STATEMENT OF WORK.pdf
PDF 142 KB Posted
- Attached to
- SVC FIELD PORTABLE SPECTRORADIOMETER Federal contract opportunity
- Solicitation number
- 80NSSC242185
About this file
This document is a Statement of Work (SOW) for a NASA requirement for a SVC Field Portable Spectroradiometer. The SOW provides detailed specifications for the spectroradiometer, including the spectral range, resolution, detectors, field of view, and physical characteristics. It lists the standard equipment to be provided, including a rugged smartphone, software, and accessories. The estimated delivery time is 5-7 weeks after receipt of order.
The related federal contract opportunity is a pre-solicitation notice indicating that NASA intends to award a sole-source contract to Spectra Vista Corporation for the spectroradiometer, as the manufacturer is the only source compatible with existing NASA instruments. The procurement will be conducted under FAR Part 12 for commercial items. Responses to this notice to determine if the requirement should be competed are due by September 6, 2024.
View the file
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
STATEMENT OF WORK
1. Objective/Requirements
The information listed below is required to complete this purchase.
2. Characteristics, Scope, and Specs
LINE PART NO. DESCRIPTION QTY.
1 HR-1024i SVC Field Portable Spectroradiometer
Optional Equipment:
4.0 Meter Fiber Optic Light Guide 25° FOV including pistol grip Radiometric Calibration for primary 4.0 meter Fiber Optic
4.0 Meter Fiber Optic Light Guide 25° FOV including pistol grip (Spare) Radiometric Calibration for spare 4.0 meter Fiber Optic 3 inch Direct Connect Full Sky Irradiance Shere Irradiance Calibration for the Full Sky Irradiance Sphere Battery Eliminator (AC Power Supply)
SVC HR-1024i Field Portable Spectroradiometer Features
Provides the exceptional spectral resolution operating across this full spectral region Use of 100% linear array technology ensures superior wavelength stability. The InGaAs and extended InGaAs detectors are cooled for enhanced stability and sensitivity All critical optical components are hard mounted to the spectrometer platform Provides fast, full spectral measurements with no moving gratings Fixed foreoptics ensure a rugged and reliable optical path even at low temperatures Field swappable fiber optic light guides (optional) are available in various lengths Internally mounted camera provides a photo record of each target measurement The internal CPU enables measurements to be made without an external computer The SVC HR-1024i internal memory stores up to a full day’s data without download The CAT S62 rugged Android smart phone with thermal camera is supplied The on-board Lithium-Ion battery is quickly changed and enhances mobility The SVC HR-1024i is so small, light and completely self-contained that it is easily operated by one person in the most challenging field conditions.
The 1/4VGA color LCD displays data graphically and facilitates stand alone set-up A second Bluetooth is provided for communication with up to 16 external sensors Internal inclinometer enables correct instrument positioning and records offset
SVC HR-1024i SPECIFICATIONS
Spectral Range 350 – 2500 nm Spectral Resolution 3.5 nm, 350 – 1000 nm
9.5 nm, 1000 – 1880 nm
6.5 nm, 1880–2500 nm
Spectral Sampling Bandwidth 1.5 nm, 350 – 1000 nm
3.8 nm, 1000 – 1880nm
2.5 nm, 1880 – 2500 nm
Detectors 512 Si array 350 – 1000 nm 256 InGaAs array 1000 – 1880 nm 256 Extended InGaAs array 1880 –2500 nm
FOV (nominal) 4º std., 8 º & 14º optional foreoptics 25º optional fiber optic
Stray Light < 0.1% @ 500 nm
NER
1.0 sec. integration time, W cm-2 nm-1sr-1
< 0.8 x 10 -9 @ 700 nm < 1.2 x 10 -9 @ 1500 nm
< 1.2 x 10 -9 @ 2100 nm Max. Radiance W cm-2 nm-1sr-1 1.5 x 10 -4 @ 700 nm Calibration Accuracy (NIST Traceable) ± 5 % @ 400 nm
± 4 % @ 700 nm ± 7 % @ 2200 nm
Minimum Integration Speed 10 ms VNIR, 1.0 ms SWIR Wavelength Reproducibility 0.1 nm Wavelength Accuracy ±1 nm VNIR, ±2 nm SWIR
SVC HR-1024i Physical Specifications
Power 7.4 V Instrument Size ≤8.75" x 11.5" x 3.0"/ 22 cm x 29 cm x 7.6 cm Weight ≤8.5 lbs., 3.8Kg Battery Weight ≤1.5 lbs., 0.7Kg Host Communications Interface RS-232, USB, Bluetooth Sealing Sealed against dirt and dust Tripod mounting Standard ¼”-20 mounting
SVC HR-1024i Standard Equipment
The SVC HR-1024i Spectroradiometer System is supplied with the following:
SVC HR-1024i Spectroradiometer As outlined above
InGaAs and extended InGaAs arrays to have
Rugged Smart Phone CAT S62 with thermal camera and Bluetooth
Foreoptic: Standard 4° FOV foreoptic (removable)
Other foreoptics are quoted
Field Case Rugged water-tight Pelican field case, high density foam
Battery Charger: (2) 120/240 VAC 50/60 Hz input
Batteries: (2) rechargeable batteries
Communication Cable: (1) two meter communication cable
Software Programs: PC S/W, runs under Windows 10 & 11
Android App installed on CAT 62 Data may to be imported into ENVI
Calibration Files: Wavelengths & std. foreoptic radiometric supplied
User Manual: Hard & Soft copy of user manual is provided in English
3. Place of Performance
Attn: Dr. Julia A. Barsi Lead, RCL, DCL, GLAMR NASA Goddard Space Flight Center 8800 Greenbelt Road
MD 20771
4. Period of Performance
Estimated delivery lead-time is 5-7 weeks ARO.