Specifications_Sheet_MRM-7650_Sampler.pdf
PDF 56 KB Posted
- Attached to
- P25-1527 H70 H2 Fuel NVR Sampler
- Solicitation number
- P25-1527
- Issued by
- California
About this file
Summary of Specifications Sheet MRM-7650 Sampler
This is a specifications sheet for a field sampler device related to California contract opportunity P25-1527 for H70 H2 Fuel NVR Sampler equipment. The document details the technical requirements for a specialized sampler designed to collect particulate samples at dispensers in an alternative fuels laboratory setting. The sampler must meet or exceed the standards established in ASTM D7650 for filter sampling recommendations and comply with ASTM D7651 specifications. The equipment is intended to enable precise measurement and analysis of particulates in hydrogen fuel applications, supporting the state's alternative fuels testing and development infrastructure.
The specifications sheet establishes the engineering and performance criteria that potential vendors must satisfy in submitting proposals for this sampling equipment. The document serves as the technical foundation for the procurement, ensuring that any awarded vendor delivers equipment meeting established industry standards and laboratory testing requirements for alternative fuel quality assurance.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| RFQ_SB_Option_P25-1527.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Specifications Sheet MRM-7650 Hydrogen Fuel Sampler
• Fuel sampler H70 rated 10,000+ psig for Non-Volatile Residue (NVR) particulate matter testing by ASTM D7650 sampling method and ASTM D7651 measuring method.
• A Mass Flow Meter pressure rated should be 15,000 psig with Totalizer. It should also have a rechargeable battery powered controller for precise measurement of Kilogram Hydrogen Sampled.
• All check valves should have a cracking pressure of 10 psig. A check valve system on Vent for 9/16” Cone and Thread.
• The filter housing should rate 10,000 psig with a SAE adaptor fitted for a connection to the MRM-7650 sampler.
• Portable vent system for hydrogen H2 sampling. A spring assembly that extends the venting hydrogen to more than 10 ft vertically with a 30 ft certified hydrogen H2 venting hose that will allow the system to extend beyond overhands.
• A sturdy Pelican Shipping/Transport case with wheels and retractable handle.