To procure two fume hoods for LMFS's newly renovated bldg. 3
Solicitation number
1333MJ24Q0024
Issued by

About this file

This document is a Statement of Need for the procurement of two fume hoods to be used in the renovated Building 3 at the Lake Michigan Field Station (LMFS) in Muskegon, MI. The hoods will be used to protect human health by exhausting air contaminated with chemicals like formalin and DMF.

The technical specifications call for a 6-foot wide fume hood with a vertical sash, gooseneck faucet, cup sink, outlet, lighting, lower flammable cabinets, and a flow alarm. The 4-foot wide hood has similar features. The delivery schedule requires the hoods to be installed by August 15, 2024. Pricing is to include ASHRAE 110 certification, shipping, and installation. The related federal contract opportunity, Solicitation Number 1333MJ24Q0024, is being issued by the Department of Commerce's National Oceanic and Atmospheric Administration to procure these two fume hoods for the LMFS's newly renovated Building 3.

View the file

Other files

File Type Posted
Sol_1333MJ24Q0024.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

DATE

STATEMENT OF NEED

145025-24-0091

1. PURPOSE:

To procure two bldg 3 fume hoods to be used in the renovated bldg 3.

2. SCOPE OR MISSION:

We are a multidisciplinary environmental research laboratory that provides scientific understanding to inform the use and management of Great Lakes and coastal marine environments. The Lake Michigan Field Station (LMFS) in Muskegon, MI is a government owned facility in support of all missions of the Great Lakes Environmental Research Laboratory (GLERL) in Ann Arbor, MI.

3. BACKGROUND:

Two fume hoods to be used in the renovated LMFS Building 3 Preserved Sample Lab (6' hood) and Dark Room (4' hood). The hoods will be used to protect human health by exhausting air contaminated with formalin, DMF, and other chemicals out of the lab. The fume hoods are necessary equipment to meet air quality safety standards in our laboratory.

4. TECHNICAL SPECIFICATIONS:

6' wide hood with a vertical sash (chain and sprocket mechanism). Gooseneck faucet (cold water input) and a cup sink mounted on the left side. One outlet and within hood lighting. Lower flammable cabinets for chemical storage. Upper furring strips. Flow alarm.

4' wide hood with a vertical sash (chain and sprocket mechanism). Gooseneck faucet (cold water input) and a cup sink mounted on the left side. One outlet and within hood lighting. Lower flammable cabinets for chemical storage. Upper furring strips. Flow alarm.

5. DELIVERY SCHEDULE:

ARO to be completed NLT 15 Aug 2024.

6. SPECIAL REQUIREMENTS:

Note: Pricing to include ASHRAE 110 certification, shipping, and installation.