B533--VANIHCS SPS Water Testing (VA-27-00000760)

Closed Pre-Solicitation Posted

Solicitation number
36C25026R0060
Agency
Responses due
Set-aside
Total Small Business

Opportunity facts

Place of performance
VA Northern Indiana Health Care System Multiple location - Statement of work for location Fort Wayne, Indiana 46805, United States

Notice details come from SAM.gov. Updated .

About this opportunity

The Department of Veterans Affairs Northern Indiana Health Care System (NIHCS) is conducting market research through this Sources Sought Notice to identify qualified vendors capable of providing water testing and reagents for sterile processing systems (SPS) at three VA facilities in Marion, Fort Wayne, and Jackie Walorski, Indiana. This pre-solicitation announcement is for information-gathering purposes only and will not result in contract awards. Qualified vendors must demonstrate at least five years of experience providing water treatment testing at medical centers, employ a water treatment specialist holding a bachelor's degree in chemical engineering or physical sciences with five years of relevant experience, and operate or contract with a state of Indiana-certified laboratory supervised by a Professional Chemist. The procurement requires vendors to provide monthly testing visits using calibrated conductivity and resistivity meters, deliver comprehensive water analysis reports, and furnish training to VA staff, with all testing performed in compliance with ANSI/AAMI ST108:2023 standards. Responses must be submitted via email to eric.sweatt@va.gov no later than April 22, 2026 at 3:00 PM EST and must not exceed 10 pages or 5 MB in file size.

This opportunity is set aside for total small business, with specific targeting of Service-Disabled Veteran Owned Small Businesses (SDVOSB), Veteran Owned Small Businesses (VOSB), and general Small Business (SB) concerns, all of which must be registered in the System for Award Management (SAM) with current representations and certifications. The appropriate NAICS code is 541380 (Testing Laboratories) with a size standard of $16.5 million. While no specific contract value is disclosed in this pre-solicitation notice, respondents are instructed to include Federal Supply Schedule contract information if applicable or indicate "open market" status, provide details on three past similar contracts, submit references and pricing lists, and disclose any planned subcontracting arrangements. The market research phase will inform the government's approach to a subsequent formal solicitation for this water testing and reagents requirement.

Notice text

2 versions

Update #2 · Latest ·

THIS IS A SOURCES SOUGHT ANNOUNCEMENT FOR INFORMATION ONLY. THIS IS NOT A SOLICITATION FOR PROPOSALS AND NO CONTRACT WILL BE AWARDED FROM THIS ANNOUNCEMENT. The purpose of this notice is to conduct market research and obtain information from qualified businesses with special interest in seeking: VA certified Service-Disabled Veteran owned Small Business (SDVOSB), VA certified Veteran Owned Small Business (VOSB) and Small Business (SB) concerns. No solicitation is being issued at this time. All information submitted in response to this announcement is voluntary; the Government will not pay for information requested nor will it compensate any respondent for any cost incurred in developing information provided to the Government. The Department of Veteran Affairs, Northern Indiana Health Care System (NIHCS) has a requirement for Water Testing and Reagents for the sterile processing system (SPS) at the Marion, Fort Wayne, and Jackie Walorski VA sites. The appropriate NAICS Code 541380, Testing Laboratories, Size Standard $16.5 Million Responses: Please submit your response in accordance with the following instructions: Responses shall be no more than 10 pages in length. Responses are due by 4/22/2026 at 15:00 PM EST. Responses shall be submitted via email to eric.sweatt@va.gov. Emails cannot exceed 5mb  Subject line of the Email shall include this Sources Sought Notice Number Firm Name NIHCS Water Testing and Reagents for the SPS Services. If your response contains proprietary information you will place Proprietary Information in front of the Sources Sought Notice Number. No marketing materials are allowed as part of this Sources Sought Notice.  The Government will not review any other information or attachments included in your submission that is not listed herein or exceeds the 10-page limitation. Contractors Federal Supply Schedule Contract that carries the desired services If you do not have a Federal Supply Schedule contract, you shall state open market Provide the Schedule Provide expiration date of your Contract Provide your GSA/FSS/SAC Contract Number Provide the SIN for water testing and reagents for SPS services If any of the items are not on your GSA/FSS contract, you will mark which items are open market. Contractor Point of Contact to include: Name Title Phone Email Company Information to include: Name Address SAM UEI Number Tax ID Number Business Size Provide information as to what water testing services are provided by your company. Provide how your company invoices for services in the commercial marketplace. Business Size/Type: Service-Disabled Veteran Owned Small Business (SDVOSB), Veteran Owned Small Business (VOSB), 8(a), Hub Zone Small Business, Women Owned Small Business (WOSB), Small Disadvantaged Business (SDB), Small Business, or Large Business. All prospective Contractors are reminded that in accordance with FAR 4.11 they shall be registered in the System for Award Management (SAM) database and shall complete annual Representations and Certifications in SAM prior to any offer.  SAM is located at http://www.sam.gov and consolidated the capabilities of CCR/FedReg, ORCA, and EPLS.  There is no cost to use SAM. FOR SDVOSB/VOSB ONLY:  In order to be eligible, interested parties must be registered in the System for Award Management (SAM) data base at https://sam.gov. For firms claiming SDVOSB/VOSB classification status, they must also be registered and certified through the Department of Veterans Affairs Office of Small & Disadvantaged Business Utilization data base, located at https://veterans.certify.sba.gov/. Please include a copy of your certification with your response. Contractor shall include 3 past contracts for the same or similar services.  Contractor shall include at minimum: Name of Company Company POC Company POC phone Company POC email Dollar amount of contract Contract number Period of Performance Description of services provided Contractor shall include any relevant comments about the attachments, if applicable. Contractor shall provide a public price list/contract price list for the services described herein. Contractor shall provide a detailed response to include your firm s ability, capability, and responsibility to deliver the services presented in this Sources Sought Notice. Contractor shall provide if they plan to sub-contract all or part of the work, and to what firm to include the subs duns number and business size. Determination by the Government does not compete with this proposed contract action based on responses to this notice is solely within the discretion of the Government. STATEMENT OF WORK SPS Water Testing Services Part I. Description of Services: The contractor shall provide all the following for testing the water for/within SPS per the ANSI/AAMI ST108:2023 at the Marion, Fort Wayne, and Jackie Walorski VA sites as defined in this Statement of Work (SOW). WORK TO BE PERFORMED: Contractors shall provide all labor, materials, equipment, transportation, lodging and supervision necessary to collect appropriate samples and deliver to a lab meeting requirement for processing of AAMI ST108. Contractors shall calibrate the conductivity/resistivity meter monthly during their testing visit and provide a water analysis report. The contractor will be part of the multidisciplinary team that assists the water working group with resolving out of range water testing, in addition to: Serves as the subject matter expert on water testing per the ST108:2023 Serves as the subject matter expert on corrective actions and to mitigate risks for out-of-range water testing parameters. Completes the monthly, quarterly, and yearly testing per the ST108:2023 and per the VHA water contract/SOW. Completes performance qualification testing per the ST108:2023 and per the VHA contract/SOW. Completes follow up testing as ordered by engineering for out-of-range water parameters per the ST108:2023 and VHA water contract/SOW. Contractor shall provide consultative services as needed such as resolving abnormal water testing results, water system analysis, and mitigation of the tests performed Part II Work Requirements Water Treatment Testing Qualifications: The contractor shall have been in business at least five years providing water treatment testing of water distribution systems similar in design at a medical center. Provide a list of at least five references similar to Department of Veterans Affairs who have utilized the services for water treatment in the last three years. One water treatment specialist who has a bachelor s degree in chemical engineering or the physical sciences from an accredited college or university shall be assigned to this contract. The water treatment specialist must have at least five years successful experience with water treatment programs and shall furnish references from at least three customers who utilize water treatment programs. The contractor must operate (or continuously maintain a contract with) a state of Indiana certified laboratory to perform complete scientific analyses of water chemistry, deposits, metallic materials affected by corrosion and provide recommendations for correction of problems related to the chemical and mechanical water treatment program. A Professional Chemist shall supervise the laboratory. On-Site Services: Provide complete written instructions, supplemented by oral instructions and demonstrations, to SPS personnel for their performance of routine water tests. The instructions cover all aspects of safe chemical handling and safe and accurate testing and methods to suit the SPS operating parameters and water treatment performance requirements. Provide one training session within one week of the start of the contract. Provide one training session annually after the new contract has been awarded. The vendor will provide a handheld resistivity meter as well as pH hardness colorimetric dipsticks to the VA. The vendor shall calibrate the resistivity meter on a monthly basis during their testing visit and provide a calibration certification with this visit as part of the report. The vendor shall provide a detailed operations manual and training to the VA staff for procedures related to VA required testing noted below: AAMI ST108 UTILITY WATER, CRITICAL WATER, AND STEAM TESTING BY VENDOR: Fort Wayne Routine (monthly) POU Testing by Vendor Water Quality Measurement Units Type of Testing Routine Sample Site Utility Routine Sample Site Critical Minimum Frequency of Testing and Testing Range Utility Water Critical Water pH pH pH Meter or Colorimetric dipsticks Medivator Basin x4 Decon sinks x3 Washer x1 Ultrasonic x1 Eye Ultrasonic X1 6.5-9.5 Quarterly 5.0-7.5 Monthly Conductivity uS/cm Conductivity Meter Medivator Basin x4 Decon sinks x3 Washer x1 Ultrasonic x1 Eye UltrasonicX1 <500 Quarterly <10 Monthly Total Alkalinity mg CaCO3/L Colorimetric dipsticks or Alkalinity test kit Medivator Basin x4 Decon sinks x3 Washer x1 Ultrasonic x1 Eye Ultrasonic X1 <400 Quarterly <8 Monthly Total Hardness mg CaCO3/L Determination of ppm as CaCO3/L by colorimetric dipsticks, titration kit, or handheld meter Medivator Basin x4 Decon sinks x3 Washer x1 Ultrasonic x1 Eye Ultrasonic X1 <150 Quarterly <1 Monthly Bacteria CFU/mL Heterotrophic Plate Count (HPC) NA Decon sinks x3 Medivator Basin x4 Washer x1 Ultrasonic x1 Eye Ultrasonic X1 <500 Quarterly <10 Monthly Endotoxin EU/mL LAL Test NA Decon sinks x3 Medivator Basin x4 Washer x1 Ultrasonic x1 Eye Ultrasonic X1 NA NA <10 Monthly Fort Wayne Point of Generation Routine (monthly) Testing by Vendor Water Quality Measurement Units Type of Testing Routine Sample Site Critical Minimum Frequency of Testing and Testing Range Critical Water pH pH pH Meter or Colorimetric dipsticks Loop Supply Loop Return 5.0-7.5 Monthly Conductivity uS/cm Conductivity Meter (in line or measurement of a collected sample) Loop Supply Loop Return <10 Monthly Total Alkalinity mg CaCO3/L Colorimetric dipsticks or Alkalinity test kit Loop Supply Loop Return <8 Monthly Total Hardness mg CaCO3/L Determination of ppm as CaCO3/L by colorimetric dipsticks, titration kit, or handheld meter Loop Supply Loop Return <1 Monthly Bacteria CFU/mL Heterotrophic Plate Count (HPC) Loop Supply Loop Return <10 Monthly Endotoxin EU/mL LAL Test Loop Supply Loop Return <10 Monthly Fort Wayne Performance Qualification Testing (quarterly) Testing by Vendor Water Quality Measurement Units Type of Testing PQ Sample Site Utility PQ Sample Site Critical PQ Sample Site Steam Minimum Frequency of Testing and Testing Range Utility Water Critical Water Steam pH pH pH Meter or Colorimetric dipsticks Decon sinks x4 Medivator Basin x4 Washer x2 Ultrasonic x2 Eye Ultrasonic X2 City Water Water Softener Utility DI After Carbon Filter Medivator 1 Supply Medivator 2 Supply Decon sinks x3 Washer x1 Ultrasonic x1 Eye Ultrasonic X1 Loop Supply Loop Return After RO After DI Sterilizer x2 6.5-9.5 Quarterly 5.0-7.5 Monthly 5.0-9.2 Quarterly Conductivity uS/cm Conductivity Meter Decon sinks x4 Medivator Basin x4 Washer x2 Ultrasonic x2 Eye Ultrasonic X2 City Water Water Softener Utility DI After Carbon Filter Medivator 1 Supply Medivator 2 Supply Decon sinks x3 Washer x1 Ultrasonic x1 Eye Ultrasonic X1 Loop Supply Loop Return After RO After DI Sterilizer x2 <500 Quarterly <10 Monthly <10 Quarterly Total Alkalinity mg CaCO3/L Colorimetric dipsticks or Alkalinity test kit Decon sinks x4 Medivator Basin x4 Washer x2 Ultrasonic x2 Eye Ultrasonic X2 City Water Water Softener Utility DI After Carbon Filter Medivator 1 Supply Medivator 2 Supply Decon sinks x3 Washer x1 Ultrasonic x1 Eye Ultrasonic X1 Loop Supply Loop Return After RO After DI Sterilizer x2 <400 Quarterly <8 Monthly <8 Quarterly Total Hardness mg CaCO3/L Determination of ppm as CaCO3/L by colorimetric dipsticks, titration kit, or handheld meter Decon sinks x4 Medivator Basin x4 Washer x2 Ultrasonic x2 Eye Ultrasonic X2 City Water Water Softener Utility DI After Carbon Filter Medivator 1 Supply Medivator 2 Supply Decon sinks x3 Washer x1 Ultrasonic x1 Eye Ultrasonic X1 Loop Supply Loop Return After RO After DI Sterilizer x2 <150 Quarterly <1 Monthly <1 Quarterly Total Organic Carbon (TOC) mg/L (ppm) NA Decon sinks x3 Washer x1 Ultrasonic x1 Eye Ultrasonic X1 Loop Supply Loop Return After RO After DI NA NA NA <1.0 Quarterly NA NA Aluminum mg/L Decon sinks x4 Medivator Basin x4 Washer x2 Ultrasonic x2 Eye Ultrasonic X2 City Water Water Softener Utility DI After Carbon Filter Medivator 1 Supply Medivator 2 Supply Decon sinks x3 Washer x1 Ultrasonic x1 Eye Ultrasonic X1 Loop Supply Loop Return After RO After DI Sterilizer x2 <0.1 Quarterly <0.1 Quarterly <0.1 Quarterly Chloride mg/L Decon sinks x4 Medivator Basin x4 Washer x2 Ultrasonic x2 Eye Ultrasonic X2 City Water Water Softener Utility DI After Carbon Filter Medivator 1 Supply Medivator 2 Supply Decon sinks x3 Washer x1 Ultrasonic x1 Eye Ultrasonic X1 Loop Supply Loop Return After RO After DI Sterilizer x2 <250 Quarterly <1 Quarterly <1 Quarterly Copper mg/L Decon sinks x4 Medivator Basin x4 Washer x2 Ultrasonic x2 Eye Ultrasonic X2 City Water Water Softener Utility DI After Carbon Filter Medivator 1 Supply Medivator 2 Supply Decon sinks x3 Washer x1 Ultrasonic x1 Eye Ultrasonic X1 Loop Supply Loop Return After RO After DI Sterilizer x2 <0.1 Quarterly <0.1 Quarterly <0.1 Quarterly Iron mg/L Decon sinks x4 Medivator Basin x4 Washer x2 Ultrasonic x2 Eye Ultrasonic X2 City Water Water Softener Utility DI After Carbon Filter Medivator 1 Supply Medivator 2 Supply Decon sinks x3 Washer x1 Ultrasonic x1 Eye Ultrasonic X1 Loop Supply Loop Return After RO After DI Sterilizer x2 <0.1 Quarterly <0.1 Quarterly <0.1 Quarterly Manganese mg/L Decon sinks x4 Medivator Basin x4 Washer x2 Ultrasonic x2 Eye Ultrasonic X2 City Water Water Softener Utility DI After Carbon Filter Medivator 1 Supply Medivator 2 Supply Decon sinks x3 Washer x1 Ultrasonic x1 Eye Ultrasonic X1 Loop Supply Loop Return After RO After DI Sterilizer x2 <0.1 Quarterly <0.1 Quarterly <0.1 Quarterly Nitrate mg/L Decon sinks x4 Medivator Basin x4 Washer x2 Ultrasonic x2 Eye Ultrasonic X2 City Water Water Softener Utility DI After Carbon Filter Medivator 1 Supply Medivator 2 Supply Decon sinks x3 Washer x1 Ultrasonic x1 Eye Ultrasonic X1 Loop Supply Loop Return After RO After DI Sterilizer x2 <10 Quarterly <1 Quarterly <1 Quarterly Phosphate mg/L Decon sinks x4 Medivator Basin x4 Washer x2 Ultrasonic x2 Eye Ultrasonic X2 City Water Water Softener Utility DI After Carbon Filter Medivator 1 Supply Medivator 2 Supply Decon sinks x3 Washer x1 Ultrasonic x1 Eye Ultrasonic X1 Loop Supply Loop Return After RO After DI Sterilizer x2 <5 Quarterly <1 Quarterly <1 Quarterly Silicate mg/L Decon sinks x4 Medivator Basin x4 Washer x2 Ultrasonic x2 Eye Ultrasonic X2 City Water Water Softener Utility DI After Carbon Filter Medivator 1 Supply Medivator 2 Supply Decon sinks x3 Washer x1 Ultrasonic x1 Eye Ultrasonic X1 Loop Supply Loop Return After RO After DI Sterilizer x2 <50 Quarterly <1 Quarterly <1 Quarterly Sulfate mg/L Decon sinks x4 Medivator Basin x4 Washer x2 Ultrasonic x2 Eye Ultrasonic X2 City Water Water Softener Utility DI After Carbon Filter Medivator 1 Supply Medivator 2 Supply Decon sinks x3 Washer x1 Ultrasonic x1 Eye Ultrasonic X1 Loop Supply Loop Return After RO After DI Sterilizer x2 <150 Quarterly <1 Quarterly <1 Quarterly Zinc mg/L Decon sinks x4 Medivator Basin x4 Washer x2 Ultrasonic x2 Eye Ultrasonic X2 City Water Water Softener Utility DI After Carbon Filter Medivator 1 Supply Medivator 2 Supply Decon sinks x3 Washer x1 Ultrasonic x1 Eye Ultrasonic X1 Loop Supply Loop Return After RO After DI Sterilizer x2 <0.1 Quarterly <0.1 Quarterly <0.1 Quarterly Bacteria CFU/mL Heterotrophic Plate Count (HPC) Decon sinks x4 Washer x2 Ultrasonic x2 Eye Ultrasonic X2 City Water Water Softener Utility DI After Carbon Filter Medivator 1 Supply Medivator 2 Supply Decon sinks x3 Medivator basin x4 Washer x1 Ultrasonic x1 Eye Ultrasonic X1 Loop Supply Loop Return After RO After DI NA <500 Quarterly <10 Monthly NA NA Endotoxin EU/mL LAL Test NA Decon sinks x3 Medivator basin x4 Washer x1 Ultrasonic x1 Eye Ultrasonic X1 Loop Supply Loop Return After RO After DI NA NA NA <10 Monthly NA NA Marion Routine (monthly) POU Testing by Vendor Water Quality Measurement Units Type of Testing Routine Sample Site Critical Minimum Frequency of Testing and Testing Range Critical Water pH pH pH Meter or Colorimetric dipsticks Decon sinks x1 Washer x1 5.0-7.5 Monthly Conductivity uS/cm Conductivity Meter Decon sinks x1 Washer x1 <10 Monthly Total Alkalinity mg CaCO3/L Colorimetric dipsticks or Alkalinity test kit Decon sinks x1 Washer x1 <8 Monthly Total Hardness mg CaCO3/L Determination of ppm as CaCO3/L by colorimetric dipsticks, titration kit, or handheld meter Decon sinks x1 Washer x1 <1 Monthly Bacteria CFU/mL Heterotrophic Plate Count (HPC) Decon sinks x1 Washer x1 <10 Monthly Endotoxin EU/mL LAL Test Decon sinks x1 Washer x1 <10 Monthly Marion Point of Generation Routine (monthly) Testing by Vendor Water Quality Measurement Units Type of Testing Routine Sample Site Critical Minimum Frequency of Testing and Testing Range Critical Water pH pH pH Meter or Colorimetric dipsticks Supply 5.0-7.5 Monthly Conductivity uS/cm Conductivity Meter (in line or measurement of a collected sample) Supply <10 Monthly Total Alkalinity mg CaCO3/L Colorimetric dipsticks or Alkalinity test kit Supply <8 Monthly Total Hardness mg CaCO3/L Determination of ppm as CaCO3/L by colorimetric dipsticks, titration kit, or handheld meter Supply <1 Monthly Bacteria CFU/mL Heterotrophic Plate Count (HPC) Supply <10 Monthly Endotoxin EU/mL LAL Test Supply <10 Monthly Marion Performance Qualification Testing (quarterly) Testing by Vendor Water Quality Measurement Units Type of Testing PQ Sample Site Utility PQ Sample Site Critical PQ Sample Site Steam Minimum Frequency of Testing and Testing Range Utility Water Critical Water Steam pH pH pH Meter or Colorimetric dipsticks Decon sinks x2 Washer x2 Ultrasonic x2 City Water Water Softener Utility DI After Carbon Filter Decon sinks x1 Washer x1 Supply After RO After DI Sterilizer x2 6.5-9.5 Quarterly 5.0-7.5 Monthly 5.0-9.2 Quarterly Conductivity uS/cm Conductivity Meter Decon sinks x2 Washer x2 Ultrasonic x2 City Water Water Softener Utility DI After Carbon Filter Decon sinks x1 Washer x1 Supply After RO After DI Sterilizer x2 <500 Quarterly <10 Monthly <10 Quarterly Total Alkalinity mg CaCO3/L Colorimetric dipsticks or Alkalinity test kit Decon sinks x2 Washer x2 Ultrasonic x2 City Water Water Softener Utility DI After Carbon Filter Decon sinks x1 Washer x1 Supply After RO After DI Sterilizer x2 <400 Quarterly <8 Monthly <8 Quarterly Total Hardness mg CaCO3/L Determination of ppm as CaCO3/L by colorimetric dipsticks, titration kit, or handheld meter Decon sinks x2 Washer x2 Ultrasonic x2 City Water Water Softener Utility DI After Carbon Filter Decon sinks x1 Washer x1 Supply After RO After DI Sterilizer x2 <150 Quarterly <1 Monthly <1 Quarterly Total Organic Carbon (TOC) mg/L (ppm) NA Decon sinks x1 Washer x1 Supply After RO After DI NA NA NA <1.0 Quarterly NA NA Aluminum mg/L Decon sinks x2 Washer x2 Ultrasonic x2 City Water Water Softener Utility DI After Carbon Filter Decon sinks x1 Washer x1 Supply After RO After DI Sterilizer x2 <0.1 Quarterly <0.1 Quarterly <0.1 Quarterly Chloride mg/L Decon sinks x2 Washer x2 Ultrasonic x2 City Water Water Softener Utility DI After Carbon Filter Decon sinks x1 Washer x1 Supply After RO After DI Sterilizer x2 <250 Quarterly <1 Quarterly <1 Quarterly Copper mg/L Decon sinks x2 Washer x2 Ultrasonic x2 City Water Water Softener Utility DI After Carbon Filter Decon sinks x1 Washer x1 Supply After RO After DI Sterilizer x2 <0.1 Quarterly <0.1 Quarterly <0.1 Quarterly Iron mg/L Decon sinks x2 Washer x2 Ultrasonic x2 City Water Water Softener Utility DI After Carbon Filter Decon sinks x1 Washer x1 Supply After RO After DI Sterilizer x2 <0.1 Quarterly <0.1 Quarterly <0.1 Quarterly Manganese mg/L Decon sinks x2 Washer x2 Ultrasonic x2 City Water Water Softener Utility DI After Carbon Filter Decon sinks x1 Washer x1 Supply After RO After DI Sterilizer x2 <0.1 Quarterly <0.1 Quarterly <0.1 Quarterly Nitrate mg/L Decon sinks x2 Washer x2 Ultrasonic x2 City Water Water Softener Utility DI After Carbon Filter Decon sinks x1 Washer x1 Supply After RO After DI Sterilizer x2 <10 Quarterly <1 Quarterly <1 Quarterly Phosphate mg/L Decon sinks x2 Washer x2 Ultrasonic x2 City Water Water Softener Utility DI After Carbon Filter Decon sinks x1 Washer x1 Supply After RO After DI Sterilizer x2 <5 Quarterly <1 Quarterly <1 Quarterly Silicate mg/L Decon sinks x2 Washer x2 Ultrasonic x2 City Water Water Softener Utility DI After Carbon Filter Decon sinks x1 Washer x1 Supply After RO After DI Sterilizer x2 <50 Quarterly <1 Quarterly <1 Quarterly Sulfate mg/L Decon sinks x2 Washer x2 Ultrasonic x2 City Water Water Softener Utility DI After Carbon Filter Decon sinks x1 Washer x1 Supply After RO After DI Sterilizer x2 <150 Quarterly <1 Quarterly <1 Quarterly Zinc mg/L Decon sinks x2 Washer x2 Ultrasonic x2 City Water Water Softener Utility DI After Carbon Filter Decon sinks x1 Washer x1 Supply After RO After DI Sterilizer x2 <0.1 Quarterly <0.1 Quarterly <0.1 Quarterly Bacteria CFU/mL Heterotrophic Plate Count (HPC) Decon sinks x2 Washer x2 Ultrasonic x2 City Water Water Softener Utility DI After Carbon Filter Decon sinks x1 Washer x1 Supply After RO After DI NA <500 Quarterly <10 Monthly NA NA Endotoxin EU/mL LAL Test NA Decon sinks x1 Washer x1 Supply After RO After DI NA NA NA <10 Monthly NA NA JWC Performance Qualification Testing (quarterly) Testing by Vendor Water Quality Measurement Units Type of Testing PQ Sample Site Utility PQ Sample Site Critical Minimum Frequency of Testing and Testing Range Utility Water Critical Water pH pH pH Meter or Colorimetric dipsticks Decon sink x2 Medivator Basin x4 Water Softener City Water Medivator 30 Supply Medivator 31 Supply NA 6.5-9.5 Quarterly 5.0-7.5 Monthly Conductivity uS/cm Conductivity Meter Decon sink x2 Medivator Basin x4 Water Softener City Water Medivator 30 Supply Medivator 31 Supply NA <500 Quarterly <10 Monthly Total Alkalinity mg CaCO3/L Colorimetric dipsticks or Alkalinity test kit Decon sink x2 Medivator Basin x4 Water Softener City Water NA <400 Quarterly <8 Monthly Total Hardness mg CaCO3/L Determination of ppm as CaCO3/L by colorimetric dipsticks, titration kit, or handheld meter Decon sink x2 Medivator Basin x4 Water Softener City Water NA <150 Quarterly <1 Monthly Total Organic Carbon (TOC) mg/L (ppm) NA NA NA NA <1.0 Quarterly Aluminum mg/L Decon sink x2 Medivator Basin x4 Water Softener City Water Medivator 30 Supply Medivator 31 Supply NA <0.1 Quarterly <0.1 Quarterly Chloride mg/L Decon sink x2 Medivator Basin x4 Water Softener City Water Medivator 30 Supply Medivator 31 Supply NA <250 Quarterly <1 Quarterly Copper mg/L Decon sink x2 Medivator Basin x4 Water Softener City Water Medivator 30 Supply Medivator 31 Supply NA <0.1 Quarterly <0.1 Quarterly Iron mg/L Decon sink x2 Medivator Basin x4 Water Softener City Water Medivator 30 Supply Medivator 31 Supply NA <0.1 Quarterly <0.1 Quarterly Manganese mg/L Decon sink x2 Medivator Basin x4 Water Softener City Water Medivator 30 Supply Medivator 31 Supply NA <0.1 Quarterly <0.1 Quarterly Nitrate mg/L Decon sink x2 Medivator Basin x4 Water Softener City Water Medivator 30 Supply Medivator 31 Supply NA <10 Quarterly <1 Quarterly Phosphate mg/L Decon sink x2 Medivator Basin x4 Water Softener City Water Medivator 30 Supply Medivator 31 Supply NA <5 Quarterly <1 Quarterly Silicate mg/L Decon sink x2 Medivator Basin x4 Water Softener City Water Medivator 30 Supply Medivator 31 Supply NA <50 Quarterly <1 Quarterly Sulfate mg/L Decon sink x2 Medivator Basin x4 Water Softener City Water Medivator 30 Supply Medivator 31 Supply NA <150 Quarterly <1 Quarterly Zinc mg/L Decon sink x2 Medivator Basin x4 Water Softener City Water Medivator 30 Supply Medivator 31 Supply NA <0.1 Quarterly <0.1 Quarterly Bacteria CFU/mL Heterotrophic Plate Count (HPC) Decon sink x2 Water Softener City Water Medivator 30 Supply Medivator 31 Supply Medivator Basin x4 <500 Quarterly <10 Monthly Endotoxin EU/mL LAL Test NA Medivator Basin x4 NA NA <10 Monthly JWC Routine (monthly) POU Testing by Vendor Water Quality Measurement Units Type of Testing Routine Sample Site Utility Routine Sample Site Critical Minimum Frequency of Testing and Testing Range Utility Water Critical Water pH pH pH Meter or Colorimetric dipsticks Medivator Basin x4 NA 6.5-9.5 Quarterly 5.0-7.5 Monthly Conductivity uS/cm Conductivity Meter Medivator Basin x4 NA <500 Quarterly <10 Monthly Total Alkalinity mg CaCO3/L Colorimetric dipsticks or Alkalinity test kit Medivator Basin x4 NA <400 Quarterly <8 Monthly Total Hardness mg CaCO3/L Determination of ppm as CaCO3/L by colorimetric dipsticks, titration kit, or handheld meter Medivator Basin x4 NA <150 Quarterly <1 Monthly Bacteria CFU/mL Heterotrophic Plate Count (HPC) NA Medivator Basin x4 <500 Quarterly <10 Monthly Endotoxin EU/mL LAL Test NA Medivator Basin x4 NA NA <10 Monthly Part III Supporting information: All work to be performed during the normal business hours of 7:00a 4:30p, Monday through Friday, excluding National Holidays. excluding federal holidays. Federal holidays are New Year's Day, Martin Luther King Day, President's Day, Memorial Day, Juneteenth, Independence Day, Labor Day, Columbus Day, Veterans Day, Thanksgiving Day, Christmas, and any other day specifically declared by the President of the United States to be a national holiday. Security: The C&A requirements do not apply, and a Security Accreditation Package is not required. Place of Performance: VA Northern Indiana Health Care System (Marion) 1700 E 38th St. Marion, Indiana 46953 VA Northern Indiana Health Care System (Fort Wayne) 2121 Lake Ave. Fort Wayne, Indiana 46805 Jackie Walorski VA Clinic (St. Joseph County VA Health Care Center) 1540 Trinity Place, Mishawaka, IN 46545. Submittals are required for all Products and Services: Provide three hard copies of all specifications of all products: name, ingredients, control range, test procedures, physical state (liquid, powder, etc.), container size, method of delivery, compatibility with present materials being used, catalog data, MSDS sheets

Update #1 ·

THIS IS A SOURCES SOUGHT ANNOUNCEMENT FOR INFORMATION ONLY. THIS IS NOT A SOLICITATION FOR PROPOSALS AND NO CONTRACT WILL BE AWARDED FROM THIS ANNOUNCEMENT.

The purpose of this notice is to conduct market research and obtain information from qualified businesses with special interest in seeking: VA certified Service-Disabled Veteran owned Small Business (SDVOSB), VA certified Veteran Owned Small Business (VOSB) and Small Business (SB) concerns. No solicitation is being issued at this time. All information submitted in response to this announcement is voluntary; the Government will not pay for information requested nor will it compensate any respondent for any cost incurred in developing information provided to the Government.

The Department of Veteran Affairs, Northern Indiana Health Care System (NIHCS) has a requirement for Water Testing and Reagents for the sterile processing system (SPS) at the Marion, Fort Wayne, and Jackie Walorski VA sites.

The appropriate NAICS Code 541380, Testing Laboratories, Size Standard $16.5 Million

Responses:

Please submit your response in accordance with the following instructions:

Responses shall be no more than 10 pages in length.

Responses are due by 4/22/2026 at 15:00 PM EST.

Responses shall be submitted via email to eric.sweatt@va.gov.

Emails cannot exceed 5mbÂ

Subject line of the Email shall include this Sources Sought Notice Number Firm Name NIHCS Water Testing and Reagents for the SPS Services.

If your response contains proprietary information you will place Proprietary Information in front of the Sources Sought Notice Number.

No marketing materials are allowed as part of this Sources Sought Notice. The Government will not review any other information or attachments included in your submission that is not listed herein or exceeds the 10-page limitation.

Contractors Federal Supply Schedule Contract that carries the desired services

If you do not have a Federal Supply Schedule contract, you shall state open market

Provide the Schedule

Provide expiration date of your Contract

Provide your GSA/FSS/SAC Contract Number

Provide the SIN for water testing and reagents for SPS services

If any of the items are not on your GSA/FSS contract, you will mark which items are open market.

Contractor Point of Contact to include:

Name

Title

Phone

Email

Company Information to include:

Name

Address

SAM UEI Number

Tax ID Number

Business Size

Provide information as to what water testing services are provided by your company.

Provide how your company invoices for services in the commercial marketplace.

Business Size/Type:Â Service-Disabled Veteran Owned Small Business (SDVOSB), Veteran Owned Small Business (VOSB), 8(a), Hub Zone Small Business, Women Owned Small Business (WOSB), Small Disadvantaged Business (SDB), Small Business, or Large Business.

All prospective Contractors are reminded that in accordance with FAR 4.11 they shall be registered in the System for Award Management (SAM) database and shall complete annual Representations and Certifications in SAM prior to any offer. SAM is located at http://www.sam.gov and consolidated the capabilities of CCR/FedReg, ORCA, and EPLS. There is no cost to use SAM.

FOR SDVOSB/VOSB ONLY:Â In order to be eligible, interested parties must be registered in the System for Award Management (SAM) data base at https://sam.gov. For firms claiming SDVOSB/VOSB classification status, they must also be registered and certified through the Department of Veterans Affairs Office of Small & Disadvantaged Business Utilization data base, located at https://veterans.certify.sba.gov/. Please include a copy of your certification with your response.

Contractor shall include 3 past contracts for the same or similar services. Contractor shall include at minimum:

Name of Company

Company POC

Company POC phone

Company POC email

Dollar amount of contract

Contract number

Period of Performance

Description of services provided

Contractor shall include any relevant comments about the attachments, if applicable.

Contractor shall provide a public price list/contract price list for the services described herein.

Contractor shall provide a detailed response to include your firm s ability, capability, and responsibility to deliver the services presented in this Sources Sought Notice.

Contractor shall provide if they plan to sub-contract all or part of the work, and to what firm to include the subs duns number and business size.

Determination by the Government does not compete with this proposed contract action based on responses to this notice is solely within the discretion of the Government.

STATEMENT OF WORK

SPS Water Testing Services

Part I. Description of Services:

The contractor shall provide all the following for testing the water for/within SPS per the ANSI/AAMI ST108:2023 at the Marion, Fort Wayne, and Jackie Walorski VA sites as defined in this Statement of Work (SOW).

WORK TO BE PERFORMED:

Contractors shall provide all labor, materials, equipment, transportation, lodging and supervision necessary to collect appropriate samples and deliver to a lab meeting requirement for processing of AAMI ST108.

Contractors shall calibrate the conductivity/resistivity meter monthly during their testing visit and provide a water analysis report.

The contractor will be part of the multidisciplinary team that assists the water working group with resolving out of range water testing, in addition to:

Serves as the subject matter expert on water testing per the ST108:2023

Serves as the subject matter expert on corrective actions and to mitigate risks for out-of-range water testing parameters.

Completes the monthly, quarterly, and yearly testing per the ST108:2023 and per the VHA water contract/SOW.

Completes performance qualification testing per the ST108:2023 and per the VHA contract/SOW.

Completes follow up testing as ordered by engineering for out-of-range water parameters per the ST108:2023 and VHA water contract/SOW.

Contractor shall provide consultative services as needed such as resolving abnormal water testing results, water system analysis, and mitigation of the tests performed

Part II Work Requirements

Water Treatment Testing Qualifications:

The contractor shall have been in business at least five years providing water treatment testing of water distribution systems similar in design at a medical center. Provide a list of at least five references similar to Department of Veterans Affairs who have utilized the services for water treatment in the last three years.

One water treatment specialist who has a bachelor s degree in chemical engineering or the physical sciences from an accredited college or university shall be assigned to this contract. The water treatment specialist must have at least five years successful experience with water treatment programs and shall furnish references from at least three customers who utilize water treatment programs.

The contractor must operate (or continuously maintain a contract with) a state of Indiana certified laboratory to perform complete scientific analyses of water chemistry, deposits, metallic materials affected by corrosion and provide recommendations for correction of problems related to the chemical and mechanical water treatment program. A Professional Chemist shall supervise the laboratory.

On-Site Services:

Provide complete written instructions, supplemented by oral instructions and demonstrations, to SPS personnel for their performance of routine water tests. The instructions cover all aspects of safe chemical handling and safe and accurate testing and methods to suit the SPS operating parameters and water treatment performance requirements.

Provide one training session within one week of the start of the contract. Provide one training session annually after the new contract has been awarded.

The vendor will provide a handheld resistivity meter as well as pH hardness colorimetric dipsticks to the VA. The vendor shall calibrate the resistivity meter on a monthly basis during their testing visit and provide a calibration certification with this visit as part of the report. The vendor shall provide a detailed operations manual and training to the VA staff for procedures related to VA required testing noted below:

AAMI ST108 UTILITY WATER, CRITICAL WATER, AND STEAM TESTING BY VENDOR:

Fort Wayne Routine (monthly) POU Testing by Vendor

Water Quality Measurement

Units

Type of Testing

Routine Sample Site

Utility

Routine Sample Site

Critical

Minimum Frequency of Testing and Testing Range

Utility Water

Critical Water

pH

pH

pH Meter or Colorimetric dipsticks

Medivator Basin x4

Decon sinks x3

Washer x1

Ultrasonic x1

Eye Ultrasonic X1

6.5-9.5

Quarterly

5.0-7.5

Monthly

Conductivity

uS/cm

Conductivity Meter

Medivator Basin x4

Decon sinks x3

Washer x1

Ultrasonic x1

Eye UltrasonicX1

Files

File Type Posted
36C25026R0060.docx DOCX document

Similar open opportunities

Opportunity Agency Set-aside Responses due
B533--VANIHCS SPS Water Testing (VA-27-00000760) VA Northern Indiana Healthcare System Fort Wayne Vamc, 46805 Marion, Indiana 46953, United States · NAICS 541380 Veterans Integrated Service Network 10 Service-Disabled Veteran-Owned Small Business
F109--Water Management Program and Chemical Testing VAMC Boston Healthcare System Department of Veterans Affairs Vamc Boston Healthcare System Boston MA 02130, United States · NAICS 561210 Veterans Integrated Service Network 1 Service-Disabled Veteran-Owned Small Business
Z1NE--RFQ Semi Annual Industrial Wastewater Discharge Testing Huntington VAMC Base Year Plus Four Option Years Department of Veterans Affairs Hershel "Woody" Williams Vamc Huntington, West Virginia 25704, United States · NAICS 561210 Veterans Integrated Service Network 5 No set-aside
H346--Legionella Water Testing United States · NAICS 541620 Veterans Integrated Service Network 22 Service-Disabled Veteran-Owned Small Business
FCC Terre Haute DRINK WATER TESTING FY27 Terre Haute, Indiana 47802, United States · NAICS 541380 Federal Correctional Complex Terre Haute Total Small Business

On GovTribe

Work this opportunity

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity