SSJ_DAQ redacted.pdf

PDF 412 KB Posted

Attached to
HIGH-SPEED NUCLEAR DATA ACQUISITION AND PROCESSING SYSTEM Federal contract opportunity
Solicitation number
FA8601-HighSpeedNuclearDAQ
Issued by
Department of the Air Force Materiel Command Lifecycle Management Center Wright Patterson Air Force Base

About this file

This document is a Single Source Justification (Simplified Acquisition) for the procurement of a High-Speed Nuclear Data Acquisition and Processing System. The key requirements are the ability to simultaneously sample at 500 MS/s on 40 channels, with 14-bit analog-to-digital conversion, individual self-triggering, and a communication interface enabling data transfer at 80 MB/s to 50 TB of data storage. The system must be compatible with existing AFIT/ENP equipment and software. The Government intends to sole source this requirement to Caen Technologies under FAR 13.106-1(b)(1) authority, as they are the only known manufacturer capable of meeting the specifications. The proposed NAICS code is 334516 with a size standard of 1,000 employees. All responsible sources may submit capability statements, proposals, or exceptions to the sole source intent, which will be considered by the Government.

View the file

Other files for this federal contract opportunity

Other files attached to HIGH-SPEED NUCLEAR DATA ACQUISITION AND PROCESSING SYSTEM, newest first.
File Type Posted
08 Statement of Work Fast Nuclear DAQ System 230821.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Single Source Justification (Simplified Acquisition)

CONTROLLED UNCLASSIFIED INFORMATION

April 2021 CONTROLLED UNCLASSIFIED INFORMATION Page 1 of 2April 2021 CONTROLLED UNCLASSIFIED INFORMATION Page 1 of 2

The measurement of short-lived nuclear events requires instrumentation with fast acquisition rates. Further, as multi-detector systems are often employed in these experiments to improve counting efficiency and/or measure additional event characteristics, there is a need for simultaneous measurement across several data acquisition channels. Real-time monitoring of experiment status is facilitated by the associated pulse processing software.

AFIT's current nuclear digital data acquisition capability is not at the state-of-the-art in digitization rate, computational speed, or storage capacity. Whereas the digitization frequency of modern data acquisition equipment is about 100 GHz, AFIT's current capability is 5 GHz at best, and usually slower. As a result, nuclear events of 100 picoseconds or less cannot be measured by students and faculty in our Nuclear Engineering program.

An appropriate High-Speed Data Acquisition and Processing System for Nuclear Measurements includes several subsystems: Digital Pulse Acquisition, Digital Pulse Processing and Transduction, Fast Optical Download, and Data Storage. These integrated hardware components are typically accompanied by control modules for instrument set-up and data acquisition, and for post-collection data analysis.

In one intended research application, the data acquisition system will interface with the recently procured dual, double-sided strip detector (dDSSD), which outputs 40 channels of data (from two 10x10 imaging arrays).

The limiting factor in image acquisition rate is the speed of the multichannel acquisition hardware and the transfer of data from the acquisition buffer into the data storage system. The currently available (analog) acquisition hardware severely limits data collection and, consequently, the range of applications in which the dDSSD would be considered an acceptable imager choice. Here, the data download capability is also key and fast throughput would allow efficient generation of rich datasets amenable to Artificial Intelligence interrogation s. In another application, the fast data acquisition system would allow coincidence detection of multiple nuclear events with the necessary timing resolution to investigate material properties with much greater accuracy.

The minimum specification for the required system are:

Ability to sample waveforms from charge-sensitive preamp of radiation detector, with unipoloar Vpeak ≤ 2 V

Simultaneous sampling at ≥ 500 MS/s on ≥ 40 channels, with all channels time-synched

≥ 14-bit analog-to-digital conversion (ADC) resolution

Internal ADC sampling clock with ≥ 50 MHz timing resolution; capability to utilize an external clock signal

Individual self-trigger with channel over/under threshold, and trigger stamp resolution of better

Single Source Justification (Simplified Acquisition)

CONTROLLED UNCLASSIFIED INFORMATION

April 2021 CONTROLLED UNCLASSIFIED INFORMATION Page 2 of 2April 2021 CONTROLLED UNCLASSIFIED INFORMATION Page 2 of 2 than 100 ps

Memory buffer ≥ 5 MS/channel

Communication interface that enables data in all active channels to simultaneously transfer data to data storage at a rate ≥ 80 MB/s, with included hardware for data storage (≥= 50 TB)

Online digital pulse processing with ability to customize gating, pulse shaping parameters and pulse discrimination (over/under threshold, coincidence, veto, etc.)

Compatible DAQ software that allows independent control of each active channel

All necessary cables and interfaces to implement ≥ 40 channels of simultaneous data acquisition, data processing, and transfer to data storage

New Equipment must be compatible with existing AFIT /ENP equipment and use the same custom software to control, acquire and analyze collected data.