilities of CMM and operating CMM for first time users.
Advanced training - cover advance concepts beyond the basic training, such as importing CAD files for parts measurement, learning scanning concepts for analog scanning probes.
Contractor must meet all requirements outlined in the Statement of Work (SOW).
Technical Evaluator Signature Print Name Date
File details come from the government source that posted it. Updated .