The document is a Statement of Work (SOW) for a requirement to provide four (4) Wet Bulb Globe Temperature (WBGT) instruments. The USAFSAM/PHD organization requires these WBGT instruments to support training for preventive medicine personnel operating in contingency environments. The WBGT instruments must meet specific salient characteristics, including the ability to measure globe temperature, dry bulb temperature, wet bulb temperature, relative humidity, and provide a WBGT index. The instruments must be water-resistant, weigh no more than 3 lbs, be battery-operated, and have an operating temperature range of -5°C to +100°C. The instruments shall also include data logging and event logging capabilities. This SOW is related to the attached federal contract opportunity, Solicitation Number COMBO-AFRL-PZLEQ-2024-0008, issued by the Department of the Air Force Materiel Command Research Laboratory for the procurement of the WBGT meters.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| RFQ (Over Micro - 25K) Wet Bulb Globe Temperature.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Minimum Requirements Document
Wet Bulb Globe Temperature Heat Stress Monitor
Document Date: 16 April 2024
Background & Scope: USAFSAM/PHD requires the new capability to perform training for preventive medicine courses regarding the effects of heat stress on personnel using a Wet Bulb
Globe Temperature (WBGT) instrument. This new requirement is a part of a series of tasks identified by the Career Field Manager (CFM) that are deemed necessary for preventive medicine personnel to operate proficiently while located in a contingency environment.
Requirements: The vendor shall provide a quantity of Four (4) Wet Bulb Globe Temperature (WBGT) instruments.
Salient Characteristics:
• Shall include Globe Temperature, Dry Bulb Temperature, Wet Bulb Temperature, % of relative humidity and a WBGT index.
• Shall provide temperature in Fahrenheit or Celsius
• Shall be designed to be water resistant to light rain or mist.
• Shall be no more than 3lbs.
• Shall be field portable.
• Shall be battery operated.
• Shall have English as its operable language.
• Shall have an operating temperature range of –5°C to +100°C.
• Shall have a dry bulb sensor accuracy and range of ±0.5 from 0°C to 120°C.
• Shall have a globe sensor accuracy and range of ±0.5 from 0°C to 120°C.
• Shall have a relative humidity sensor accuracy and range of ±5% from 20 % to 95% (non-condensing.
• Shall have a waterless wet bulb (humidity) sensor accuracy and range of 1.1°C between 0°C and 80°C.
• Shall include data logging at user selected intervals and an event logging mode.
• Shall contain an index of U.S. Navy PHEL Charts, ACGIH TLV, and flag conditions for the U.S. Navy/Marine Corp Ashore.