Statement of Work BBR.pdf
PDF 174 KB Posted
- Attached to
- Cannon Instruments Bending Beam Rheometer
- Solicitation number
- 693C7321Q000003
View the file
Other files
| File | Type | Posted |
|---|---|---|
| RFQ NO. 693C7321Q000003.pdf | ||
| BRAND NAME JUSTIFICATION.docx | DOCX document | |
| PROVISION 52.204-26.pdf | ||
| Sec 889 Offeror Representation Form.pdf | ||
| PROVISION 52.204-24.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Statement of Work
This statement of work is for the purchase of a Bending Beam Rheometer used in AASHTO T 313 and ASTM D 6648. The Bending Beam Rheometer is used to measure deflection of an elastic beam. The results from this device will be used to provide stiffness modulus values and m-value for asphalt binder material.
Specification/Requirements
A. Bending Beam Rheometer Apparatus Designed and meets AASHTO T 313 and ASTM D 6648 test requirements.
• Due to space limitations and current configuration, the overall dimensions of the control unit should be no greater than
ₒ 29”W x 28”D x 22” H
• Temperature Range: 0 to -40 degrees Celsius
• Temperature Stability: +/- 0.03 °C, +/- 0.01 °C resolution
• Data Output: USB
• Meet or Exceed ASTM D 6648, AASHTO T 313
• Factory Calibration in accordance with ISO 9001
• Minimum Specimen Bath Volume: 1.33 Gallons(5L)
• Pneumatic pressure regulators to adjust pressure on main input line, air bearing and load shaft support.
• Provide a precision calibration kit which meets or exceeds ASTM D 6648 and
AASHTO T 313. This includes guage block, stainless thin beam, ¼” compliance beam, four 100 gram weights with NIST Traceable certifications, aluminum molds with seperators, Digital thermometer with probe and data storage management software.
B: Electrical Requirements
1. Operate on standard U.S. power systems.
C: Service & Support
1. Phone support of equipment installation and initial hardware/software set up as well as support initial calibration requirements.
Address: 1827 Jack Delozier Dr.
Sevierville, TN 37876
2. The Bending Beam Rheometer supplier must be capable of support through phone, e-mail, and network. Warranty for one year.
File details come from the government source that posted it. Updated .