Specifications X-Ray_4 SEP 2024.docx
DOCX document 23 KB Posted
- Attached to
- X-Ray Inspection System
- Solicitation number
- PANDTA24P0000010427
About this file
This document appears to be a set of specifications for an X-Ray Inspection System that Sierra Army Depot (SIAD) is seeking through a federal contract opportunity. The key details include:
The required system must have a tunnel size capable of scanning objects up to 59 inches wide by 70.9 inches high, a conveyor with a pulling weight of at least 1,200 kg (2,646 lbs) and bidirectional movement, and the ability to connect to external monitors and recall at least 10 previously scanned images. Technically, the system must provide at least 28 mm of steel penetration, 38 AWG wire resolution, a 100% duty cycle, less than 0.1 mR per hour dose per exam, and less than 0.5 mR per hour radiation leakage.
The related federal contract opportunity is a pre-solicitation notice from the Department of the Army Materiel Command Contracting Command Detroit Arsenal, seeking sources who can provide the required X-Ray Inspection System. This is a market research effort, and no solicitation exists at this time. Interested parties are requested to submit capability statements, including their socioeconomic status, UEI, and CAGE code, to the contracting officer.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
X-Ray Inspection System Desired Specifications
General
· Tunnel Size: Must be able to scan objects up to 59 inches x 70.9 inches (W x H)
· Conveyer: Pulling weight must be no less than 1200 kg (2,646 lbs) and belt movement must be bidirectional (forward and reverse)
· Operating System: Must be able to connect to external monitors via HDMI cable. Must contain an image recall function that can show at least 10 previously scanned objects. Must include controls for belt movement with options to stop, move forward and backward. Comment by Worrell, Tracy E CIV USARMY ACC (USA): Do we want to address the operator control panel and ability to at a minimum move objects forward, backward and stop? Comment by Worrell, Tracy E CIV USARMY ACC (USA): The Autoclear specs call out a specific quantity for Autoimage recall, maybe we can use that as a minimum standard Autoimage recall must be 10 previously scanned images or more or something like that
· System Power: 220 VAC +/- 10%
Technical
· Steel Penetration: Must be at least 28 mm
· Wire Resolution: Must be at least 38 AWG
X-Ray Generator
· Duty Cycle: 100%
· Dose per Exam: Must be less than 0.1 mR per hour
· Radiation Leakage: Must be less than 0.5 mR per hour