The document is a Salient Characteristics Certification for a small business set-aside solicitation (W911QX25QA051) issued by the Department of the Army Contracting Command at Aberdeen Proving Ground. The solicitation requires the procurement of one (1) Air-lift Bench-top Bioreactor System with detailed technical specifications, including a 10-liter working volume, temperature control from 5 degrees above ambient to 60 degrees Celsius, and peristaltic pumps for programmable acid/base addition and defoaming.
Key system requirements include a benchtop design weighing less than 80 kilograms, NRTL/UL certification for U.S. electrical systems (120V or 240V), air condenser, thermally controlled jacketed vessel, mixer capabilities of 0-1000 rpm, dissolved oxygen and pH controls, and air inflow volume of 0.5-10 standard liters per minute. The bioreactor system must be delivered FOB Destination to the Army Research Laboratory Center in Adelphi, Maryland, with the solicitation representing a 100% small business set-aside opportunity.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Combination Synopsis-Solicitation for Commercial Purchases (Under SAT)_Final.docx | DOCX document | |
| Adelphi Special Instructions and Information Full-Text Attachment.doc | DOC document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Salient Characteristics
Commercial Availability Certification:
I have evaluated the commercial marketplace and determined that all of the salient characteristics documented below are available in products/services offered in the commercial marketplace.
Signature (typed name): TBA after Contract Award_____________ Technical Point of Contact (TPOC) or Contracting Officer’s Representative (COR)
The Contractor shall provide one (1) Air-lift Bench-top Bioreactor System with the following specifications:
a) Shall be a benchtop system.
b) Shall be sterilizable in an autoclave with dimensions of 25 inches long x 25 inches wide x 40 inches deep.
c) Shall have peristaltic pumps: programmable for acid addition, base addition, defoaming, and/or feeding.
d) Shall have built-in pressure regulator.
e) Shall have multi-functional advanced dissolved oxygen (D.O.) and potential of hydrogen (pH) control.
f) Shall have temperature control with a range of five (5) degrees Celsius (°C) above ambient to 60 °C.
g) Shall include a mixer with speed of zero (0) to 1000 revolutions per minute (rpm).
h) Shall have controllable air inflow volume of 0.5-10 standard liter per minute (slpm).
i) Shall have a working volume of at least ten (10) Liters (L).
j) Shall have a thermally controlled jacketed vessel.
k) Shall include one (1) sampling tube from top into bottom.
l) Shall be equipped with air condenser.
m) Shall weigh less than 80 kilograms (kg).
n) Shall be Nationally Recognized Testing Laboratory (NRTL) or Underwriters Laboratories (UL) certified for use in the U.S.A. with 120 volts (V) or 240 V single phase electrical systems.
o) The Contractor shall deliver the one (1) airlift bioreactor-benchtop FOB Destination to The Army Research Laboratory Center, Adelphi Maryland 20783.