NARDA 520 SYSTEM SPECS
NBM-520 BROADBAND FIELD METER
Result units mW/cm2, W/m2, V/m, A/m (for flat probes) % (for shaped probes) Display range 0.01 to 9999 V/m
0.0001 to 265.3 A/m
0.0001 to 9999 W/m2
0.0001 to 9999 mW/cm2
0.0001 to 9999 % Result types (RMS, isotropic) Actual (ACT), Maximum (MAX), Average (AVG), Spatial Averaging (SPATIAL) Averaging time 4 s to 30 min (2 s steps), selectable by PC software Spatial averaging discrete or continuously, selectable by PC software Alarm function 2 kHz audible signal (4 Hz repetition), threshold adjustable by PC software
HF-3061 H-FIELD PROBE
Frequency range (b) 300 kHz to 30 MHz Type of frequency response Flat Measurement range
0.012 to 16 A/m (CW)
0.012 to 0.7 A/m (True RMS)
5.4 μW/cm² to 10 W/cm² (CW)
5.4 μW/cm² to 18 mW/cm² (True RMS) Dynamic range 62 dB CW damage level 35 A/m 46 W/cm² Peak damage level (c) 350 A/m 4.6 kW/cm² Sensor type Diode based system Directivity Isotropic (Tri-axial)
HF-0191 H-FIELD PROBE
Frequency range (b) 27 MHz to 1 GHz Type of frequency response Flat Measurement range
0.018 to 16 A/m (CW)
0.018 to 1 A/m (True RMS) 12 μW/cm² to 10 W/cm² (CW) 12 μW/cm² to 38 mW/cm² (True RMS) Dynamic range 59 dB CW damage level 20 A/m 15 W/cm² Peak damage level (c) 200 A/m 1.5 kW/cm² Sensor type Diode based system Directivity Isotropic (Tri-axial) Readout mode / spatial assessment 3 separate axes
EF-9091 E-FIELD PROBE
Frequency range 100 MHz to 90 GHz Type of frequency response Flat Measurement range
0.7 to 400 V/m (CW)
0.7 to 61.4 V/m (True RMS) from 2 V/m for frequencies > 60 GHz 130 nW/cm² to 42 mW/cm² (CW) 130 nW/cm² to 1 mW/cm² (True RMS) from 1 μW/cm² for frequencies > 60 GHz Dynamic range 55 dB CW damage level 1600 V/m 700 mW/cm² Peak damage level (b) 1900 V/m 1 W/cm² Sensor type Diode based system Directivity Isotropic (Tri-axial) Readout mode / spatial assessment Combined 3-axis (RSS)
EF-5092 E-FIELD PROBE
Frequency range (b) 300 MHz to 50 GHz Type of frequency response Flat Measurement range 18 to 1090 V/m 0.085 to 315 mW/cm² Dynamic range 35 dB CW damage level 2400 V/m 1500 mW/cm² Peak damage level (c) 47 kV/m 600 W/cm² Sensor type Thermocouple (inherent True RMS detection) Directivity Isotropic (Tri-axial) Readout mode / spatial assessment Combined 3-axis (RSS)
EF-5091 E-FIELD PROBE
Frequency range (b) 300 MHz to 50 GHz Type of frequency response Flat Measurement range 8 to 614 V/m 17 μW/cm² to 100 mW/cm² Dynamic range 37 dB CW damage level 1500 V/m 0.6 W/cm² Peak damage level (c) 27 kV/m 200 W/cm² Sensor type Thermocouple (inherent True RMS detection) Directivity Isotropic (Tri-axial) Readout mode / spatial assessment Combined 3-axis (RSS)
EF-0691 E-FIELD PROBE
Frequency range (b) 100 kHz to 6 GHz Type of frequency response Flat Measurement range (c)
0.2 to 650 V/m (CW)
0.2 to 22 V/m (True RMS) 10 nW/cm² to 112 mW/cm² (CW) 10 nW/cm² to 0.13 mW/cm² (True RMS) Dynamic range 70 dB CW damage level 1000 V/m 265 mW/cm² Peak damage level (d) 10 kV/m 26 W/cm² Sensor type Diode based system Directivity Isotropic (Tri-axial) Readout mode / spatial assessment 3 separate axes
EB-5091 SHAPED E-FIELD PROBE
Frequency range (b) 3 MHz to 50 GHz (EB5091) Type of frequency response Shaped, see table on page 2 Measurement range 0.5 to 600 % of Standard (Power Density) 85 μW/Dynamic range 30 dB cm² to 500 mW/cm² CW damage level 2000 % of Standard 700 mW/cm² Peak damage level (c) 32 dB above Standard Sensor type 6C0o0m Wbin/cemd² diodes/ thermocouples Directivity Isotropic (Tri-axial) Readout mode / spatial assessment Combined 3-axis (RSS)