Product Specifications
Direct Radiography System Specifications
1) Digital Image Acquisition Parameter - X-ray Detector (Quantity 1) o Monolithic flat panel o Bit depth- 16 bit resolution o Image pixel height x width - 2048 x 2048 pixels o Digital zoom – 1 x or greater o Pixel pitch – 0.2 mm (200 micron) o Screen - Gadox Scintillator o Frame Rate - 15 frames per second at full resolution.
o Window width - Total Area: 16" x 16" (409.6mm x 409.6mm) o Interface – Fiber optic interface o Sensitivity - Ultrahigh o Output Grade: best quality – CT grade level
2) Workstation Specifications (Quantity 2) o The Image Acquisition and Processing system shall include two computers to acquire the radiograph, run the detector, process the radiographs, and analyze the data.
o The computers shall be a Dell or Hewlett-Packard as it will be updated, serviced, and maintained by the ATC Information Technology Division. No custom built computers will be accepted.
o The system software shall be able to be run in a Windows 10, 64-bit environment.
o The processor shall contain at least 6 cores.
o The workstations shall contain two hard drives o The workstations shall at least meet the minimum recommended UEFI secure boot and TPM 2.0 requirements.
o See below for minimum system requirements for DR Acquisition and Review workstation with monitor Operating System - Windows® 10 Operating System Processor - Eight-Core Intel® Xeon® Processor
CPU - 8 GB RAM
Storage - 1 TB SSD Operating System Drive Hard Drive - 3TB of Data Storage in RAID 0 Configuration DVD+/-RW Drive LAN - 10/100/1000 Network Interface Graphic Card - NVidia High Performance Display Adaptor Monitor - 30” (762 mm) 4 MP Color Display Input Device - Wireless Keyboard and Mouse
3) Software - Requirement
Digital Image Export Parameter o File Types: TIFF 32 bit / 16 bit / 8 bit, BMP, JPEG, RAW, DICONDE o Bit depth- 16 bit resolution o Image pixel height x width - 2048 x 2048 pixels
Image Acquisition Software and Review with Analysis tools. (Quantity 1 each) o Windows® based (Compatible with Windows 10) o DICONDE compliant o Tools to help ensure ASTM compliance o High performance image processing and measurement functions utilizing GPU technology o Automatic parallelization for systems with multiple CPU’s and GPU’s o Live Imaging up to 30 frames per second o Live Averaging, Live Histogram, Live line Profile, Live Rotation between portrait and landscape modes o Live Measurements with up to 1-pixel resolution o Live image gain/offset calibration and defective pixel correction o Automatic creation of Technique sheet o Capture video into AVI files o Read and store images in TIFF 32 bit / 16 bit / 8 bit, BMP, JPEG, RAW, DICONDE o Cropping support for reduced file size and versatility o Histogram o Line Profile o Rotation between portrait and landscape modes o Measurements
4) Installation o Installation by NDT Branch, Warfighter Directorate, ATC o Vendor shall provide all hardware, software and license keys.
o Vendor shall provide remote support if required.
5) Onsite Advanced Training (2 Days)
Direct Radiography Image Acquisition System training shall be at the Aberdeen Test Center, Aberdeen Proving Ground, Maryland.
Submitted
6) Imaging Software Acquisition and Analysis Compliance
Standards:
o American Society for Testing and Materials (ASTM) E2698, Radiological Examination
Using Digital Detector Arrays o American Society for Testing and Materials (ASTM) E747, Standard Practice for Design, Manufacture and Material Grouping Classification of Wire Image Quality Indicators (IQI) Used for Radiology
Image Acquisition System shall meet the following minimum requirements:
o Geometrical unsharpness calculated to be less than 0.5 mm.
o The Digital Detector Array shall have a pixel pitch less than or equal to 0.2 mm.
o The required quality level (2-2t) as determined by an IQI type (ASTM E747 Wire) shall be achievable for the inspection of hard body armor.
7) Warranty o Acquisition and Review Workstation shall be free from any defects and confirm to the Workstation Specifications (Section 5.0) for a period of 5 years.