STATEMENT OF WORK.pdf
PDF 123 KB Posted
- Attached to
- G2131-I GAS ANALYZER Federal contract opportunity
- Solicitation number
- 80NSSC243985
About this file
This document is a Statement of Work for a federal contract opportunity. The key details are:
The government has a requirement for a high-precision analyzer for isotope measurement of delta[13]C from CO2 and concentrations of CO2 and CH4 in ambient air, specifically the Picarro G2131-i model. NASA/NSSC intends to issue a sole source contract to Picarro Inc. under the authority of FAR 13.106-1(b)(1)(i), as Picarro is the only manufacturer that can provide equipment compatible with NASA's existing instruments. The contract will be for commercial items and services using FAR Part 12 and FAR Part 13 procedures. The government is seeking responses and capabilities from interested organizations by 7/26/2024 to determine if this will be a competitive procurement. The NAICS code for this effort is 334516.
View the file
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
STATEMENT OF WORK
1. Objective/Requirements
The information listed below is required to complete this purchase.
2. Characteristics, Scope, and Specs
PART
NO.
DESCRIPTION QTY.
G2131-i High-Precision Analyzer for Isotope Measurement of delta[13]C from CO2 and Concentrations of CO2 and CH4 in ambient air
• Guaranteed delta[13]C-CO2 precision (1 sigma over 1 hour in 5 min average) :<0.1‰
• Measurement interval: 2 sec; Rise/Fall Time (10- 90% / 90 - 10%): 30 sec
• Automatic water correction to report dry mole fraction
• Compatible with various peripherals for dissolved gas, solids, discrete samples or flux chambers
• Leverages Picarro's unique CRDS technology to make time-based laser measurement for quantifying spectral features in an optical cavity
• Proprietary wavelength monitor (WLM) controls laser wavelength to within 2 MHz for precise quantification of spectral features
• Optimized flow path and small cavity (< 35ml) for fast sample turnover rate
• Temperature & pressure stabilization within 0.005 °C and 0.0002 atm
• Field deployable; lightweight <55 lbs (25 kg) & compact in size 17" W x 7" H x 17.5" D (43 x 18 x
45 cm)
• Includes: Keyboard, Mouse, and Mouse Pad Includes external pump and free remote technical support
SH1002 Freight Prepaid and Added, for shipping in USA. We will contract our freight forwarder to deliver to your destination (please specify). The amount specified here represents shipping and insurance charges, all of which will be prepaid by Picarro as a courtesy service and charged back to customer on the invoice.
S0512 Additional N2 Background Gas Mode for Standard Analyzer CO2 Concentration Range (380-2000ppm)
• Installation and calibration of a separate mode that uses N2 as a background gas.
• For use with Combustion Module (A0201), Automate Prep Device (A0304) or any other frontend that uses N2 as a carrier gas.
• • Available for new analyzer purchase only.
S0511 13C optimization for Low CO2 Concentration (200 ppm) in N2 and/or Air background gas
• Ensures low concentration isotopic CO2 performance with appropriate corrections down to 200 ppm CO2, and up to 5% H2O.
• Available for new analyzer purchase only.
S0506 Enriched 13C Isotopic CO2 Calibration (up to 6500permil)
• To test and calibrate analyzer for enrichment levels up to 6500permil at 380ppm CO2.
• Applicable to G2131-i and G2201-i only.
A0201 Combustion Module For use as part of the CM-CRDS system for bulk stable isotope analysis using 13C.
• Basic furnace design with large capacity 25mm reactor for combustion/reduction
• 50 position carousel autosampler
• Can be extended with two additional carousels for a total of 148 samples
• Easy replaceable inserts for removal of ash
• Must be combined with an interface (A0301 or A2100) and an iso-CO2 analyzer
A2100 Picarro Caddy Interface Kit
• Designed for coupling a Picarro Combustion Module to a Picarro isotopic CO2 Analyzer
• Includes all hardware and software interfacing material
A0924 O2 and N2 dry gas kit for Combustion Module – CGA. This kit contains plumbing required to connect O2 and N2 to the Picarro Combustion Module (A0201), kit includes:
• High-purity pressure regulators
• CGA connector and Swagelok fittings
• Rotameter
• Tubing, cutters, and assembly instructions.
C0198 CM-CRDS Consumables Kit for Caddy – 1000 samples Consumable kit for 1000 samples, use with Caddy (A2100). Kit contains:
• Quartz tube, catalysts & inserts
• 1000 tin capsules
A0304 AutoMate Prep Device
• Current lead time 9 weeks
• Automatic preparation system for acidification of solid carbonates and DIC in water.
• Fully automated delta(13)C analysis when coupled with Picarro’s Caddy™ interface and iso- CO2 WS-CRDS system.
• Suitable for shipboard, field, and laboratory automated analysis of carbon isotopes of solid carbonate and DIC. Release 2024
C0310 3-year Consumables Kit for AutoMate Prep Device (A0302) Kit includes:
• 2 sets of needle kit
• 1 set of replacement pumps
• 1x replacement flow meter
S2084 On-site installation and training for 1 (one) Picarro analyzer. Includes travel and lodging. 1
3. Place of Performance
Goddard Space Flight Center Greenbelt MD 20771 United States
4. Period of Performance
Estimated delivery lead-time is 8-16 weeks ARO.
File details come from the government source that posted it. Updated .