This is a technical specification document for a Hemispherical Electron Analyzer for Angle-resolved Photoelectron Spectroscopy (ARPES) system under Code 6812. The system comprises three main components: a hemispherical electron analyzer, sample manipulator, and ultrahigh vacuum chamber.

The hemispherical electron analyzer must have a minimum 135mm mean radius, 30mm working distance, and 0.5-1500 eV energy range with high throughput capabilities. Key performance requirements include transmission and angular lens modes with ±15° and ±10° acceptance respectively, angular resolution <0.1° from 0.1mm spot, >850 energy channels, >650 angular channels, and 3 meV FWHM energy resolution. The system requires a minimum 40mm multichannel plate detector with digital camera capabilities (1 Gbit/s readout, >7bit resolution, >120 fps), mu-metal housing bakeable to 150°C, and computer operation using a PC from non-Chinese manufacturers (excluding Lenovo, Huawei, or Apple) with Windows-based software for XPS, UPS, and AES experiments. The sample manipulator must provide 5-axis motorized control (x,y,z,θ,Φ) for 90mm diameter sample holders accommodating 75mm wafers, with specified travel ranges and electrical isolation capabilities. The ultrahigh vacuum chamber requires mu-metal or stainless steel construction with mu-metal shielding, bakeable to ≥150°C, leak rate ≤5x10⁻¹⁰ mTorr, and multiple ports for various analytical instruments and connections to existing Code 6812 equipment. The system must operate standalone due to NRL firewall restrictions preventing remote access.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Specifications for Code 6812 Hemispherical Electron Analyzer for Angle-resolved Photoelectron Spectroscopy (ARPES) system (analyzer, sample manipulator, and ultrahigh vacuum chamber)

1. Hemispherical electron analyzer

a. Minimum 135mm mean radius

b. Minimum working distance 30 mm

c. Energy range: 0.5 eV – 1 500 eV

d. High throughput

e. Transmission and Angular lens modes of operation

i. Transmission mode lens acceptance: ±15°

ii. Angle modes up to ±10°

iii. Angular resolution: < 0.1° from 0.1 mm spot

iv. Energy channels: > 850

v. Angular Channels: > 650

f. Electrostatic deflection capability

i. Energy resolution (trans. and ang. mode): 3 meV FWHM at 2 eV pass energy

ii. Pass energies ranging from 2 eV to 100 eV

g. Mu-metal housing

i. Bakeable to 150 °C

ii. Magnetic fields < 0.5 uT

h. Minimum 40mm multichannel plate detector

i. Digital camera detector

ii. 1 Gbit/s readout rate

iii. > 7bit resolution

iv. < 0.01 cps noise per channel

v. > 120 frames per second (TV frame rate)

vi. Pulse and ADC acquisition modes

i. Electronics

i. 110/120 VAC 60Hz

ii. Stable/low noise

iii. Mountable in 19” cabinet

iv. 25m cabling lengths

j. Computer and Software

i. The hemispherical analyzer must be computer operated using a PC from a non-Chinese manufacturer (no Lenovo, Huawei, or Apple)

ii. Software must be provided to perform XPS, UPS, AES experiments in standard transmission mode and in angle-resolved mode

iii. The operating system must be Windows-based

iv. Data should be stored in CasaXPS format for manipulation and analysis

v. Data should be exportable in an acceptable form for publication (Excel, ASCII, JPEG, PDF)

2. Sample Manipulator

a. 5-axis (x,y,z,θ,Φ) of motion

b. Motorized control

c. Sample stage for 90mm diameter sample holder (of a 75mm wafer)

i. Electrical isolation for grounding or biasing of sample

ii. BNC feedthrough for electrical connection

d. x,y travel >25mm

e. z travel >200mm

f. θ travel up to 60° depending on chamber geometry

g. Φ travel +/- 180 ° allowing for analysis of every point on a 75mm wafer

h. Bakeable to 150 °C

3. Ultrahigh vacuum chamber

a. Manufactured from mu-metal or stainless steel with mu-metal shielding for all internal surfaces to reject magnetic fields ≤ 0.5uT using only UHV compatible manufacturing techniques.

b. Bakeable to ≥ 150 °C

c. Leak rate ≤ 5x10-10 mTorr

d. Ports for:

i. Connection to Code 6812 preparation chamber (8” CF)

ii. Connection to Code 6812 vacuum pumps (8” CF)

iii. Hemispherical analyzer

iv. Monochromatic x-ray source (4.5” CF)

v. Dual anode x-ray source (2.75” CF)

vi. Ultraviolet photon source (2.75” CF)

vii. Small spot electron gun (4.5” CF)

viii. Secondary electron detector (2.75” CF)

ix. Sputter ion source (2.75” CF)

x. Electron flood gun (2.75” CF)

xi. Motorized sample manipulator

xii. Sample transfer manipulator (2.75” CF)

xiii. Vacuum suitcase (6” CF)

xiv. Camera viewport (2.75” CF)

xv. Transfer viewport (≥ 4.5” CF)

xvi. Lighting viewport (2.75” CF)

xvii. Extra ports for pumping and other needs

e. Integration with existing vacuum equipment, surface analysis techniques, and sample transfer mechanisms (vacuum cart and vacuum suitcase).

4. Software Operation

a. NRL firewalls prevent “remote access” to computers and instruments, so the equipment should/must be capable of “standing alone”.