SIGNAL ANALYZER

PURCHASE DESCRIPTION

1. GOVERNMENT NEEDS: A signal analyzer as described in this purchase description.

2. SPECIFICATIONS: The following specifications detail the basic characteristics of the signal analyzer:

2.1 GENERAL SPECIFICATONS

2.1.1. Power Requirements: Voltage and frequency 100 to 120 V, 50/60/400 Hz

2.1.2. Display

- Resolution: 1280 x 768 minimum, WXGA

- Size: 10” diagonal minimum

2.1.3. Data Storage: 160 GB nominal (removable solid state drive)

2.1.4. Dimensions; 7” H x 16” W x 14” L minimum

2.2 FREQUENCY AND TIME SPECIFICATIONS

2.2.1. Frequency Range: DC coupled: 9 kHz to 26.5 GHz; AC coupled: 10 MHz to 26.5 GHz

2.2.2. Aging Rate: ± 1 x 10–6 / year

2.2.3.: Temperature Stability:

20 to 30 °C

± 2 x 10–6

Full Temperature Range ± 2 x 10–6

2.2.4. Achievable Initial Calibration Accuracy: ± 1.4 x 10–6

2.2.5 Trigger: Free run, line, video, external 1, RF burst, periodic timer

2.2.6 Resolution Bandwidth

Range (–3.01 dB bandwidth)
1 Hz to 3 MHz (10 % steps), 4, 5, 6, 8 MHz
Bandwidth Accuracy (power)
1 Hz to 750 kHz
± 1.0 % (± 0.044 dB) nominal
820 kHz to 1.2 MHz (< 3 GHz CF)
± 2.0 % (± 0.088 dB) nominal
1.3 to 2.0 MHz (< 3 GHz CF)
± 0.07 dB nominal
2.2 to 3 MHz (< 3 GHz CF)
± 0.15 dB nominal
4 to 8 MHz (< 3 GHz CF)
± 0.25 dB nominal

2.2.7. Analysis Bandwidth: Maximum bandwidth – 10 MHz

2.2.8. Video Bandwidth Range - 1 Hz to 3 MHz (10 % steps), 4, 5, 6, 8 MHz, and wide open (labeled 50 MHz)

2.3 AMPLITUDE ACCURACY AND RANGE SPECIFICATIONS

2.3.1. Amplitude Range – Preamp on/off – 100 kHz to 26.5 GHz

2.3.2. Input Attenuator Range – 0 – 70 dB in 10 dB steps

2.3.3. Maximum Safe Input Level

+30 dBm (1 W)
Input attenuation ≥ 10 dB, preamp off
+30 dBm (1 W)
Input attenuation ≥ 20 dB, preamp on

2.3.4. DC Volts: AC coupled - ± 50 Vdc; DC coupled - ± .2 Vdc

2.3.5 Frequency Response

(10 dB input attenuation, 20 to 30 °C, s = nominal standard deviation)

Specification

95th percentile (≈ 2s)

9 kHz to 10 MHz
± 0.8 dB
± 0.5 dB
10 MHz to 3 GHz
± 0.65 dB
± 0.4 dB
3 to 7.5 GHz
± 1.5 dB
± 0.5 dB
7.5 to 13.6 GHz
± 2.0 dB
± 0.8 dB
13.6 to 19 GHz
± 2.0 dB
± 1.0 dB
19 to 26.5 GHz
± 2.5 dB
± 1.3 dB

Preamp on

100 kHz to 3 GHz
± 0.7 dB
3 to 13.6 GHz
± 1.0 dB
13.6 to 19 GHz
± 1.1 dB
19 to 26.5 GHz
± 2.5 dB

2.3.6. Input Attenuation Switching Uncertainty

Specifications Additional Information Attenuation > 2 dB, 50 MHz (ref freq) ± 0.32 dB ± 0.15 dB typical preamp off Relative to 10 dB 0 kHz to 3.0 GHz ± 0.30 dB nominal (reference setting) 3.0 to 7.5 GHz

± 0.50 dB nominal

7.5 to 26.5 GHz ± 0.70 dB nominal

2.3.7 Input voltage standing wave ratio (VSWR) (≥ 10 dB attenuation) 10 MHz to 3 GHz

< 1.3 nominal

3 to 7.5 GHz

< 1.4 nominal

7.5 to 26.5 GHz

< 1.9 nominal

2.4 DYNAMIC RANGE SPECIFICATIONS

2.4.1. 1 dB gain compression (two-tone)

Total power at input mixer Preamp off

50 MHz to 7.5 GHz

+7 dBm noiminal

7.5 to 13.6 GHz

+3 dBm noiminal

13.6 to 26.5 GHz

+0 dBm noiminal

Preamp on

50 MHz to 26.5 GHz –19 dBm nominal

2.4.2. Displayed average noise level (DANL) - Input terminated, sample or average detector, averaging type = Log, 0 dB input attenuation, IF Gain = High, 20 to 30 °C

Preamplifier Off Preamplifier On

1 to 10 MHz
–143 (–148) dBm
–153 (–158) dBm
10 MHz to 1.5 GHz
–147 (–150) dBm
–160 (–163) dBm
1.5 to 6 GHz
–143 (–147) dBm
–158 (–161) dBm
6 to 7.5 GHz
–141 (–145) dBm
–155 (–160) dBm
7.5 to 13.6 GHz
–139 (–142) dBm
–155 (–160) dBm
13.6 to 20 GHz
–134 (–140) dBm
–153 (–157) dBm
20 to 24 GHz
–132 (–138) dBm
–151 (–155) dBm
24 to 26.5 GHz
–124 (–129) dBm
–142 (–147) dBm

2.4.3 Third-order intermodulation distortion (TOI) Preamp off (Two –20 dBm tones at input mixer spaced by100 kHz, 0 dB attenuation, 20 to 30 °C)

10 to 500 MHz
+11 dBm, (+15) dBm
500 MHz to 2 GHz
+12 dBm, (+15) dBm
2 to 3 GHz
+11 dBm, (+15) dBm
3 to 7.5 GHz
+12 dBm, (+17) dBm
7.5 to 13.6 GHz
+11 dBm, (+15) dBm
13.6 to 26.5 GHz
+10 dBm, (+14) dB

2.5 POWERSUITE MEASUREMENT SPECIFICATIONS

Dynamic Range Typical

Adjacent Alternate

Without noise correction

-66dB

-69dB

With noise correction

-73dB

-78dB

2.6 INPUTS AND OUTPUTS

2.6.1 Front panel

RF input

Connector

Type-N female, 50 Ω nominal

2.6.2 Rear Panel

10 MHz out

Connector

Output amplitude

Frequency BNC female, 50 Ω nominal

≥ 0 dBm nominal

10 MHz ± (10 MHz x frequency reference accuracy)

Ext Ref In

Connector

Input amplitude range

Input frequency

Frequency lock range BNC female, 50 Ω nominal

–5 to 10 dBm nominal

10 MHz ± nominal

± 5 x 10–6 of specified external reference input frequency

Trigger 1 input

Connector

Impedance

Trigger level range BNC female

> 10 kΩ nominal

–5 to 5

Trigger 1 output

Connector

Impedance

Level BNC female

50 Ω nominal

5 V TTL nominal

USB ports compatible with USB 3.0

2.7 I/Q ANALYZER

2.7.1. Frequency span – 10 Hz to 10 MHz

2.7.2 Resolution Bandwidth

Range

Overall
100 mHz to 3 MHz
Span = 1 MHz
50 Hz to 1 MHz
Span = 10 kHz
1 Hz to 10 kHz
Span = 100 Hz
100 mHz to 100 Hz

2.7.3 Data acquisition (standard 10 MHz IF path)

Time record length
4,000,000 IQ sample pairs
Sample rate
30 MSa/s
ADC resolution
14 Bits

3. CALIBRATION - An ANSI Z540-1-1994 calibration/certificate shall be provided with the unit.

4. INFORMATION ASSURANCE REQUIREMENTS: This system will operate in stand-alone mode only (no network connection of any kind). If there becomes such a time where this system requires the use of another PC (laptop, tablet, etc.) or network connection it will conform to all requisite DoD and Army cybersecurity requirements. Additionally, since this system operates in stand-alone mode there will be no ability to connect via either wireless

(e.g Bluetooth) or wired (e.g. Ethernet cable), to a network. If these parameters change then the system will be subject to Change Configuration Board (CCB) approval for operation.

5. WARRANTY: A one year minimum warranty shall be furnished with the provided equipment. The warranty shall cover all components. The warranty period begins when the LEAD Contracting Officer provides official confirmation that the equipment has satisfied all operational provisions of the contract.

6. PRESERVATION, PACKAGING AND DELIVERY: The contractor shall utilize standard commercial methods for preservation and packaging appropriate for the provided equipment and acceptable to commercial carriers. As a minimum, all areas susceptible to damage from exposure to the elements shall be preserved and/or packed to prevent damage. The contractor shall be responsible for ensuring the equipment is delivered to LEAD in good condition.

7. INSPECTION AND ACCEPTANCE: The provided equipment shall be examined for design, construction, materials, components, and workmanship to determine compliance with the contract and this specification. Any deviation from these requirements shall be cause for rejection.

File details come from the government source that posted it. Updated .