This document is a Sources Sought announcement for x-ray sources for the US Geological Survey's (USGS) Pacific Coastal and Marine Science Center (PCMSC). The USGS is conducting market research to acquire two specific x-ray sources: a Thermo Kevex 130 Kv X-ray Source for their Rotating X-Ray Computed Tomography (RXCT) scanner and a Cs-137 Source for their Multi-Sensor Core Logger (MSCL). The equipment, valued at approximately $1 million total, supports critical research in geologic hazard and mineral resource assessments.
The solicitation (No. 140G0325Q0096) requires potential vendors to submit detailed product specifications electronically to yangzhideng@usgs.gov by May 2, 2025, at 3:00 PM PST. The announcement uses NAICS code 334517 (Irradiation Apparatus Manufacturing) with a size standard of 1,200 employees, and Product Service Code 6525 for imaging equipment. This is a market research effort only, and the government is not committed to an award. Responses must include technical capabilities to meet the specified requirements, which include precise x-ray source specifications such as tube voltage range, power output, and specific technical parameters for the imaging equipment.
View the file
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
The USGS Pacific Coastal and Marine Science Center (PCMSC) acquired a GeoTek Rotating X-Ray Computed Tomography (RXCT) scanner in 2018 for approximately $650,000 and GeoTek Multi-Sensor Core Logger (MSCL) in 2003 (upgraded in 2012, and 2024) for approximately $350,000 (amount includes upgrade costs). The RXCT is only CT machine across all USGS labs. By having both the RXCT and MSCL together in the same lab, the PCMSC Core Lab stands out as a top core lab in the United States. These machines routinely support research regarding, geologic hazard and mineral resource assessments across several USGS mission areas. Therefore, these GeoTek scanners not only help the USGS protect human life and infrastructure but also facilitate management of mineral resources. The required items and the technical specifications are the following:
· Thermo Kevex 130 Kv X-ray Source for RXCT S/N21
· Cs-137 Source for MSCL-S and MSCL-SXZ
· Disposal of Old Cs-137 Source
· Onsite installation/system realignment and training
1. The Thermo Kevex PXS10-WB X-ray source is the only source that Geotek uses in their Rotating X-ray Computed Tomography System.
2. Maximum tube voltage: 130 Kv
3. Operating volt range: 45-130 Kv
4. Tube current operational range: 0-500uA µA
5. Maximum Power output: 65W
6. Minimum spot size of 6 microns at 4 watts, for high-resolution imaging
7. X-ray beam angle: 115°, round beam
8. Focus to object distance: 10.0 ±0.5mm
9. Target material: Tungsten
10. X-ray output window material: Beryllium
11. Input power: 24-26 VDC, ^A max
12. Control interface: RS-232
13. Small, round spot optimized over the range of operating voltage and power, for sharp, distortion-free images
14. 14mm spot to window spacing, for high geometric magnification
15. X-ray tube, power supply, and control electronics combined into one compact package, for easy system integration easy
16. Option to rotate the tube head +90° or -90° relative to home position, for flexible configuration and installation
17. Auto-conditioning, for protection of the source during start-up
18. Digital interface, for easy access to diagnostics and operating logs