This document contains detailed technical specifications for a Theta Flow Optical Tensiometer used in scientific research. The device offers multiple experimental measurement capabilities, including static and dynamic contact angle measurements, surface/interfacial tension analysis, topography profiles, and surface free energy calculations using various scientific methods. Key technical specifications include a CMOS 1" USB 3.0 digital camera with 5MP resolution, maximum measuring speed of 3422 fps, and an integrated touch display.
The tensiometer requires a computer with a 2 GHz processor, 2 GB RAM, 120 GB hard drive (SSD recommended), and Windows 10 operating system. It operates within specific environmental conditions: ambient temperature of 15-30°C, pressure of 700-1060 hPa, and humidity of 20-80% non-condensing. The device has integrated sensors for monitoring temperature, humidity, and system levelness, and uses OneAttension software that includes all measurement modes. Physical specifications include dimensions of 765 x 230 x 435 mm, weight of 29 kg, and input voltage range of 100-240 VAC.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| NDAA 889 Certification.pdf | ||
| USDA NDAA 889 Certification Form 2020.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Specifications for Theta Flow Optical Tensiometer with required parts for research:
Specifications:
Available measurements
Experiments available: Static contact angle, advancing and receding contact angles, batch contact angle, meniscus contact angle, surface / interfacial tension by pendant drop method, topography profiles, surface free energy and dilatational interfacial rheology by oscillating drop method.
Surface free energy methods: Zisman Plot, OWRK/Extended Fowkes, van Oss Acid-Base, Wu, Neumann’s Eq. of State, Schultz 1 and 2
Software and hardware
Measuring range (°, mN/m): 0…180, 0.01…2000
Accuracy (°, mN/m): ± 0.1, ± 0.01
Maximum sample size (mm): Unlimited x 100 x 320 (w. stage).
Integrated sample holders: Yes
Maximum resolution (pixels): 2592 x 2048 (5MP)
Maximum measuring speed (fps): 3422
Camera: CMOS 1’’ USB 3.0 digital camera with zoom.
Image focusing: Software-controlled autofocus, manual fine focus in optics
Image quality: Enhanced with DropletPlus technology
Camera protection: Inside instrument covers
Camera view angle (°): -4…2.5, with digital scale
Light source and size: High power monochromatic LED, 62 x 62 mm.
Field of view (diagonal in mm): 1.44…45.3
Measurement indicator LED: Yes
Integrated touch display Yes
Environment monitoring: Integrated digital ambient temperature, relative humidity and system levelness sensors
Disposable tip dispensing: Yes
Software: OneAttension, includes all measurement modes
Dimensions – Basic frame (LxWxH, mm): 765 x 230 x 435
Weight – Basic frame (kg): 29
Input voltage (Vac): 100…240
Frequency (Hz): 50…60
Input current (mA): 330…650
Fuses: T1.0A 5x20 mm
System requirements
Recommended system requirements for Theta Flow: 2 GHz processor, 2 GB RAM, 120 GB hard disk drive*, 1920x1080 resolution, 1 USB 3.0 port
*SDD hard disk (min. 500MB/s) needed for high speed recording with high resolution.
Operating system requirements: Windows 10 (32 or 64 bit).
Environmental conditions
Ambient temperature (°C): 15…30
Ambient pressure (hPa): 700…1060
Ambient humidity (%): 20-80 (non-condensing)