This document is a Technical Evaluation Sheet for a 1.5 Tesla (T) Mobile Magnetic Resonance Imaging (MRI) System solicitation (36C25624Q0713) issued by the Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 16 (VISN 16). The solicitation requires the MRI system to meet specific technical requirements, including a 1.5T magnet, 2D FSE-based imaging, wide bore, low gradient noise, digital imaging capabilities, and various application packages. The MRI system must also include specific RF coils and additional requirements such as a CD burner, PACS compatibility, patient safety features, and operator/technical support. The solicitation is a firm-fixed price requirements contract with a NAICS code of 621512 (standard size $19M). Responses are due by July 19, 2024, at 2:00 PM CST, and all questions must be submitted by July 16, 2024, at 10:00 AM CST.

View the file

Other files for this federal contract opportunity

Other files attached to W065--Mobile MRI in Springfield MO, newest first.
File Type Posted
36C25624Q0713 0001_1.docx DOCX document
36C25624Q0713.docx DOCX document
P01. Site Plan C501 Utilities.pdf PDF
P01. Site Plan C202 Central Site Plan.pdf PDF
36C25624Q0713_1.docx DOCX document

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Technical 36C25624Q0713

Review for fully function/fully operable 1.5 Tesla (T) Mobile Magnetic Resonance

Imaging (MRI) System

COMPLY NOT COMPLY

TECHNICAL EVALUATION EXPLANATION HOW IT COMPLIES OR DOES NOT

COMPLIES WITH THE REQUIRMENT

TECHNICAL EVALUATOR NAME: ___________________________________

1.5 Telsa Mobile MRI System must meet the following minimum requirements

State of the art 1.5T Magnet

Must have 2D FSE-base magnetic resonance image compilation imaging

The MRI must have a wide bore for patients that are claustrophobic and/or obese

Gradient noise level of 99 db or below without reducing performance

Magnetic Resonance Image Compilation (generate multiple image contrasts in a single MRI scan)

The MRI must have Total Digital Imaging (TDI). TDI helps deliver consistently sharper images than conventional MR systems with higher SNR, superior homogeneity and uniformity Equipment will be capable of sending images to a minimum of 3 destinations (printer, PAC’s, Hospital RIS System, etc)

MRI System will include these application packages as a minimum

Synthtic MRI Capabilities

Fat Suppression Technology

MAGIC Suite

3d ASL, HEART, and Promo

RF Coil Suite

Silent Suite

Neuro Suite

Angio Suite

Cardiac Suite

Body Suite

Breast Suite

Viso Works

Ortho Suite

Pediatric

SPACE

TWIST

BLADE

Diffusion

SWI

Argus Flow Quantification

These Coils must be included as a minimum

Head Coil

Neck Coil

Body Coil

Spine Coil

Angio Coil

4 channel shoulder array, small and large coils

Large and Small CP Flex coils

TMJ Coil

8 channel foot/ankle/knee coil

8 channel wrist coil

Additional MRI Trailer and System Requirements

CD Burner

Full compatibility with hospital PACS System

DICOM 3.0

Patient intercom system

Safety and Warning Systems Visual and Audible Alarms

Smoke Detectors, Fire Extinguisher

Emergency Lighting Battery

Phone and data ports

480v 50' power cable

Hydraulic patient lift, stairs and railing

HVAC System to maintain trailer environment systems

Operators and Technical support Manuals

Fully System Service (Monday-Friday after normal duty hours local time excluding Federal

Holidays

Regular scheduled maintenance to be performed after hours

24 Hour telephone technical support toll free line, response within 1 hour

550 LBs maximum patient table weight