Tab 04 SOW.pdf

PDF 103 KB Posted

Attached to
Hamamatsu Micro x-ray source for defect detection in metallic AM structures
Solicitation number
80NSSC26942207Q-1
Issued by

About this file

This is a Statement of Work for the procurement of a micro x-ray source designed for non-destructive testing of thin-walled, additively-manufactured metallic structures. The contractor shall manufacture or supply one micro x-ray source unit that meets specific technical performance criteria, including voltage range of 40-130kV, current of 10-300 µA, maximum power of 39W, focal spot size not exceeding 10 micrometers, cone angle of 45-53 degrees, and focus-to-object distance of 7-15 mm. The source design must be as compact as possible to facilitate integration into restricted spaces.

The contractor is responsible for two primary tasks: manufacturing or supplying the micro x-ray source meeting all specifications and performing factory acceptance testing to verify performance metrics prior to shipment. Deliverables include the hardware unit, a compatible power supply with all necessary cables and computer communication cables, control software with operating license, technical documentation including operation manuals and safety guidelines, safety standards certification demonstrating compliance with applicable IEC/EN and FCC standards, and test data including a Certificate of Conformance with final verification of voltage, power, and spot size. Key milestones include order acknowledgement within one week of receipt, pre-shipment verification and testing, and final delivery and acceptance at destination. No government-furnished property or information will be provided, as this is an off-the-shelf procurement.

View the file

Other files

File Type Posted
Tab 10 RFQ 80NSSC26942207Q-1.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

STATEMENT OF WORK

1.0 Introduction/Background

The purpose of this effort is to procure a micro x-ray source to be used for imaging defects inside thin-walled, additively-manufactured, metallic structures. High-resolution x-ray capability is required to ensure the structural integrity and quality assurance of these metallic components. This procurement addresses the need for an x-ray source to support non-destructive testing requirements.

2.0 Scope of Work

This Statement of Work covers the procurement, fabrication (if applicable), and delivery of one (1) micro x-ray source. The contractor shall ensure the hardware meets the specific size, power, and resolution performance metrics defined herein.

3.0 Objectives

The objective of this requirement is to acquire a micro x-ray source that combines high voltage and power with a small spot size to enable precise defect detection. The successful delivery of this hardware will allow for the detailed inspection of thin-walled metallic parts.

4.0 Description of the Work/Contractor Tasks

The Contractor shall furnish all necessary labor, equipment, and materials to deliver a micro x-ray source meeting the technical specifications listed below.

4.1 Technical Specifications

The micro x-ray source must meet or exceed the following performance criteria:

Voltage: range 40 - 130kV Current: 10 – 300 µA Max Power: 39W Focal Spot Size: Maximum 10um Cone Angle: 45 - 53 degrees Focus to Object Distance: 7 – 15 mm Form Factor: The source design must be as compact as possible to facilitate integration into restricted spaces.

4.2 Tasks

Task 1: Manufacture/Supply one (1) micro x-ray source meeting the specifications in Section 4.1.

Task 2: Perform factory acceptance testing to verify performance metrics (kV, Wattage, Spot Size) prior to shipment.

5.0 Deliverables

The Contractor shall provide the following deliverables to the Government/Buyer:

Deliverable Description

Micro X-Ray Source Hardware unit meeting specifications in Section 4.0.

Power Supply Compatible power supply with all cables as well as cables for communication with computer

Software Controller Control software and license to operate and control x-ray source from computer

Technical Documentation

Operation manuals, interface control drawings, part drawings, and safety guidelines.

Safety Standards Certification

Documentation certifying compliance to the requirements of applicable IEC/EN and FCC standards

Test Data Certificate of Conformance and final test data verifying voltage, power, and spot size.

6.0 Other Considerations

6.1 Milestones

The following milestones will be monitored to ensure timely completion:

Milestone 1: Order Acknowledgement (1 week After Receipt of Order).

Milestone 2: Pre-shipment verification/testing.

Milestone 3: Final Delivery and Acceptance at destination.

6.2 Government-Furnished Property/Information

Property: None. No Government property will be furnished to the contractor.

Information: None. No Government information is required for this off-the-shelf procurement.

File details come from the government source that posted it. Updated .