SPECIFICATIONS

o Bench Top Footprint: Maximum dimensions of 35"1x 40"W x 48"H
o Operating Voltage 120 VAC/60 Hz
o Nominal Pipette Volumes: 10 pl- 1.2 ml
o Linearity: <0.03 mg
o Repeatability: < 0.02 mg
o Resolution: 0.01 mg
o Maximum permissible error (k=1): < 0.02 mg (in the range of 0 -1.2 I | 0 -1,200 pl)
o Sensitivity to temperature drift: < I 1 + 10-6 K between +10'C and +30'C
o compliant with lso 8655-6:2002 specifications

DELIVERY

Delivered within 180 days of date of award