Statement of Work_Oxygen Analyzer(s)- 20 Mar 2024.pdf

PDF 190 KB Posted

Attached to
Oxygen Analyzers Federal contract opportunity
Solicitation number
FA860124Q0106
Issued by
Department of the Air Force Materiel Command Lifecycle Management Center Wright Patterson Air Force Base

About this file

This document is a Statement of Work (SOW) for the procurement of two (2) Oxygen Analyzer systems by the Air Force Petroleum Agency's Aerospace Fuels Laboratory at Kadena Air Base. The key requirements for the Oxygen Analyzers include the capability to perform oxygen purity testing within 0-100% range with +/- 0.1% accuracy, have no internal pump, use 110-120V 60Hz power, and be delivered to Wright-Patterson AFB within 12 weeks of contract award. Training, installation, and maintenance services are not required. The related federal contract opportunity is a Request for Quote (RFQ) under Solicitation Number FA860124Q0106, seeking the same Oxygen Analyzer systems. The solicitation is a commercial item procurement using Simplified Acquisition Procedures, with quotes due by June 26, 2024. The Government will award a Firm Fixed Price contract to the responsible offeror submitting the lowest priced, technically acceptable quote.

View the file

Other files for this federal contract opportunity

Other files attached to Oxygen Analyzers, newest first.
File Type Posted
Combo Synopsis-Solicitation.pdf PDF
Solicitation - FA860124Q0106.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Statement of Work

For

Oxygen Analyzer

20 March 2024

I. Background:

The Air Force Petroleum Agency’s Aerospace Fuels Laboratory located at Kadena Air Base is seeking to purchase two (2) Oxygen Analyzer systems to determine oxygen purity of aviation breathing oxygen as well as trace analysis of gaseous samples via methods CGA G-4.3-2018 and CGA G-7.1-2018, respectively.

II. Minimum Technical Requirements:

The following is a list of requirements for an Oxygen Analyzer:

a. The Oxygen Analyzer shall:

i. Perform oxygen purity within the range of 0-100% with an accuracy of +/-

0.1%.

ii. Have no internal pump for flow.

iii. Have a sample inlet connection of ¼”.

iv. The capability to calibrate the high end with pure (100%) oxygen and a low end using pure inert gas such as nitrogen (N2).

v. Operate on power requirements of 110-120 V, 60 Hz.

vi. Be a benchtop not to exceed 400 mm (H) x 240 (W) x 350 (D)

b. Be delivered to Wright-Patterson AFB, OH within 12 weeks of contract award.

c. Be supplied with manuals (paper and electronic) for installation/setup and operation.

III. Training:

Training is not required.

IV. Installation:

Installation is not required.

V. Maintenance/Service:

Maintenance/service is not required.

VI. Delivery:

Delivery Address:

Aerospace Fuels Laboratory

AFPET/PTPLA (FP2070)

2430 C Street, Building 70, Area B Wright Patterson AFB OH 45433-7632

Aerospace Fuels Laboratory POC: Melissa Teague

Email: melissa.teague@us.af.mil Phone: 312-785-0238

VII. Lead Time:

We require this item by the earliest possible delivery date to not exceed 12 weeks after receipt of award.

VIII. Warranty:

Free from defects in material and workmanship, under normal and proper use in accordance with user manual, for the period of one year after receipt.