Water Filtration Contract - Attachment B - Testing Requirements.pdf
PDF 345 KB Posted
- Solicitation number
- NOI-CC-25-006804
About this file
This document is a technical standard from the Association for the Advancement of Medical Instrumentation (ANSI/AAMI ST108:2023) that details water quality requirements for medical device processing. The standard defines three categories of water quality: Utility Water, Critical Water, and Steam, each with specific performance qualification levels for parameters like pH, total alkalinity, bacteria count, endotoxin levels, and ionic contaminants.
The document provides comprehensive guidance for healthcare facilities on water quality monitoring, including recommended minimum testing frequencies at water generation systems and point-of-water-use locations. Monitoring requirements vary by water category, with Critical Water requiring more frequent testing (monthly) compared to Utility Water (quarterly). The standard emphasizes the importance of a planned, systematic quality improvement process, risk analysis, and ongoing verification of water quality to ensure patient safety during medical device processing.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Water Filtration Contract - Attachment A - List of Supported Equipment.xlsx | XLSX spreadsheet | |
| 1. SOW - Water Filtration Contract - PWS.docx | DOCX document | |
| 1. JOFOC - Sole Source - redact.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
e) ineffective cleaning/disinfection due to water contaminants interfering with device processing chemicals.
6 Categories of water quality for medical device processing
6.1 Introduction
The water quality needed for the various stages of medical device processing is determined by the type of medical device and by the disinfection or sterilization process used. For example, the processing of stainless steel surgical instruments that are steam sterilizable has different water quality requirements than the processing of flexible endoscopes, which require high-level disinfection or low-temperature sterilization.
6.2 Three categories of water quality
Table 2 describes three categories of water quality in terms of the characteristics that are important for medical device processing and the level of treatment that may be needed:
a) Utility Water: water as it comes from the tap that can require further treatment at the facility to achieve water quality measurement values outlined in Table 2. This water is mainly used for flushing, washing, and intermediate rinsing (e.g., rinsing between cleaning and disinfection).
NOTE The decision regarding the need to treat incoming tap water to provide adequate water quality for medical device processing should be undertaken in every facility that processes medical devices.
b)- Critical Water: water that meets the water quality measurement values detailed in Table 2; to achieve these values, the water generally requires extensive treatment by a multi-step process that can include a pretreatment, does include primary treatment (e.g., RO and/or DI), storage, distribution, and can include final treatment to provide a level of assurance that microorganisms and inorganic and organic material are removed from the water. See section 8.3.2.2 for recommended treatment methods. This water is mainly used for the final rinse after high-level disinfection, for the final rinse for critical devices prior to sterilization and feedwater for process steam production.
NOTE Using Critical Water for all stages of medical device processing can be unnecessary, costly, and can cause damage to water system or processing equipment (see Annex D).
c) Steam: vaporized water that is produced by a centralized boiler or a generator/heat exchanger near the sterilizer. When the steam is tested as a condensate, it should meet the specified criteria as defined in Table 2.
NOTE See i Annex D for additional information.
Utility Water is predominantly used for medical device processing with the exception of final rinse, where Critical Water is recommended. At the point of production, Utility Water and Critical Water should meet the requirements as specified in Table 2. Tap water can require treatment to meet these requirements. The bacteria level for Critical Water in Table 2 (< 10 CFU/mL [< 10,000 CFU/L]) is consistent with the Clinical & Laboratory Standards Institute (CLSI) guidelines.
[46]
© 2023 Association for the Advancement of Medical Instrumentation ■ ANSI/AAMI ST108;2023 17 cosgroveca Highlight cosgroveca Highlight cosgroveca Highlight
Table 2—Categories and performance qualification levels of water quality for medical device processing
Water Quality Measurement Units Utility Water Critical Water Steam* pH @ 25 °C: pH 6.5 - 9.5 5.0 - 7.5 5.0 — 9.2**
Total Alkalinity mg CaCO3/L <400 <8 <8
Bacteria CFU/mL <500*** <10 N/A
Endotoxin EU/mL N/A*** <10 N/A Total Organic Carbon
(TOC) mg/L (ppm) N/A <1.0 N/A
Color and Turbidity Visual Colorless, clear, without sediment
Colorless, clear, without sediment
Colorless, clear, without sediment
Ionic Contaminants
Aluminum mg/L <0.1 <0.1 <0.1
Chloride mg/L <250 <1 <1
Conductivity µS/cm <500 <10 <10
Copper mg/L <0.1 <0.1 <0.1
Iron mg/L <0.1 <0.1 <0.1
Manganese mg/L <0.1 <0.1 <0.1 , Nitrate mg/L <10 <1 <1
Phosphate mg/L <5 <1 <1
Sulfate mg/L <150 <1 <1
Silicate mg/L <50 <1 <1
Total Hardness mg CaCO3/L <150**** <1 <1
Zinc mg/L <0.1 _ <0.1 <0.1
* Steam parameters are for monitoring as steam condensate **See 6:3 a) for specifics. For local steam generation, the condensate pH should be 5.0 to 7.5. For boiler-treated steam, most boilers should be treated to maintain a condensate pH of 7.5 to 9.2 ***When Utility Water is used after chemical high-level disinfection as a final rinse, the bacteria should be <10 CFU/mL and endotoxin <10 EU/mL **** If hardness is greater than 150 mg/L a water softener is recommended unless used for washing where the cleaning chemistry is capable of handling higher levels of hardness
6.3 Rationale for Table 2 water quality parameters and other considerations:
a) pH: Water having extremes in pH can cause damage of device surfaces. In addition, an extreme pH can interfere with the efficacy of cleaning agents and disinfectants (see E.2.18). Utility water is expected to have a pH in the range of 6.5 to 9.5. The pH of the water will be dependent on the disinfection chemistry (e.g., chlorine, monochloramine) and water treatment strategy of the local municipality. Potable water is typically in the higher portion of this range. For Critical Water the specification is based upon typical values for the water produced. For boiler-treated steam that travels through an extensive distribution network that can be oxidized (e.g., black iron), acidic pH can result in corrosion of the piping and transfer of rust to the sterilizer load, therefore pH should be controlled to maintain a pH of 7.5 to 9.2. For steam produced by a dedicated generator the pH can be in the range of 5.0 to 7.5 (see EN 285).
b) Alkalinity: Total alkalinity is the measure of dissolved solids which are alkaline in nature such as bicarbonate, carbonate, and hydroxide. High alkalinity levels increase the tendency of hard water salts to form scale. High alkalinity in the final rinse water increases potential for concentration of alkaline dissolved solids onto the device during drying. This dried alkaline residue can then be reactivated under the high temperature and humidity conditions of a steam sterilizer causing devices to become stained or corroded. The performance of
18 © 2023 Association for the Advancement of Medical Instrumentation • ANSI/AAMI ST108:2023 birrieltorresgj processing department, infection prevention and control) and integrated into the overall quality process in the health care facility. Variables in the system may be controlled to achieve assurance of product quality and process efficacy.
Monitoring frequency will vary, depending on the quality improvement goals, the health care facility policies and procedures for the handling of unfavorable or unplanned events, and the type of process variable.
An investigation should be conducted for any problem relating to any aspect of water quality. Problems and updates should be reported to those involved with medical device processing.
There should be a planned, systematic, and ongoing process for verifying compliance with procedures. Quality processes can be enhanced by audits that are conducted on a regular basis. The information from these activities should be summarized and made available to individuals, groups, or teams as needed.
Table 5 and Table 6 describe the water quality monitoring that shall be performed as part of a quality improvement process within their area. Continuous quality improvement involves conducting risk analyses (see Clause 5).
Table 5—Frequency for water quality monitoring at water generation system
Water quality measurement Type of testing
Routine monitoring sampling site
Minimum frequency of testing*
Utility Water Critical Water pH pH meter** or Colorimetric dipsticks (sample tested within 15 minutes)
After the last treatment step
Quarterly Monthly
Conductivity Conductivity meter (in line or by measurement of a collected sample)
After the last treatment step, Storage tanks (if used)
Quarterly Daily
Total Alkalinity Colorimetric dipsticks Alkalinity test kit**
After the last treatment step, storage tanks (if used)
Quarterly Monthly
Total Hardness
Determination of ppm as CaCO3 by Calorimetric dipsticks, Titration kit**, or Handheld meter**
After the last treatment step
Quarterly Monthly
Bacteria Heterotrophic plate count (see Annex H)
Loop out and loop return points
N/A Monthly
Endotoxin LAL test (see Annex H) Loop out and loop return points
N/A Monthly
*NOTE 1 The recommendations for frequency of testing in this table are the recommended minimum frequency. If problems or issues arise with the water quality, it may be necessary to increase the frequency until they are resolved.
*"NOTE 2 When using these tests, the user should carefully follow the manufacturer's written IFU for accurate results. When measuring Critical Water levels, the water sample must be filled to the brim for these test kits, sealed and, if the sample is hot, allowed to cool to room temperature before testing. Testing within 15 minutes of sample collection prevents carbon dioxide absorption that can result in'an inaccurate pH.
34 © 2023 Association for the Advancement of Medical Instrumentation ■ ANSI/AAMI ST108:2023 cosgroveca
Table 6—Frequency for water quality monitoring at point-of-water-use
Water quality measurement Type of testing
Routine monitoring sampling site
Minimum frequency of testing*
Utility Water
Critical Water
Steam pH_ pH meter** or Colorimetric dipsticks (sample tested within 15 minutes)
At the point the distribution loop enters the processing area or first POU on the distribution loop
Quarterly Monthly Quarterly
Conductivity Conductivity meter** or Colorimetric dipsticks
At the point the distribution loop enters the processing area or first POU on the distribution loop
Quarterly Monthly Quarterly
Total Alkalinity Colorimetric dipsticks or Alkalinity test kit**
At the point the distribution loop enters the processing area or first POU on the distribution loop
Quarterly Monthly Quarterly
Total hardness
Determination of ppm as CaCO3 by Calorimetric dipsticks, Titration kit** ,or Handheld meter**
At the point the distribution loop enters the processing area or first POU on the distribution loop
Quarterly Monthly Quarterly
Bacteria Heterotrophic plate count (see Annex H)
Each location of point-of-use in department
Quarterly Monthly N/A
Endotoxin LAL test (see Annex H) Each location of point-of-use in department
N/A Monthly N/A
Visual Inspection
Visual Inspection of inside of equipment - Look for residue, staining, scaling, and discoloration (Annex I)
Spray Arms/Inside Chamber Walls/Inside Interior of Machine
Daily Daily Daily
*NOTE 1 The recommendations for frequency of testing in this table are the recommended minimum frequency. If problems or issues arise with the water quality, it may be necessary to increase the frequency until they are resolved.
**NOTE 2 Test type needed to measure Critical Water and Steam levels. Steam condensate must be filled to the brim, sealed, and allowed to cool before testing to prevent carbon dioxide absorption.
The minimum frequency of testing aligns with established water monitoring programs performed during medical device manufacturing. Device processing is an extension of the medical device supply chain, so a continuation of the water quality requirements is necessary for continued patient safety. The testing frequency described in Table 5 and Table 6 are recommended minimum frequencies to demonstrate the water system remains in a state of control. The risk assessment performed by the multidisciplinary team may identify sampling points that will require an increase in testing frequency (e.g., less frequently used point-of-use, portion of circulation loop containing a dead-leg).
© 2023 Association for the Advancement of Medical Instrumentation ■ ANSI/AAMI ST108:2023 35 birrieltorresgj
| Page 1 |
| Multi-Disciplinary Group Responsibilies |
| Frequency for Water Quality Monitoring |
| Table 2: Water Quality Summary |
| Table 5: Frequency of Monitoring: Generation System |
| Table 6: Frequency of Monitoring: Point of Use |
| Annex B: Risk Analysis |
| Proposed Validation Plan |
| Page 2 |
| Page 3 |
| Page 4 |
| Page 5 |
| Page 6 |
| Page 7 |
| Page 8 |
| Page 9 |
| Page 10 |
| Page 11 |
| Page 12 |
| Page 13 |
| Page 14 |
| Page 15 |
| Page 16 |
| Page 17 |
| Page 18 |
| Page 19 |
| Page 20 |
| Page 21 |
| Page 22 |
| Page 23 |
| Page 24 |
| Page 25 |
| Page 26 |
| Page 27 |
| Page 28 |
| Page 29 |
| Page 30 |
| Page 31 |
| Page 32 |
| Page 33 |
| Page 34 |
| Page 35 |
| Page 36 |
| Page 37 |
| Page 38 |
| Page 39 |
| Page 40 |
| Page 41 |
| Page 42 |
| Page 43 |
| Page 44 |
| Page 45 |
| Page 46 |
| Page 47 |
| Page 48 |
| Page 49 |
| Page 50 |
| Page 51 |
| Page 52 |
| Page 53 |
| Page 54 |
| Page 55 |
| Page 56 |
| Page 57 |
| Page 58 |
| Page 59 |
| Page 60 |
| Page 61 |
| Page 62 |
| Page 63 |
| Page 64 |
| Page 65 |
| Page 66 |
| Page 67 |
| Page 68 |
| Page 69 |
| Page 70 |
| Page 71 |
| Page 72 |
| Page 73 |
| Page 74 |
| Page 75 |
| Page 76 |
| Page 77 |
| Page 78 |
| Page 79 |
| Page 80 |
| Page 81 |
| Page 82 |
| Page 83 |
| Page 84 |
| Page 85 |
| Page 86 |
| Page 87 |
| Page 88 |
| Page 89 |
| Page 90 |
| Page 91 |
File details come from the government source that posted it. Updated .