This document is a government solicitation for a Request for Quote (RFQ) for one (1) High-Temperature Tube Furnace. The Army Research Laboratory (ARL) requires a benchtop system capable of holding temperature continuously between 800-1500°C in a heating zone of at least 250mm, with a maximum temperature of 1600°C or above. The furnace must meet specific technical requirements for temperature control, tube size, heating and cooling rates, inert gas flow, and certification. This is a small business set-aside solicitation, with the opportunity converting to full and open competition if no acceptable small business offers are received. Offers are due within 5 business days, and delivery is required within 30-60 days of contract award. Award will be made to the offer most advantageous to the government, based on price, technical capabilities, and past performance.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Provisions and Clauses- Tube Furnace.pdf | ||
| Solicitation - Combination Synopsis - Tube Furnace.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Salient Characteristics
Commercial Availability Certification:
I have evaluated the commercial marketplace and determined that all of the salient characteristics documented below are available in products/services offered in the commercial marketplace.
Signature (typed name):
Technical Point of Contact (TPOC)
The Army Research Laboratory (ARL) requires one (1) High-temperature Tube Furnace. The Contractor shall provide one (1) High-temperature Tube Furnace that meets the following characteristics:
• Shall be a benchtop system.
• Shall be able to hold a tube horizontally and include tube supports if necessary.
• Shall be able to hold temperature continuously between 800-1500 ºC in a heating zone of at least 250mm. The maximum temperature shall be 1600 ºC or above.
• Shall be able to control temperature with precision +/- 10 ºC.
• Shall be able to accommodate tube sizes with an outer diameter of at least
100mm.
• Shall be able to control heating and cooling rate at least 10 ºC/min below 1100ºC and at least 5 ºC/min above 1100ºC
• Shall be able to support the use of inert gas flow over the sample and include necessary flanges to do so.
• Shall be able to program multiple heating, cooling, and holding segments on a temperature controller.
• Shall be Nationally Recognized Testing Laboratory (NRTL) or Underwriters
Laboratories (UL) certified for use in the U.S.A. with 120V or 240V single phase electrical systems.
The Contractor shall deliver one (1) High-temperature Tube Furnace FOB Destination to The Army Research Laboratory Center, Adelphi Maryland 20783.
Contracting Officer: