Water Treatment Trailer - Salient Characteristics.xlsx
XLSX spreadsheet 16 KB Posted
- Attached to
- Fort Bliss Water Purification System Unit
- Solicitation number
- PANHCA-26-P-0000_031551
- Issued by
-
Department of the Army Medical Command
About this file
This is a salient characteristics document for a water treatment trailer system required by the William Beaumont Army Medical Center (WBAMC) at Fort Bliss, Texas. The document outlines the technical and operational specifications for a multi-stage water filtration system designed to connect to leased sterilization trailers in the Sterile Processing Department. The specifications detail the system's filtration capabilities, water quality output standards, and integration requirements with existing sterilization equipment. Key characteristics include filtration stages for sediment removal, activated carbon filtration, and membrane-based purification to meet medical-grade water standards required for sterilization processes.
The underlying federal contract opportunity is a Sources Sought Notice issued by the Medical Readiness Contract Office-W, representing preliminary market research rather than a formal solicitation. The anticipated contract will be Firm Fixed-Price and is classified under NAICS code 221310 (Water Supply and Irrigation System) with a small business size standard of $41 million. The Government is seeking capabilities and qualifications from both small business entities (including 8(a), HUB-Zone, Women-Owned Small Business, and Service-Disabled Veteran-Owned Small Business) and large businesses to establish an adequate competitive pool. The type of formal solicitation issued will depend on responses received to this sources sought notice, and potential offerors are prohibited from contacting Government personnel other than the identified Contract Specialist.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Sources Sought Notice Water treatment Unit_WBAMC.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Sheet3
| Salient Characteristics | ||||
| Palce of Performance: WBAMC Loading Dock | Date:5/7/2026 | |||
| Requirement POC: | Quantity: 1 | |||
| Description: Water treatment Trailer | RP#: | |||
| Summary: Mobile Sterilization trailers will need a seperate means of treating their supply water do to the special requirements needed for the cold water, hot water, and RO for final rinse. | ||||
| # | Description of Minimum Requirement | |||
| and/or Objective | Minimum Requirement | Objective | ||
| 1 | 60 gpm RO unit | 1 | 60 gpm RO unit. Permeate should be able to drain to nearby trench drain. Use | |
| 1a | RO membranes | 8 | Replacement RO Filters | |
| 2 | 2000 Gal Water storage tank | 1 | 2000 gal Water storage tank | |
| 3 | Chemical Injection Quill | 1 | Chemical Injection Quill | |
| 4 | Multibag Filter Housing | 1 | Housing to use 40 inch filters for prefiltering water of sediment and particles. | |
| 4a | Bag filters | 100 | .1 Micron Absolute rating | |
| 4b | Filters | 100 | .5 micron Absolute 40-inch filters | |
| 5 | 7.5 HP Water Pump | 1 | 7.5 HP vertical distribution pump | |
| 6 | VFD | 1 | 10 HP 460 V | |
| 7 | Bladder tank | 1 | 44 Gallon stainless steel vessel w/ bladder | |
| 8 | Globe valve | 1 | 1" for blending water to meet Water quality requirement of sterilization equipment. | |
| 9 | TDS meter | 1 | GF Signet 9950 to measure TDS in mg/l | |
| 10 | Water softner system | 1 | Water softner to soften water going to RO and water Heaters. | |
| 11 | Electric Heaters | 4 | 120 v electric heaters | |
| 12 | 120 Gal Electric Water Heaters | 2 | Hot water supply. Must be able to keep the hot water between 120-140 degrees Faranheit. | |
| 13 | 2 gpm Recirculating pump | 1 | For recirculating hot water | |
| 14 | Containment | Equipment should fit inside of a 52' tractor trailer with a wood flooring so holes can be made for routing pipe. | ||
| 15 | Maintenance | Maintenance plan should be provided to maintain the individual equipment and the structure. Maintence plan to include on-call personel. | ||
| 16 | Installation | Must include turnkey installation of equipment. Must include connection of equiopment to each other and to the mobile sterilization unit(s). | ||
| 17 | Site Visit | Must include site visit to see availability of exsisting utilites and availble space. | ||
| 18 | Monthly water testing | Monthly water test to test cold water and hot water supply. | ||
| 18a | Min Sediment Filtration | MIN-MAX | Cold Water: 0-5 Micron |
Hot Water: 0-5 Micron 18b Total Hardness as CaCo3 MIN-MAX Cold Water: 50-170 ppm CaCo3 Hot Water: 0-17 ppm CaCo3 18c Total Dissovled Solids MIN-MAX Cold Water: 100-170 ppm Hot Water: 50-250 ppm 18d Total Alkalinity as CaCo3 MIN-MAX Cold Water: 70-180 ppm Hot Water: 50-180 ppm 18e pH MIN-MAX Cold Water: 6.5-8.5 Hot Water: 6.5-9.5 18f Conductivity MIN-MAX Cold Water: 0-1000 umhos/cm Hot Water: 40-500 umhos/cm 18g Total Silica MIN-MAX Cold Water: 0-50 ppm Hot Water: 0.1-2.5ppm 18h Chloride MIN-MAX Cold Water: 0-10 ppm Hot Water: 0-10 ppm 18i Cu-Copper MIN-MAX Cold Water: 0-0.1 ppm Hot Water: 0-0.1 ppm 18j Fe-Iron MIN-MAX Cold Water: 0-0.1 ppm Hot Water: 0-0.1 ppm 18k Zn-Zinc MIN-MAX Cold Water: 0-0.1 ppm Hot Water: 0-0.1 ppm 18l Al-Aluminium MIN-MAX Cold Water: 0-0.1 ppm Hot Water: 0-0.1 ppm 18m Mn-Manganese MIN-MAX Cold Water: 0-0.1 ppm Hot Water: 0-0.1 ppm
| 18n | Color/Clarity | Water must be colorless |
| 19 | covering | Must include seperate raised platform with a roof and shall be enclosed to prevent the outside enviorment from entering the space. The platform shall have a width of 120" and the floor at a height of 47" to include a roof that extends an additional 120" from the platforms floor. Floor shall be hospital grade and include air curtain to help space pressure. Will need to be made out of metal with steel beams to withstand the elements |
Sheet2
| Cold Utility | Hot Utility | ||||
| Specification | Units | min | max | min | max |
| Min Sediment Filtration | Micron | 0 | 5 | 0 | 5 |
| Total Hardness as CaCo3 | ppm CaCo3 | 50 | 170 | 0 | 17 |
| Total Dissovled Solids | ppm | 100 | 500 | 50 | 250 |
| Total Alkalinity as CaCo3 | ppm | 70 | 180 | 50 | 180 |
| pH | 6.5 | 8.5 | 6.5 | 9.5 | |
| Conductivity | umhos/cm | 0 | 1000 | 40 | 500 |
| Total Silica | ppm | 0 | 50 | 0.1 | 2.5 |
| Chloride | ppm | 0 | 10 | 0 | 10 |
| Cu-Copper | ppm | 0 | 0.1 | 0 | 0.1 |
| Fe-Iron | ppm | 0 | 0.1 | 0 | 0.1 |
| Zn-Zinc | ppm | 0 | 0.1 | 0 | 0.1 |
| Al-Aluminium | ppm | 0 | 0.1 | 0 | 0.1 |
| Mn-Manganese | ppm | 0 | 0.1 | 0 | 0.1 |
| Color/Clarity | Water should be colorless |