Attachment_1_-_Product_Specifications.pdf
PDF 542 KB Posted
- Attached to
- HPLC 1260 Infinity II LC System
- Solicitation number
- W91ZLK-18-Q-0027
About this file
Product Specifications
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Attachment_2_-_Past_Performance.docx | DOCX document | |
| Attachment_3_-_J&A_Brand_Name_Only.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Attachment 1 - Specifications for an High Performance Liquid Chromatograph (HPLC)
Requirement: To acquire a state of the art Agilent HPLC instrument that provides maximum performance and versatility for an environment with ever changing requirements with the system configuration described below.
High Performance Liquid Chromatograph System – Minimum Requirements
1. The instrument must be a bench-top 1260 Infinity II LC system. The instrument must be a new model and currently in production. Refurbished or demonstrator instruments are not acceptable.
2. System must include:
1260 Infinity II Flexible pump with the following:
Quaternary pump with power range 800 bar and flow rate 5mL/min Includes active seal wash, solvent cabinet and bottles, ISET, and BlendAssist System Tool Kit Max uptime kit Stay Safe cap Poroshell 120 EC-C18 3.0x 100mm
1260 Infinity II vialsampler with the following:
100 uL metering devices 100uL sample loop Integrated needle flush port Standard drawer for 2mL vials Column identification kit Sample cooler
1260 Infinity II Multicolumn Thermostat column compartment with the following:
Valve drive Quick-connect HE large ID Quick-connect HE large ID HF 4 column selector with 0.12mm cap kit and 0.17mm cap kit
1260 Infinity II Diode Array Detector WR with the following:
120 Hz data acquisition rate Deuterium lamp Tungsten lamp Standard flow cell 10mm, 13uL, 120bar Programmable slit widths of 1, 2, 4, 8, and 16nm
OpenLab CDS ChemStation Edition PC Bundle including Software, core license, and PC Instrument driver with license 3D UV (PDA) add-on license
Instrument Controller
3. The controller must have a minimum of a 24” flat panel multimedia monitor and a printer.
4. The controller must have a Windows 10 system.
5. The controller must have a LAN connection to the instrument and include all necessary drivers.
Data System
6. The instrument controlling data system must be a minimum 64-bit running under Microsoft windows 10 format.
7. Must be designed according to the requirements of regulated labs and provide validation details.
8. Must be capable of using providing electronic notebook functionality to include electronic signatures and meet the requirements of 21 CFR Part 11.
9. Must automatically generate audit trails.
10. Must have built-in system suitability tests for the instrument and system including detector sensitivity, peak resolution, and noise.
11. Must be able to seamlessly transfer existing methods, data files, and/or spectral libraries from older LC instruments.
12. Must be able to view data collection live and easy to use live preview option.
13. Must be able to generate reports using template or by creating custom reports.
14. Must have ability to perform custom calculations.
15. Must be able to save spectra and a copy of the exact method used for their acquisition for traceability requirements in an accredited laboratory.
16. All raw data must be saved including calibration curves and the system must allow for post run reprocessing of any data point, eliminating the need to reanalyze the sample.
17. Must have a peak exploration tool to spot trends, retention time shifts, outliers, and artifacts to quickly diagnose instrument problems.
Support
18. Must include delivery and complete installation of the entire system to including software and transfer of methods.
19. Must include on-site familiarization training (for up to 4 people) for each component of the system and 3 consecutive days (8 hours per day) of on-site method and application consulting by a certified Agilent service engineer.
20. Must include a full 3-year extended warranty which includes all repairs (parts, labor, and travel). Must also include unlimited telephone technical help for hardware, software, applications, and consumables.
21. Telephone and/or online technical support for applications and method development from the manufacturer must be available for the entire useful life of the system.
Technical Assessment of Offeror Product
22. Technical assessment will consist of vendors providing peak resolution data to senior analyst prior to award.
23. Senior analyst will assess technical performance of all proposals in order to verify minimum specification requirements.
File details come from the government source that posted it.