18 January 2017
NRL 5524
PURCHASE DESCRIPTION
OSCILLOSCOPE, DIGITAL 1 GHZ
GMG: OSC1GHZ-18
Solicitation No: SPRMM1-17-R-YK33 / SCAT: 4314
1.0 GENERAL This procurement requires a four channel 1 GHz digital storage oscilloscope with a real time sample rate of 5 GS/sec when four channels are active.
2.0 CLASSIFICATION This equipment shall meet the requirements of MIL-PRF-
28800F class 3 for Navy shipboard, submarine, shore, and Marine Corps applications with the exceptions specified in section 4.1.
3.0 OPERATIONAL REQUIREMENTS The specifications provided below are the minimum requirements and accuracies that will meet the Government's need.
The oscilloscope shall at least meet these requirements and accuracies and may provide better performance.
3.1 Vertical System
3.1.1 Input Channels: The equipment shall be capable of simultaneous and synchronous acquisition with 4 input channels.
3.1.2 Bandwidth (-3 dB roll-off): DC to 1 GHz with 50 Ohm input impedance at ≥ 5 mV/div.
3.1.3 Rise Time: ≤375 ps 10% to 90%.
3.1.4 Transient Capture: The equipment shall be capable of capturing the representation of transients ≤1 ns in duration.
3.1.5 Bandwidth Limit: The equipment shall have a low-pass filter with a nominal cut-off frequency of 30 MHz or less.
3.1.6 Sensitivity: 1 mV/div to 5 V/div at 1 MOhm input impedance and 1 mV/div to 1
V/div at 50 Ohm input impedance.
3.1.7 DC Offset: At least equal to the sensitivity multiplied by the number of vertical divisions.
3.1.8 DC Gain Accuracy: ±2% of full scale.
3.1.9 Vertical Resolution: The analog to digital converter shall provide eight bits of resolution.
3.1.10 Effective Bits: The analog to digital converter shall provide 5.5 bits of resolution at 1 GHz at the maximum sampling rate.
3.1.11 Input Impedance: Selectable 1 MOhm ±5% shunted by 15 pF typical or 50 Ohm
±5%.
3.1.12 Maximum Input: 200 V (DC + peak AC, 150 Vrms minimum) at 10 KHz or less with 1 MOhm input impedance and 5 Vrms with 50 Ohm input impedance.
3.1.13 Channel Isolation: 100:1 up to 100 MHz; 30:1 above 100 MHz.
3.1.14 Input Coupling: AC and DC coupling with 1 MOhm input impedance and DC coupling with 50 Ohm input impedance. Ground coupling or displayed ground indicator for both input impedances.
3.2 Horizontal System
3.2.1 Time Base Range: 200 ps/div to 20 s/div.
3.2.2 Time Base Accuracy: 50ppm (0.005%) + (0.2)(sample period).
3.2.3 Pre-trigger Delay Range: 0-100% of the record length.
3.2.4 Post-trigger Delay: 20 ns to 10 s.
3.3 Acquisition System
3.3.1 Real Time Sample Rate: 5 GS/s per channel when four channels are active.
3.3.2 Record Length: 40 Mpoints with 1 channel active, 20 Mpoints with 2 channels active, and 10 Mpoints with 4 channels active.
3.3.3 Waveform Update Rate: 200,000 waveforms/s with no filtering, processing, or math functions applied. A waveform shall consist of 1024 points minimum.
3.4 Trigger system
3.4.1 Trigger Sensitivity: 0.5 divisions DC to 100 MHz, 1 division 100 MHz to 1 GHz.
3.4.2 Trigger Sources: Channels 1- 4 and external.
3.4.3 Trigger Holdoff Range: 100 ns – 320 ms
3.4.4 Trigger Out: The equipment shall have a trigger out connector that can trigger an external device.
3.5 Parametric Measurements
3.5.1 Automatic Measurement: The equipment shall be capable of automatically measuring and displaying the following parameters of the displayed signal:
period, frequency, width (+ and -), rise time, fall time, peak to peak, min, max, overshoot (+ and -), duty cycle, area, RMS amplitude, std. deviation, and mean.
3.5.2 Measurement Display: Simultaneous display of six automatic measurements.
3.5.3 Math Function: The equipment shall be capable of performing the following functions on two waveforms: add, subtract, multiply, divide, average, invert, square, square root, min, max, integrate, and differentiate.
3.5.4 Math Display: Simultaneous display of four math waveforms.
3.5.5 Eye Pattern: The equipment shall be capable of automatically measuring and displaying the following communications related eye pattern measurements:
eye height, eye width, jitter, and duty cycle distortion.
3.6 Fast Fourier Transform (FFT)
3.6.1 Frequency Range: DC to 1 GHz.
3.6.2 Window Functions: Hanning, Rectangular and Flattop Window functions shall be provided.
3.6.3 Measurement Display: FFT amplitude and frequency.
3.7 Display The equipment shall be provided with a color LCD display screen that shall be fixed to the equipment case and not rotate, pivot, or tilt.
3.7.1 Display Size: ≥10" measured diagonally.
3.7.2 Display Grid: Ruled, with eight vertical divisions and ten horizontal divisions minimum.
3.8 Touch screen A touch screen user interface shall be provided. The screen shall allow the operator to access and select all parametric measurements listed in section 3.5. Screen shall also allow the user to access, select, and name waveform files without the use of a mouse or keyboard.
3.9 Automatic Setup The equipment shall be equipped with a single push-button control that will initiate automatic adjustment of the vertical and horizontal deflection factors and trigger level for an optimized display of the input signal. This function shall operate with signals exceeding 1% duty cycle, 10 mV amplitude, and frequencies of 50 Hz.
3.10 Setup Storage The equipment shall be capable of saving a setup file that can be transferred for use in identical equipment. A USB port must be provided to accommodate a flash drive for storing and uploading setup files.
3.11 Waveform Storage The equipment shall be capable of storing and replaying four waveforms with associated vertical and horizontal settings in nonvolatile memory. A USB port shall be provided to accommodate a flash drive for storing waveforms.
3.12 Waveform Analysis The equipment shall have a feature that will allow signal evaluation of selected events or portions from captured waveforms. The feature shall allow the user to see a larger representation of the selected portion of the waveform.
4.0 GENERAL REQUIREMENTS
4.1 Environmental
4.1.1 Operating temperature:
4.1.1.1 Without accessing storage devices (hard drive or CD-ROM): 0º C to +45º C.
4.1.1.2 With accessing storage devices (hard drive or CD-ROM): +5º C to 40º C.
4.1.2 Non-operating temperature: -20º C to +60º C.
4.1.3 Operating altitude: 0 to 3000 meters.
4.2 Digital Interface A USB port to support a digital interface in accordance with
MIL-PRF-28800F and a RJ-45 connector to support 10/100 Base-T Ethernet interface for remote access and control shall be provided.
4.3 Printing Capability An inherent printing capability is required.
4.4 Power source MIL-PRF-28800F nominal power source requirements are invoked. Maximum power consumption 390W.
4.5 Battery Restrictions Per MIL-PRF-28800F, Lithium and Mercury batteries are prohibited without prior authorization. If equipment qualifies for one or both of the exceptions stated below, then the offeror shall submit the required information at the time of submission of proposals. If equipment does not qualify for an exception, then a request for approval for the use of Lithium and Mercury batteries shall be submitted with production lot delivery, after contract award. Approval shall apply only to the specific model proposed.
Exception 1: Per NAVSEA S9310-AQ-SAF-010, the use of non-rechargeable lithium coin cell batteries meeting the following criteria is approved for all uses and does not require individual testing and review by the Naval Ordnance Safety and Security Activity (NOSSA).
- Unmodified, commercially available off-the-shelf (COTS) item;
- Used in single-cell configuration;
- Maximum nominal output of 3 Volts;
- Maximum rated capacity of 1 Ah.
The coin cell manufacturer, part number, nominal voltage, and rated capacity shall be provided at the time of submission of proposals.
Exception 2: Per NAVSEA S9310-AQ-SAF-010, the use of COTS electronics and equipment powered by lithium ion rechargeable (secondary) batteries meeting the following criteria is approved for all uses and does not require individual testing and review by NOSSA.
- Unmodified, COTS battery;
- Underwriters Laboratories (UL) listed;
- Used in the device as recommended by the manufacturer. Modifications to the devices may only be made in accordance with the manufacturer’s recommendations; e.g., addition of memory;
- Recharged only by devices expressly designed for recharge of the specific battery in use;
- No more than four cells in series (less than or equal to 18-Volt output);
- Rated for no more than 100 Watt-hours, as listed in the manufacturer’s specification or calculated by multiplying the capacity in Ah by the maximum working (nominal performance) voltage.
The lithium ion rechargeable cell manufacturer, part number, configuration (number of cells in series), nominal voltage and rated capacity shall be provided at the time of submission of proposals.
4.6 Accessories Two 1 GHz X10 probes that have a nominal length of 1.3 to 1.5 meters and provide full bandwidth performance in accordance with section 3.1 when used with the equipment, and two 250 MHz (minimum) X10 probes with a nominal length of 2.0 meters. Scope probes will be provided with the following accessories: dual (ground and center conductor) wire wrap adapter, retractable clip tip adapter, micro-retractable clip tip adapter; Scope Probe Ground: 6 in ground lead, wire wrap adapter, alligator clip adapter.
4.7 Accessory Pouch The equipment shall be provided with a detachable accessory pouch attached to the equipment or an integrated compartment for storage of equipment accessories.
4.8 Equipment Case The equipment case shall be designed for bench top use. No openings are permitted in the top surface of the enclosure. The case shall include a front panel cover or protective handles that protrude from the front panel.
4.9 Calibration Interval The calibration interval shall be 12 months. At the end of this interval, a minimum of 85% of the equipment shall remain in tolerance.
4.10 Calibration Procedure The procedure, software, and special interfaces/adapters that are needed for the equipment calibration shall be provided.
4.11 Dimension Nominal 330 mm (13”) H x 450 mm (17.7”) W x 450 mm (17.7”) D.
4.12 Weight Not to exceed 15 kg (33 lbs).
4.13 Wireless Communication Any capability of the equipment to communicate wirelessly, including but not limited to Wi-Fi and Bluetooth, shall be disabled.
4.14 Technical Manual The maintenance philosophy for this unit shall be level 2
(per MIL-PRF-28800F) and require maintenance to the module level of the unit.
The technical manual shall conform to the level 2 maintenance philosophy.
This level would be used for most equipment where maintenance and repair is an expected phase of equipment lifecycle. Board level maintenance and troubleshooting information is required. A Use and Installation manual
(Operator’s Manual) shall be provided separately. Maintenance and Servicing manual shall be provided to two levels of maintenance, unit operational verification level and the module level.
Information required for performance verification shall include:
• Instructions to verify equipment performance,
• List the equipment required for verification tests,
• Step-by-step instructions for test connections,
• Acceptable result criteria,
• Calibration information,
• Self-test routines.
Maintenance information shall include:
• Parts lists to the component level,
• Schematics and component layout drawings,
• Block and schematic diagrams.
• List of required test equipment and connection diagrams, and
• Sequential instructions for disassembly, repair, replacement, and reassembly shall be provided.
• Board level maintenance and troubleshooting information,
• Step-by-step instructions for troubleshooting and fault isolation,
• Expected signal levels,
• Test data sheets will be included, and as required,
• The instructions will define localizing a defective circuit card.
Parts lists shall include:
• Parts lists shall be shown on illustrations or a separate listing that includes an index or reference to other illustrations.
• Part number, cage code, and generic description.
The technical manual shall be provided in both printed and electronic formats.
The printed format shall be otherwise normally provided. The electronic format shall be in Portable Document Format (PDF) - ISO 32000-1:2008. Two separate CD/DVDs are required, one shall contain the Use and Installation manual and one shall contain the Maintenance and Service manual.
The statement of copyright release:
“REPRODUCTION AND DISTRIBUTION OF THIS TECHNICAL MANUAL IS
AUTHORIZED FOR U.S. GOVERNMENT PURPOSES.”
shall be printed in the first two pages of each technical manual and on the surface of the CD/DVD supplied.
4.15 Training Materials Training materials that demonstrate the features, detailed operations and procedures with step-by-step instructions for using the equipment shall be provided. The training material shall be delivered in technical manual format or interactive CD-ROM format.
4.16 Additional Requirements
4.16.1 Human Readable Identification Labeling:
4.16.1.1 Equipment: Per MIL-PRF-28800F, a human readable label shall be provided for all production lot units conforming with MIL-STD-130N and permanently affixed on the equipment in an easily readable location. Required fields on the label are; CAGE, part number, and serial number. Size of the label shall conform to the size of the equipment.
4.16.1.2 Case labeling: When a hard transit case is required, a human readable metal plate per MIL-PRF-28800F shall be provided for all production lot units conforming with MIL-STD-130N and permanently affixed to the front of the transit case. Required fields on the label are; CAGE, part number, and serial number. Size of the label shall conform to the size of the case. Pressure sensitive adhesive transfer tape is required to hold the plate to the hard transit case such as 3M™ 9472LE. When a soft transit case is required, labeling with permanent placement such as a metal plate affixed with rivets, screws or adhesives shall be provided.
4.16.2 Shipping container: For production lot units the package or carton containing the equipment for shipment shall be marked per MIL-STD-129P.
4.16.3 Other Additional Requirements: Shall be specified in the CDRL of the solicitation.