W91YTZ22Q0128_Urinalysis Analyzer Questions from Contractors with Government Responses

Question 1. In the SON 1.1.7- the volume states 6,000 tests per year however, in table 1 annual volume shows 600 for most tests and 400 for Mucus, Casts, and Crystals, so seems to be conflicting volume. Can you confirm the annual volume for this requirement? Also, does Fort Lee do 100% micros? Typically, we see a smaller % of total volume with micros.

Government Response: the initial table values were incorrect and the correct annual test volumes for urine chemistries is 6,000 and urine microscopy is 4,000. The Statement of Need has been revised accordingly.

Analyte

Annual Volume

Color 6000 Clarity 6000 Glucose 6000 Bilirubin 6000 Ketone 6000 Specific Gravity 6000 pH 6000 Blood 6000 Protein 6000 Urobilinigen 6000 Nitrite 6000 Leukocyte 6000

WBC 6000

RBC 6000

Bacteria 6000 Squamous Epithelial Cells 6000 Mucus 4000 Casts 4000 Crystals 4000

Question 2. Can you clarify that the 5 page response is referring to only the response and doesn't include the 4 pages of the SOW? So SOW response would be no more than 9 pages??

Government Response: Ignore the 5 Page limit for technical response requirement stated in the ADDENDUM to 52.212-1. If you are embedding responses and including the SOW, please limit your page length to 15 pages.