This document outlines detailed technical requirements for a vacuum furnace, specifying precise specifications for scientific laboratory equipment. The furnace must operate on 208V or 220V power, have an interior capable of accommodating a 12-inch cubic microfluidics board, and reach a maximum temperature of 1200 Celsius. Additional critical requirements include a digital vacuum gauge, compact dimensions of 30 × 30 × 36 inches or less, capability for both inert gas and vacuum operation, and the ability to reach vacuum levels of 5E-4 torr or lower. The system must also include a compatible vacuum pump with a minimum pumping speed of 300L/min, ensuring comprehensive functionality for advanced research and manufacturing processes involving high-temperature, low-pressure environments.

View the file

Other files

File Type Posted
N00173-26-R-GF03 CSS.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Vacuum Furnace Requirements:

• 208V or 220V power to interface with existing electrical connections

• Interior capable of holding a 12 inch x 12 inch x 12 inch assembly to accommodate largest considered microfluidics board.

• Max Temp 1200 Celsius (1473K) or greater.

• Digital Vacuum gauge included

• Dimensions: 30 inch × 30 inch × 36 inch or less to fit in existing lab space

• Capability for both inert gas and vacuum operation.

• In Vacuum mode, capable of reaching 5E-4 torr or lower.

• A compatible vacuum pump, with pumping speed of 300L/min or greater.