Statement of Work PURCHASE of a LC/MSD System with DAD (PR# 11208702)

1. Requirement. The contractor shall provide timely delivery of Government requested material to the address provided below unless otherwise specified by the Government. The Government is procuring one

(1) Liquid Chromatograph-Prep-Mass Spectrometer Detection System as specified below.

Line

Part Number

Part Description

Total Quantity

1
G6125BW
LC/MSD Bundle Including LC/MSD and 1260 Infinity II LC System w/OpenLAB CDS ChemStation
1
2
N/A
Installation and Familiarization of Unit
1

a. The items listed above shall be received approximately 25 days after award or sooner.

2. Government Furnished Property or Assistance. Not applicable. No requirement for Government Furnished Property.

3. DELIVERY/LOGISTICS.

3.1 The contractor shall deliver the data/supplies/services approximately 25 days after award to:

Attn: Nicole Rippeon U.S. Army, Edgewood Chemical Biological RDCB-DRC 8510 Ricketts Point Road

BLDG E3400

Gunpowder, MD 21010 410-436-7291

3.2 Inspection and Acceptance of the items will be conducted at destination by authorized ECBC government personnel.

4. Safety Security Considerations. None

5. Security Considerations. None

6. Government Point of Contact.

U.S. Army - RDECOM

RDCB-DRC

Nicole Rippeon Bldg E3400 5183 Blackhawk Road

APG, MD 21010-5424

410-436-7291

Specifications for LC/MSD System with DAD System

General System Specifications

Single point of control
Single Point data system method capability with full control
Supported Data Systems
Must have OpenLab CDS ChemStation Edition and MassHunter Data Analysis capabilities
Time Programming
Must have:

SIM Ion Selection Scan Functions Fragmentor Voltage External CAN valves MS Diverter Valve Drying Gas Glow Nebulizer Pressure Drying Gas Temperature Capillary Voltage (Positive) Capillary Voltage (Negative)

Wide Range of Ionization Sources
Must have:

Electrospray (ESI) Multimode Source (Simultaneous ESI and APCI) Atmospheric pressure chemical ionization (APCI) Atmospheric pressure photoionization (APPI)

Detector
High energy conversion dynode and electron multiplier horn
Vacuum System
One Turbo-molecular pump with one mechanical pump
Parameter
Measure
Specification

Mass Range

10-2,000 m/z

SIM Sensitivity, ESI Positive
1pg of reserpine injected on column, quantifying the m/z 609 ion
Signal-to-noise ration, RMS >100:1
SIM Sensitivity, APCI Positive
1pg of reserpine injected on column, quantifying the m/z 609 ion
Signal-to-noise ratio, RMS >30:1
Maximum scan speed
u/sec (Da/sec)
10,400 u/sec (Da/sec)
Polarity Switching
Switching time from positive to negative polarity
300 msec
Mass resolution (Autotune)
Full width at half maximum
Unit resolution (< 0.7 u (Da))
Mass stability (drift)
Over 12 hours
< 0.1 u (Da) or 100 ppm, whichever is larger

Mass accuracy

+0.005 u (Da) (only applicable with optional Cerno MassWorks Accurate Mass Software for LC/MSD using OpenLab CDS ChemStation Edition

Minimum SIM dwell time

5 msec

Maximum SIM transitions

100 SIM ions per time segment

Dynamic Range

>6 x 106 (electronic)