Statement of Need For 60 MDG/SGGX (Hemodialysis) Description: The Joint VA/DoD Hemodialysis/Nephrology Unit requires the replacement of six (6) portable Reverse Osmosis (RO) units. Currently, one of six (6) units is operational with no back-up availability. An RO unit is intended for use with hemodialysis systems to remove organic and inorganic substances and microbial contaminants from the water used for treating hemodialysis patients or related therapies. Functional machines allow our staff to provide emergent/urgent hemodialysis treatments for patients in the Intensive Care, Inpatient, COVID and Chronic unit (Case-by-case basis, example-Central RO non-functionable). A portable RO unit treats city water are an essential piece of equipment for acute treatments. Without functioning portable RO units, acute hemodialysis treatments will not be able to be performed, patients requiring acute services will be required to be diverted to civilian facilities emergently resulting in delays of treatment, further damage to internal organs and possibly death. This is a “brand name or equal” request.
POC: Kimberly A. Green Unless determined otherwise, the contractor will provide the following:
1. Equipment.
| 1.1 Installation and set-up must be included |
| 1.2 Installation and set-up must be performed between the hours of 6am-2pm Mon-Fri |
| 1.3 Must meet ANSI/AAMI 13959 standards |
| 1.4 Must include all accessories and provided at time of set up |
| 1.5 Must be energy efficient |
| 1.6 Must be “Heat Forward Technology” meaning a heat disinfection is performed daily |
| 1.7 Inlet temperature must be 40-104 degree Fahrenheit |
| 1.8 Inlet pressure must be between 5 to 120 psi |
| 1.9 Minimum inlet water flow must be 1.5 gpm |
2. Perform testing and certification ensuring RO elements meets NSF/ANSI regulatory and standard requirements.
2.1 NSF/ANSI 58: RO Drinking Water Treatment Systems and Elements
2.1.1 The American National Standard for point-of-use (POU) RO systems and elements. Designed to reduce contaminants that may be present in public or private drinking water
2.2 NSF/ANSI/CAN 61: Drinking Water System Components – Health Effects
2.2.1 Covers products used in the treatment and distribution of drinking water (including RO elements) > Evaluates the health effects of contaminants that may be contributed to drinking water from the materials in the treatment or distribution equipment.
3. Training
3.1 Must include end-user (30 personnel) operational training on device
3.2 Must include Medical Equipment Repair Technician (3 personnel) training on service/maintenance
4. Electrical requirements: 115/230V, 60/50Hz, 15/7.5Amps
5. Equipment warranty (12 months) must be included
6. All freight, handling, & labor charges must be included
7. Point of Contact after contract is awarded: Coordinate with Mr. Dan Dettwiler at DGMC at 707-423-7506.