This document is a Technical Evaluation Sheet for a 1.5 Tesla (T) Mobile Magnetic Resonance Imaging (MRI) System solicitation (36C25624Q0713) issued by the Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 16 (VISN 16). The solicitation requires the MRI system to meet specific technical requirements, including a 1.5T magnet, 2D FSE-based imaging, wide bore, low gradient noise, digital imaging capabilities, and various application packages. The MRI system must also include specific RF coils and additional requirements such as a CD burner, PACS compatibility, patient safety features, and operator/technical support. The solicitation is a firm-fixed price requirements contract with a NAICS code of 621512 (standard size $19M). Responses are due by July 19, 2024, at 2:00 PM CST, and all questions must be submitted by July 16, 2024, at 10:00 AM CST.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| 36C25624Q0713 0001_1.docx | DOCX document | |
| 36C25624Q0713.docx | DOCX document | |
| P01. Site Plan C501 Utilities.pdf | ||
| P01. Site Plan C202 Central Site Plan.pdf | ||
| 36C25624Q0713_1.docx | DOCX document |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Technical 36C25624Q0713
Review for fully function/fully operable 1.5 Tesla (T) Mobile Magnetic Resonance
Imaging (MRI) System
COMPLY NOT COMPLY
TECHNICAL EVALUATION EXPLANATION HOW IT COMPLIES OR DOES NOT
COMPLIES WITH THE REQUIRMENT
TECHNICAL EVALUATOR NAME: ___________________________________
1.5 Telsa Mobile MRI System must meet the following minimum requirements
State of the art 1.5T Magnet
Must have 2D FSE-base magnetic resonance image compilation imaging
The MRI must have a wide bore for patients that are claustrophobic and/or obese
Gradient noise level of 99 db or below without reducing performance
Magnetic Resonance Image Compilation (generate multiple image contrasts in a single MRI scan)
The MRI must have Total Digital Imaging (TDI). TDI helps deliver consistently sharper images than conventional MR systems with higher SNR, superior homogeneity and uniformity Equipment will be capable of sending images to a minimum of 3 destinations (printer, PAC’s, Hospital RIS System, etc)
MRI System will include these application packages as a minimum
Synthtic MRI Capabilities
Fat Suppression Technology
MAGIC Suite
3d ASL, HEART, and Promo
RF Coil Suite
Silent Suite
Neuro Suite
Angio Suite
Cardiac Suite
Body Suite
Breast Suite
Viso Works
Ortho Suite
Pediatric
SPACE
TWIST
BLADE
Diffusion
SWI
Argus Flow Quantification
These Coils must be included as a minimum
Head Coil
Neck Coil
Body Coil
Spine Coil
Angio Coil
4 channel shoulder array, small and large coils
Large and Small CP Flex coils
TMJ Coil
8 channel foot/ankle/knee coil
8 channel wrist coil
Additional MRI Trailer and System Requirements
CD Burner
Full compatibility with hospital PACS System
DICOM 3.0
Patient intercom system
Safety and Warning Systems Visual and Audible Alarms
Smoke Detectors, Fire Extinguisher
Emergency Lighting Battery
Phone and data ports
480v 50' power cable
Hydraulic patient lift, stairs and railing
HVAC System to maintain trailer environment systems
Operators and Technical support Manuals
Fully System Service (Monday-Friday after normal duty hours local time excluding Federal
Holidays
Regular scheduled maintenance to be performed after hours
24 Hour telephone technical support toll free line, response within 1 hour
550 LBs maximum patient table weight
File details come from the government source that posted it. Updated .