Or_Equal_for_Bid.xlsx
XLSX spreadsheet 10 KB Posted
- Attached to
- SPECTROSCOPY SYSTEM FOR KEESLER AFB
- Solicitation number
- W91YTZ-13-T-0179
- Issued by
-
Department of the Army Medical Command
About this file
OR EQUAL
View the file
Other files
| File | Type | Posted |
|---|---|---|
| A11.doc | DOC document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Sheet2
| Salient Characteristics | Company Name: | |
| Please Fill in Equal with description | ||
| 1 | Solid-state white light illumination | |
| 2 | Fully-automated focus, stage movement and illumination | |
| 3 | Detector design containing a single element MCT detector and a linear array MCT imaging | |
| 4 | Must operate in mid-IR, near-IR or dual range to cover the full range of imaging applications in one instrument | |
| 5 | Collects up to 170 scans sec | |
| 6 | Frequency range: 8300-350 cm 1 using a Ge-coated KBr beam | |
| 7 | Resolution: better than 0.4cm-1 | |
| 8 | Better than 55,000:1 RMS signal-to-noise (4cm-1 spectrall -resolution, 5 seconds, transmission measurement), better than 190,000:1 RMS (1 minute, 4cm-1 spectral resolution, transmission measurement) | |
| 9 | Mode change with no mechanical changeover or realignment necessary | |
| 10 | Variable pixel resoultion, 1.5micro (ATR Imagining mode only), 6.25micro, 25miceo, and 50micro | |
| 11 | Array frequency range: 7800-720cm-1 |