Salient Characteristics (Redacted).pdf
PDF 313 KB Posted
- Solicitation number
- W911QX25QA091
About this file
This document is a Salient Characteristics template for a federal contract solicitation for two New Brunswick Innova® 42R Stackable Incubator Shakers. The detailed specifications require a laboratory incubator shaker with advanced features including 24/7 continuous operation, stackable configuration, triple-eccentric counterbalanced drive, selectable orbit sizes, compatibility with multiple platforms and accessories, and integration with VisioNize® Lab Suite for remote monitoring and data logging. Additional requirements include temperature range of 20°C below ambient to 80°C, speed range of 25-400 RPM, front door opening, a platform accommodating fifteen programming steps, and precise dimensional and weight specifications. The document is signed by a Contracting Officer with a signature date of 8/13/2025, indicating this is a current procurement document for laboratory equipment to be installed at the Adelphi Laboratory Center (ALC).
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Template last updated 07/28/2016
Salient Characteristics Revisions marked in yellow
Commercial Availability Certification:
I have evaluated the commercial marketplace and determined that all of the salient characteristics documented below are available in products/services offered in the commercial marketplace.
Signature (typed name): _ ________________________
Technical Point of Contact (TPOC) or Contracting Officer’s Representative (COR)
The Contractor shall provide quantity two (2) New Brunswick Innova® 42R Stackable Incubator Shaker, or equivalent, with the following minimum specifications:
1. Shall have reliable and vibration-free operation for continuous (24/7) use.
2. Shall be designed for stackable configuration to maximize valuable laboratory space utilization (to include a stacking kit (catalog no. M1335-0800), universal platform (catalog no. 2231001157), and professional installation at the Adelphi Laboratory Center (ALC).
3. Shall have a triple-eccentric counterbalanced drive.
4. Shall offer selectable orbit sizes of 2.5 cm (1 inch) or 5.1 cm (2 inches)
5. Shall be compatible with both Universal Platforms, allowing for flexible combinations of accessories, and Dedicated Platforms for high-capacity, single-vessel-type applications.
6. Shall be fully compatible with a wide range of accessories including, but not limited to, flask clamps, sticky pads, microplate racks, and test tube racks.
7. Shall integrate seamlessly with the VisioNize® Lab Suite for remote monitoring of critical parameters including temperature, speed, and program runtime.
8. Shall allow users to define critical values (e.g., door-opening time, orbital speed) via VisioNize® to trigger on-screen, audible, SMS, and/or email alerts.
9. Shall provide a comprehensive audit trail of shaker parameters through data logging capabilities within the VisioNize® Lab Suite, supporting reproducibility and documentation.
10. Shall be constructed of high-quality materials and precision fabrication, ensuring decades of reliable and worry-free operation.
11. Shall accommodate a platfore size of 46 × 46 centimeters (cm) (18 × 18 inches)
12. Shall provide a maximum vessel volume (without static shelf, depending on flask format) of 6 L
13. Shall be compatible with alge and plants cells through the use of fluorescent bulbs, with or without intensity modulation.
14. Shall accommodate parallel static cultivation with at least one static shelve and a culture drawer for Agar plates.
15. Shall maintain a temperature range of 20 Celsius (°C) below ambient (min. 4 °C) to 80 °C (max. 70 °C with photosynthetic light bank, 60 °C with relative humidity (rH) monitoring).
File details come from the government source that posted it. Updated .