This document outlines specifications for an automated slide imaging system procurement. The Systematic Entomology Laboratory is seeking to acquire a GLISSANDOTM 20SL Digital Pathology Scanner or equal for rapid imaging of slide-mounted insect specimens. Minimum requirements include an all-in-one polarizing plus brightfield slide scanner capable of imaging one 50mm x 75mm slide or two 25mm slides at 0.5 microns per pixel resolution. The system must be able to automatically detect tissues, generate high resolution z-stacked images, and produce focus-stacked images of complete specimens. Related to this is federal contract opportunity number 12305B23Q0140 for an Automated Slide Scanner being procured by the Department of Agriculture Agricultural Research Service Field Research Implementation and Information Delivery Northeast Area.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Terms and Conditions.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Specifica ons
Systema c Entomology Laboratory (SEL) is acquiring an automated slide imaging system for rapid imaging of slide‐mounted insect specimens. This equipment will support mission cri cal research and cura on projects.
This is a Brand Name or Equal item. The item provided must be as specified or meet the salient characteris cs specified. Please see FAR Provision 52.211‐6 Brand Name or Equal, AGAR Provision
452.211‐70 Brand Name or Equal, and AGAR Clause 452.211‐71 Equal Products Offered in the Contract
Terms and Condi ons a achment.
1) Descrip on: An automated slide imaging system for rapid imaging of slide‐mounted insect specimens.
2) Make/Model: GLISSANDO 20SL DIGITAL PATHOLOGY SCANNER FOR 20 SLIDES.
3) Manufacturer: Objec ve Imaging Ltd., Unit 1, Quy Court, Colliers Lane, Stow cum Quy, Cambridge, CB25 9AU, United Kingdom.
4) Required salient characteris cs:
Minimum requirements:
An all‐in‐one polarizing plus brightfield slide scanner/computer designed for either one 50mm x
75mm slides or two 25mm x 75mm histological slides or two 25mm x 50mm geological slides
Plan Fluo 10x/0.40 objective for generating images, magnifications from 0.1x to 400x
Fast tiling technology with predictive focus and automatic region definition, sample detection and focus setup
Plan Fluor 10x objective lens, 0.5 microns per pixel; Optional Plan Apo 20x/0.75 or Plan Fluor
40x/0.75 objectives
Color Camera Sensor Size 2464 x 2056 pixels with sensor image resolution 0.25 μm (with 20X objective)
Barcode Reader image‐based 1D and 2D (Datamatrix, QR Code)
Scan times for a 15x15 mm area using fast Tiling is 60 seconds
Integrated control computer, Windows 10 mini‐PC with Ethernet connection.
On‐board Storage 500 GB solid state drive
Upgrade capacity with 20 specimen slide loader
Tes ng and Verifica on
The slide imaging system will be tested by SEL scien sts upon arrival to verify the system operates as adver sed, providing high resolu on images of slide mounted insect specimens. To meet expecta ons, the system must be able to (1) automa cally detect ssues on the slide, (2) automa cally generate high resolu on z‐stacked images of the complete specimen, and (3) produce a focus stacked image of the complete specimen. This will be tested on mul ple slides to determine proper func onality.
Data Requirements
All manuals required for instruc on on proper opera on of the slide imaging system are to be provided.
Packaging and Shipping
The system must be securely packaged and shipped so that it arrives undamaged and in working order to the specified USDA des na on.