Minimum Essential Characteristics for Two (2) Each Immunoassay Systems:

1. Shall be a System Requirements: An instrument that is a laboratory workstation with a capability to simultaneously measure multiple agents (toxins, viruses and bacteria) from a single sample

2. Shall be electrochemiluminescence (ECL) signal detection technology required

3. Shall be plate based immunoassay system capable of 96-well 9, 16, and 25 spot plates

4. Shall have a detection electronics - Cooled CCD Camera

5. Shall be plate Read Time - 7 minutes

6. Shall be dynamic assay range of 10^5 over the spectrum of assays

7. Shall be temperature operational range 15-30°C minimum

8. Shall have system foot print of less than 18 x 18 inches (L x W)

9. Shall have voltage: 120 VAC ± 10%

10. Shall be two copies of operators, service/installation literature required.