Water Treatment Trailer - Salient Characteristics.xlsx

XLSX spreadsheet 16 KB Posted

Attached to
Fort Bliss Water Purification System Unit
Solicitation number
PANHCA-26-P-0000_031551
Issued by

About this file

This is a salient characteristics document for a water treatment trailer system required by the William Beaumont Army Medical Center (WBAMC) at Fort Bliss, Texas. The document outlines the technical and operational specifications for a multi-stage water filtration system designed to connect to leased sterilization trailers in the Sterile Processing Department. The specifications detail the system's filtration capabilities, water quality output standards, and integration requirements with existing sterilization equipment. Key characteristics include filtration stages for sediment removal, activated carbon filtration, and membrane-based purification to meet medical-grade water standards required for sterilization processes.

The underlying federal contract opportunity is a Sources Sought Notice issued by the Medical Readiness Contract Office-W, representing preliminary market research rather than a formal solicitation. The anticipated contract will be Firm Fixed-Price and is classified under NAICS code 221310 (Water Supply and Irrigation System) with a small business size standard of $41 million. The Government is seeking capabilities and qualifications from both small business entities (including 8(a), HUB-Zone, Women-Owned Small Business, and Service-Disabled Veteran-Owned Small Business) and large businesses to establish an adequate competitive pool. The type of formal solicitation issued will depend on responses received to this sources sought notice, and potential offerors are prohibited from contacting Government personnel other than the identified Contract Specialist.

View the file

Other files

File Type Posted
Sources Sought Notice Water treatment Unit_WBAMC.docx DOCX document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Sheet3

Salient Characteristics
Palce of Performance: WBAMC Loading DockDate:5/7/2026
Requirement POC:Quantity: 1
Description: Water treatment TrailerRP#:
Summary: Mobile Sterilization trailers will need a seperate means of treating their supply water do to the special requirements needed for the cold water, hot water, and RO for final rinse.
#Description of Minimum Requirement
and/or ObjectiveMinimum RequirementObjective
160 gpm RO unit160 gpm RO unit. Permeate should be able to drain to nearby trench drain. Use
1aRO membranes8Replacement RO Filters
22000 Gal Water storage tank12000 gal Water storage tank
3Chemical Injection Quill1Chemical Injection Quill
4Multibag Filter Housing1Housing to use 40 inch filters for prefiltering water of sediment and particles.
4aBag filters100.1 Micron Absolute rating
4bFilters100.5 micron Absolute 40-inch filters
57.5 HP Water Pump17.5 HP vertical distribution pump
6VFD110 HP 460 V
7Bladder tank144 Gallon stainless steel vessel w/ bladder
8Globe valve11" for blending water to meet Water quality requirement of sterilization equipment.
9TDS meter1GF Signet 9950 to measure TDS in mg/l
10Water softner system1Water softner to soften water going to RO and water Heaters.
11Electric Heaters4120 v electric heaters
12120 Gal Electric Water Heaters2Hot water supply. Must be able to keep the hot water between 120-140 degrees Faranheit.
132 gpm Recirculating pump1For recirculating hot water
14ContainmentEquipment should fit inside of a 52' tractor trailer with a wood flooring so holes can be made for routing pipe.
15MaintenanceMaintenance plan should be provided to maintain the individual equipment and the structure. Maintence plan to include on-call personel.
16InstallationMust include turnkey installation of equipment. Must include connection of equiopment to each other and to the mobile sterilization unit(s).
17Site VisitMust include site visit to see availability of exsisting utilites and availble space.
18Monthly water testingMonthly water test to test cold water and hot water supply.
18aMin Sediment FiltrationMIN-MAXCold Water: 0-5 Micron

Hot Water: 0-5 Micron 18b Total Hardness as CaCo3 MIN-MAX Cold Water: 50-170 ppm CaCo3 Hot Water: 0-17 ppm CaCo3 18c Total Dissovled Solids MIN-MAX Cold Water: 100-170 ppm Hot Water: 50-250 ppm 18d Total Alkalinity as CaCo3 MIN-MAX Cold Water: 70-180 ppm Hot Water: 50-180 ppm 18e pH MIN-MAX Cold Water: 6.5-8.5 Hot Water: 6.5-9.5 18f Conductivity MIN-MAX Cold Water: 0-1000 umhos/cm Hot Water: 40-500 umhos/cm 18g Total Silica MIN-MAX Cold Water: 0-50 ppm Hot Water: 0.1-2.5ppm 18h Chloride MIN-MAX Cold Water: 0-10 ppm Hot Water: 0-10 ppm 18i Cu-Copper MIN-MAX Cold Water: 0-0.1 ppm Hot Water: 0-0.1 ppm 18j Fe-Iron MIN-MAX Cold Water: 0-0.1 ppm Hot Water: 0-0.1 ppm 18k Zn-Zinc MIN-MAX Cold Water: 0-0.1 ppm Hot Water: 0-0.1 ppm 18l Al-Aluminium MIN-MAX Cold Water: 0-0.1 ppm Hot Water: 0-0.1 ppm 18m Mn-Manganese MIN-MAX Cold Water: 0-0.1 ppm Hot Water: 0-0.1 ppm

18nColor/ClarityWater must be colorless
19coveringMust include seperate raised platform with a roof and shall be enclosed to prevent the outside enviorment from entering the space. The platform shall have a width of 120" and the floor at a height of 47" to include a roof that extends an additional 120" from the platforms floor. Floor shall be hospital grade and include air curtain to help space pressure. Will need to be made out of metal with steel beams to withstand the elements

Sheet2

Cold UtilityHot Utility
SpecificationUnitsminmaxminmax
Min Sediment FiltrationMicron0505
Total Hardness as CaCo3ppm CaCo350170017
Total Dissovled Solidsppm10050050250
Total Alkalinity as CaCo3ppm7018050180
pH6.58.56.59.5
Conductivityumhos/cm0100040500
Total Silicappm0500.12.5
Chlorideppm010010
Cu-Copperppm00.100.1
Fe-Ironppm00.100.1
Zn-Zincppm00.100.1
Al-Aluminiumppm00.100.1
Mn-Manganeseppm00.100.1
Color/ClarityWater should be colorless