8-Channel System Specifications – Animal Disease Research Unit, Pullman, WA

Types of analysis and applications
Binding: Yes/No, kinetics affinity, steady-state affinity, affinity maturation, competition/cross-reactivity

Antibody: isotyping, epitope binning Target concentration in complex biologic samples: relative and absolute

Affinity constant

(KD)

100 pM – 1 mM

Kinetic rate ranges (minimums)
Association: 101 to 106 M-1sec-1 Dissociation: 10-6 to 10-1sec-1
Molecular weight minimum detection limit
~300 Daltons
Quantitation minimum
Sensor options should include sub-microgram per ml detection
Real-time minimum data acquisition rate
5 Hz
Baseline noise maximum
< 3 pM RMS
Baseline drift maximum
< 40 pm/hour
Sample type compatibility*
Proteins, peptides, antibodies, DNA, RNA, small molecule, liposomes, virus, and compatible with relatively crude sample homogenates.
Probe-based
Minimum two probes

Some probe options include regenerative technology

Throughput
8 probes in parallel, 96-well plate format minimum
Orbital shaking
Variable with a maximum of 1,500 RPM
Sample capacity
Minimum of two loaded 96 well plates
Sample volume
Minimum 200 uL (96-well plate format)
Integrated data acquisition and analysis (includes monitor)
Assay setup and Analysis templates Real-time display of results

Algorithms for multiple binding models One-run measurement of quantitation, kinetics, and regeneration

Probe binding methods
1. Binding to biotinylated or Avi-tagged biomolecules

2. Covalent binding through amine group of proteins

3. Binding to His-tagged proteins (antibody-based method, Ni-based method)

4. Binding to antibodies (Fc method, FAB method)

* Microfluidic flow-thru technology (SPR) is not compatible with crude homogenates, which may occlude the system and stop sample flow. Probe-based (BLI) technology is fully compatible with crude homogenates.