Specifications PTR-MS.docx

DOCX document 14 KB Posted

Attached to
Proton Transfer Reaction Mass Spectrometer Federal contract opportunity
Solicitation number
12405B20Q0096
Issued by
Department of Agriculture Agricultural Research Service Field Research Implementation and Information Delivery

About this file

This document contains specifications for a federal procurement of a Proton Transfer Reaction Mass Spectrometer. The United States Department of Agriculture Agricultural Research Service is seeking to procure a Proton Transfer Reaction Mass Spectrometer that meets or exceeds the minimum specifications provided, including an inlet system with a flow rate between 50-250 ml/min, hydronium ion source, time of flight detection system with 1500 m/delta m resolution for m/z above 79, pulse frequency up to 80 kHz, sensitivity greater than 500 cps/ppbv for benzene at 1500 m/delta m resolution, and detection limit below 10 pptv for benzene at 1 minute. Offerors must provide a quote on company letterhead by September 9, 2020 with pricing, delivery terms to Gainesville, Florida by 60 days after receipt of order, and evidence of SAM registration. Award will be made based on best overall value determined by technical capability, past performance, and price.

View the file

Other files for this federal contract opportunity

Other files attached to Proton Transfer Reaction Mass Spectrometer, newest first.
File Type Posted
Commercial Items Provisions - Clauses.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Minimum Specifications for the Proton Transfer Reaction Mass Spectrometer

Configuration specifications:

· Instrument: Proton Transfer Reaction Mass Spectrometer

· Inlet system: inlet flow rate f 50 ml – 250ml/min

· Ion Source: Hydronium ions

· Ion Purity: > 99%

· Detection System: Time of Flight (positive ion mode)

· Detection Time resolution <=27ps (37GHz)

Performance Specifications

· Mass resolution (typical): > 1500m/ delta m (FWHM) for m/z > 79

· Pulse frequency: up to 80 kHz

· Sensitivity: m/z (Benzene): typical >500 cps/ppbv @ 1500 m/delta m (FWHM)

· Detection limit: m/z 79 (benzene) < 10 pptv (1min)

Delivery: FOB Destination. Deliver to the USDA-ARS, 1600 SW 23rd Drive, Gainesville, FL 32608. The Government requires delivery on or before 60 ARO, however, each offeror shall include their proposed delivery schedule as part of their quotation.

Quoted Price: any delivery costs as well as any applicable duties, brokerage, or customs fees.

File details come from the government source that posted it. Updated .