About this opportunity

The Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 1 is conducting market research to identify industry capabilities for replacing the outdated Milli-Q Direct 8 Water Purification System at the VA Boston Healthcare System Central Biorepository. This Sources Sought notice seeks information regarding vendors' interest, capabilities, and qualifications to provide and install a replacement water purification system with spare parts and warranty. The replacement system must integrate fully with existing infrastructure including plumbing, reservoir configuration, drain routing, electrical supply, and bench layout while matching or exceeding the Direct 8's production capacity and water quality specifications. Interested firms must be registered in the System for Award Management (SAM) and submit documentation demonstrating capability to perform in a hospital environment, including installation, training, and current contact information, along with their socio-economic status. The response deadline for this market research effort is December 8, 2025 at 2:00 PM Eastern Time. It is important to note that this is a Sources Sought notice for market research purposes only and not a formal solicitation; therefore, no proposals are being requested at this time and no contract will be awarded based on this notice.

This opportunity carries no set-aside designation and does not identify a current incumbent contractor by name. The replacement water purification system must produce both Type I and Type III laboratory-grade water with 18.2 megohm centimeter resistivity and provide approximately 8 liters per hour of purified water to minimize operational disruption. No specific award value or budget range is disclosed in this pre-solicitation notice, as this is preliminary market research to assess vendor capabilities and interest prior to formal solicitation development. The notice applies to analytical laboratory instrument manufacturers classified under NAICS Code 334516 and is categorized under Product Service Code 4610 for water purification equipment. While no formal solicitation timeline is provided, the December 8, 2025 response date indicates the government is gathering information to inform its procurement planning within the near term.

Notice text

Sources Sought #36C24125Q0868

THIS IS A SOURCES SOUGHT NOTICE ONLY: This announcement is seeking information from the industry, which will be used for preliminary planning purposes for procurement. No proposals are being requested or accepted with this synopsis. THIS IS NOT A SOLICITATION FOR PROPOSALS AND NO CONTRACT SHALL BE AWARDED FROM THIS SYNOPSIS. THIS IS STRICTLY MARKET RESEARCH.

The purpose of this Sources Sought is to gain knowledge regarding interest, capabilities, and qualifications of various members of industry to provide and install a replacement system for the existing outdated Milli-QÃ’ Direct 8 Water Purification System at the VA Boston HC System Central Biorepository, including spare parts and warranty.

To maintain compatibility with the existing Milli-Q Direct 8 infrastructure and avoid extensive renovations or downtime, the replacement system must: Integrate fully with the current plumbing, reservoir configuration, drain routing, electrical supply, and bench layout; match or exceed the Direct 8 s production capacity, water quality specifications, and operational functions, and prevent the need to purchase unrelated proprietary reservoirs, fittings, electrical components, or bench modifications. In addition, the replacement system sought is required to be compatible with the existing system s footprint to minimize impact on existing infrastructure and operational research disruption. Please see compatibility requirements listed below.

All interested firms must be registered in SAM (System for Award Management) to be eligible for the award of Government contracts. Interested parties responding to this Sources Sought request shall submit the following information at a minimum:

Company name and address point of contact with phone number and SAM UEI.

Documentation relating to capability of performance, including installation in a hospital environment, training, and a current point of contact and phone number.

Socio-economic status to identify the business class as being a large business, SDVOSB, VOSB, small business, Woman Owned, etc.

Detailed specifications of proposed product and installation including product part number, delivery, and support., cost estimate, estimated performance time, and country of manufacture.

Any SDVOSB/VOSB firms responding to this Sources Sought Notice must also provide proof of certification from SBA prior to submission of offers to a solicitation.

This market research is for informational and planning purposes. This is not a solicitation and does not obligate the Government to issue a solicitation. Please be advised that the government will not pay any costs for responses and information submitted. All submissions must reference Sources Sought #36C24126Q0806 and be submitted via email to emily.hatcher@va.gov.

The replacement system must meet or exceed the performance, interface, and installation characteristics listed below and must integrate with the laboratory s current configuration.

1. System Capability

Must produce both Type I (ultrapure) and Type III (reverse-osmosis or pretreated) laboratory-grade water.

Must achieve Type I water resistivity up to 18.2 M ·cm and low total organic carbon (TOC) suitable for sensitive research applications.

Must provide a continuous production rate similar to or greater than the existing Milli-Q Direct 8 (approximately 8 liters/hour).

Must include point-of-use dispensing capability or an equivalent mechanism.

2. Feed Water and Operating Requirements

Must operate using potable tap water as feed, within standard laboratory feed-water parameters (conductivity, pressure, and pH ranges suitable for RO-based systems).

Must function within a typical laboratory feed pressure range of approximately 1 6 bar (15 90 psi).

Must include or be compatible with pretreatment solutions to handle chlorine, hardness, or particulates as needed.

3. Installation Compatibility (Critical Requirement)

The replacement system must be compatible with the existing Milli-Q Direct 8 installation footprint and utilities, including:

Reservoir Compatibility

Must connect to the current reservoir or provide a fully compatible replacement reservoir.

Must interface with standard laboratory reservoir fittings, tubing diameters, and level-sensor configurations similar to those used on the Direct 8.

Plumbing

Must be compatible with existing feed-water plumbing, quick-connect fittings, and tubing sizes currently used by the Direct 8.

Must allow use of the existing drain line for waste/reject water without requiring facility modification.

Electrical

Must operate on standard laboratory power (approximately 100 240 VAC, 50/60 Hz) and must not require installation of dedicated circuits or new electrical infrastructure.

Bench / Physical Footprint

Must fit within the same bench-top or under-bench footprint as the current system or offer wall-mount options that do not require major remodeling.

Must be modular enough to integrate into the current working area without structural modification.

4. Consumable Management and Traceability

Must offer automated identification and tracking of consumables, including cartridges and filters, to ensure proper installation, compatibility, and timely replacement.

Must support auditability by providing automatic alerts and basic traceability for consumable replacement history.

5. Reliability and Serviceability

Must be designed for reduced maintenance frequency compared with the current 2010-era system.

Must allow routine maintenance (filter changes, sanitization, troubleshooting) to be performed without specialized tools or facility disruption.

Must have readily available consumables and service parts.

RESULTS INFORMATION: Respondents will not be notified of the results, and it is at the discretion of the Government to utilize results. Firms responding to this Sources Sought announcement, who fail to meet all requirements as stated above, and who fail to provide ALL the required information requested, will not be used to help the Government make the acquisition decision, which is the intent of this Sources Sought announcement.

Files

File Type Posted
36C24125Q0868_1.docx DOCX document

Similar open opportunities

Opportunity Agency Set-aside Responses due
F109--Water Management Program and Chemical Testing VAMC Boston Healthcare System Department of Veterans Affairs Vamc Boston Healthcare System Boston MA 02130, United States · NAICS 561210 Veterans Integrated Service Network 1 Service-Disabled Veteran-Owned Small Business
Reverse Osmosis System Modifications Minneapolis, Minnesota 55417, United States · NAICS 238220 Veterans Integrated Service Network 23 No set-aside
B533--VANIHCS SPS Water Testing (VA-27-00000760) Marion, Indiana 46953, United States · NAICS 541380 Veterans Integrated Service Network 10 Service-Disabled Veteran-Owned Small Business
H146--Hach Water Quality Monitoring Service NAICS 811310 Veterans Integrated Service Network 1 No set-aside
Osmometer Equipment Replacement (OsmoPro Max) Biloxi, Mississippi 39531, United States · NAICS 423450 Veterans Integrated Service Network 16 No set-aside

On GovTribe

Work this opportunity

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity