Salient Characteristics V2 (Data Pack).doc
DOC document 79 KB Posted
- Attached to
- One (1) Benchtop Dynamic Light Scattering instrument Federal contract opportunity
- Solicitation number
- W911QX23Q0176
About this file
This document outlines the technical requirements for one benchtop dynamic light scattering instrument being solicited by the Department of the Army Materiel Command Army Contracting Command Aberdeen Proving Ground. The solicitation seeks an instrument capable of determining particle sizes from 0.3 nanometers to 10 micrometers using both back-scatter dynamic light scattering and multiple measurement angles. The instrument must be able to conduct measurements of particle size, zeta potential, molecular weight, particle concentration, and perform temperature-controlled experiments from 1 to 90 degrees Celsius on inorganic, organic, and biological particles including proteins and fluorescent samples. Accompanying software is required for automated data analysis and output of measurement results.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Combination Synopsis-Solicitation for Commercial Purchases (Under SAT) (Amendment 00001) (Data Pack).docx | DOCX document | |
| Salient Characteristics (Data Pack).doc | DOC document | |
| Evaluation Criteria (Comparative) (Stylus Pro).docx | DOCX document | |
| Provisions and Clauses Full-Text Attachment (for Combo Synopsis-Solicitation) (Data Pack).doc | DOC document | |
| Combination Synopsis-Solicitation for Commercial Purchases (Under SAT) (Data Pack).docx | DOCX document |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Salient Characteristics
Commercial Availability Certification:
I have evaluated the commercial marketplace and determined that all of the salient characteristics documented below are available in products/services offered in the commercial marketplace.
Signature (typed name): __Alexander J. Winton__________________
Technical Point of Contact (TPOC) or Contracting Officer’s Representative (COR)
The contractor shall provide one (1) Benchtop Dynalmic Light Scttering instrument meeting the following characteristics:
· Capable of determining particle sizes in the range of 0.3 nanometer to 1 0 micrometer. Provides autocorrelation function and displays particle size, standard deviation, mean and mode as well as polydispersity index, refractive index and particle distribution graph. Able to conduct particle size measurement using back-scatter dynamic light scattering.
· Able to conduct zeta potential measurements in the range of -200 mV to +200 mV.
· Able to perform particle size measurements that are not restricted to a single measurement angle (instrument capable of utilizing more than a single angle when making a particle size measurement, though dual angle measurements not required to be conducted simultaneously).
· Able to conduct measurements on inorganic, organic, and biological particles & proteins.
· Capable of performing particle concentration analysis.
· Able to conduct Particle Size measurements on fluorescent particle samples.
· Accompanying software that provides data analysis and output for particle size, zeta potential, molecular weight.
· Software able to conduct automated temperature ramping/cooling measurements.
· Adjustable cell holder temperature between 1o Celsius and 90o Celsius.
· Capable of conducting Molecular weight analysis in the range of 1000 to 20,000 Daltons based on measurements.
· Provide computer and software for operating the particle analyzer and providing data analysis & read-outs.
Contracting Officer: