Salient Characteristics
The Aberdeen Proving Grounds (APG) Cermanic and Transparent Materials Branch (CTMB) requires the procurement of a laboratory scale bench top hot press furnace with the following minimum requirements:
· At least 2500°C maximum temperature
· At least 2300°C normal operating temperature
· All graphite heating zone (graphite heating elements and graphite insulation)
· Greater than or equal to 25°C/minute furnace ramp up
· Usable work zone to be at least 3” diameter x 4” height (maximum die dimensions)
· Less than or equal to 50°C maximum exterior chamber temperature
· 15 ton (30,000 lbs.) automated press with a 200 lb. or less minimum set point
· Turbo vacuum with an ultimate vacuum equal to or lower than 5.0 X10-6 Torr in a dry chamber without sample at room temperature
· Gas purification (Gettering furnace with oxygen analyzer) capable of purifying Argon and Nitrogen with titanium or iron getter charge (respectively)
· Automatic retractable thermocouple above 1000°C rated for high vacuum (≥ 10-5 Torr) and Argon atmosphere
· Optical pyrometer for >1000°C
· Electronic data display and acquisition of ram position and load indicator with touch screen display
· Ram position indicator capable of ram displacement measurements with a repeatability of 0.0001” or better
· Mass spectrometer (RGA) with high pressure analyzer for chemical species analysis of evolved gases during furnace operation with data acquisition and associated software for mass versus time
· Power supply of 240 VAC, 1 PH, 60 Hz
· Delivery within 6 months (26 weeks)
· On-site (ARL APG) unpacking, set up and training
· Standard Spare parts kit to include:
· Graphite heating element
· graphite insulation kit
· graphite rods and pyrolytic spacers
· extra thermal couples
· O-rings, fuses
· titanium getter charge
· Iron gettering charge with regeneration kit.