SPECIFICATIONS
Portable x-ray system Item: Portable x-ray system
1 Product Features
Name of Item: Portable x-ray system
1. Capabilities (if applicable): General:
1.1. Portable x-ray system for taking radiography for fishes and other aquatic organisms.
1.2. The x-ray system will be used to determine the location of implanted tags and inspect specimens for spinal and skeletal deformities.
1.3. The x-ray system will be used in both a field and laboratory setting, so must be portable and capable of being powered by battery.
1.4. Components of the portable x-ray shall include:
1.4.1. Portable x-ray generator
1.4.2. Wireless direct radiography (DR) panel
1.4.3. Patient Image Acquisition System (PACS)
1.4.4. Accessories for operation, safety, transport, and storage
2. Portable battery-powered x-ray generator – technical specifications
2.1. X-Ray filament output: 20 - 10 mA (current in milliAmperes, mA) @ 40 - 90 kVDC, equivalent to range of 2.0-10 mAs (milliAmperes seconds, mAs)
2.2. Light beam collimator with integrated adjustable dual laser pointers
2.3. Power: rechargeable lithium ion battery
2.4. Weight: < 25 lbs with battery
2.5. Additional components: battery charger, exposure switch and cord, and spare lithium ion battery
2.6. Warranty for generator: at least 1 year (1 year for batteries)
3. Wireless Direct Radiography (DR) cassette style flat panel – technical specifications
3.1. Scintillator: Direct Deposit Cesium Iodide (Csl)
3.2. Pixel Pitch: 100-154 μm
3.3. Pixels: 2816 X 2304 (approximately 6.5 million)
3.4. Image Size: 14" x 17" (35 x 43 cm)
3.5. X-ray Voltage Range: 40-150 kVp
3.6. A/D Conversion: 16 bits
3.7. Output Data: 16 bit grayscale
3.8. Wireless: Built in Access Point (AP) with 25' range.
3.9. Data Transfer: 802.11 a/g/n/ac/ Wireless LAN, up to 867 Mbps, 1Gbps Ethernet
3.10. Preview Image Access Time: Approximately 4-5 seconds after exposure
3.11. Exposure Control: Full Field Automatic Exposure Detection or manual
3.12. Weight: <10 lb.
3.13. Power (Voltage/Frequency): AC 100-240 V 50/60 Hz
3.14. Batteries and Charger: Two to Three (2-3) Lithium Ion rechargeable batteries & mobile battery charger
3.15. Standard Components: Sensor unit, interface box, cables
3.16. Panel Housing Material: Carbon Fiber Front (needs to be shock, vibration, and scratch resistant) and High Strength Aluminum Alloy Back
3.17. Protective enclosure for DR Panel
3.18. Warranty for panel: 5 years (batteries 1 year)
4. Patient Image Acquisition System (PACS) – technical specifications
4.1. Imaging Software Package
4.1.1. Image processing to includes gradation/equalization processing and grayscale print management
4.1.2. Zoom, PAN, magnifier, ROI, crop, rotate, mirror, etc.
4.1.3. Send, Image CD/ Flash Drive Burner
4.1.4. Image output files available in DICOM, JPEG, TIFF, BMP, etc.
4.1.5. Measuring of distances, angles, areas and density
4.1.6. Insertion of image annotations, e.g. free texts, arrows, ellipses etc.
4.2. Mobile workstation
4.2.1. Laptop, rugged laptop, or rugged tablet with the following minimum specifications:
4.2.1.1. i5 Intel Core
4.2.1.2. Windows 10 Professional
4.2.1.3. 500 GB Hard Disk
4.2.1.4. 6-8 GB RAM
4.2.1.5. Workstation pre-configured and with the imaging software installed
5. Accessories for operation, safety, and storage
5.1. Protective carrying case for x-ray generator
5.2. Protective carrying case for wireless DR panel
5.3. Tripod for x-ray generator
5.4. Protective apron
6. Vendor Training on System Operation
6.1. Vendor training on system operation
2. Delivery
Prices quoted must be FOB Destination
3. Delivery Location:
Product shall be delivered to the following location.
Total price includes all applicable shipping fees, the Government is tax exempt.
U.S. Fish and Wildlife Service Abernathy Fish Technology Center 1440 Abernathy Creek Road Longview, Washington 98632
4. Installation:
Prices quoted must include installation. All labor and parts associated with and required for installation are the sole responsibility of the contractor. No assistance or equipment will be provided by the government. Installation shall be completed according to industry standards.
5. Technical Point of Contact (POC):
Doug Peterson Doug_peterson@fws.gov 406-465-4076