TITLE: 110 GHz Vector Network Analyzer

Requesting Laboratory / Division / Group:

Communications Technology Laboratory Radio Frequency Technology Division Radio Frequency Electronics Group 672.01

General Statement of Need:

NIST is developing a millimeter-wave standard reference material. To deliver this new standard, NIST must measure the scattering (S-) parameters of cavity resonators from 1 GHz up to 110 GHz. NIST must measure 4 different types of resonators: split post dielectric resonators, split cylinder resonators, Fabry Perot open resonators, and balanced circular disk resonators. The split cylinder resonators, Fabry Perot open resonators, and balanced circular disk resonators require proprietary software by Keysight.

Items and Ancillary Services:

Item 1: 110 GHz Vector Network Analyzer Quantity: 1 Specifications: 1.1 All necessary power cables for installation

1.2 All necessary connectors for installation

1.3 Rackmount kit with handles for installation

1.4 Low frequency limit of 1 kHz

1.5 High frequency limit without extender heads 67 GHz

1.6 High frequency limit with extender heads 110 GHz

1.7 Maximum power output at 67 GHz must be greater than 0 dBm

1.8 Minimum IF bandwidth 10 Hz

1.9 Maximum IF bandwidth 1 kHz

1.10 4 ports cable of measuring a-waves and b-waves simultaneously and independently

1.11 Dual sources cables of delivering 0 dBm at all ports at all frequencies

1.12 1.85 mm NMD connectors on each port

1.13 Front end access jumpers

1.14 Ability to drive 3rd party extender heads

1.15 Direct IF access

1.16 Embedded LO, frequency-offset options

1.17 An internal combiner to route two or more sources to a single port

1.18 Rear access loops to combiners to allow external sources and external filters

1.19 Internal bias tees

1.20 Demonstrated measurement capabilities with N1501AE11/67

Balanced Type Circular Disk Resonator (BCDR) and Resonant Cavity Method – Option 007 for

N1501AKEAD710/20/24/28/35/40/50/60/80

Service Requirements / Item Specifications:

Section Requirements/Specifications Planning Considerations ☒ N/A

☐ List any visits required to allow potential vendors to see site specific constraints.

[Explain Requirements.] ☐ US citizen required on-site (foreign nationals require pre-registration 30 days prior to visit.)

☐ Vendor Representative On-Site more than three (3) days.

☐ Other Considerations:

[List other considerations to ensure an accurate quote.] Shipping & Delivery ☐ N/A

Shipping Address:

☒ NIST-Boulder

325 Broadway Boulder, CO 80305

☐ [Alternate Shipping Address] Shipping Criteria:

☐ Partial Delivery Acceptable

☐ [Enter Shipping and Delivery information by item.] ☒ Direct Delivery to Building Required Installation of system in Building 81 required

☐ Building has Loading Dock Containerization Preference: [Enter Container Type] Other Requirements:

[List Any Other Requirement] Delivery Date Criteria:

☒ Delivery of Goods, Installation, and Training shall be competed no later than: 11/30/2023 ☐ Delivery No Earlier Than: Click or tap to enter a date.

☐ Other:

[List Other Shipping, Delivery or Special Requirements.] Electronic Media ☐ N/A

[Enter Electronic Media Requirements by Item.]

☒ Software ☐ Electronic Manuals

☒ License Key ☐ Other:

[List Any Other Electronic Media Requirements] ☐ Email address for Software Delivery:

[Enter Address] Installation ☐ N/A

[Enter Installation Requirements by Item]

☒ Vendor Technician on Site ☐ Rigging ☐ Uncrating / Unpackaging ☐ Removal of Packaging Material ☒ Equipment Set Up ☐ Start-Up Services ☐ Turnkey Installation ☐ Other:

[Enter Details]

NOTE:

When installation is required, TPOC shall be the coordination point.

Facility / Utility Considerations ☒ N/A

[List Site Specific Utility Considerations by Item] ☐ OFPM Work Order Numbers

[List Work Order Numbers] ☐ Power Available Voltage: 120 (V) Maximum Current: 20 (A) Phase: [Choose Phase] Frequency: [Typically 60Hz] (Hz) NEMA Plug Type: [Enter Plug Type] (i.e.: NEMA L5-20P) Environment: [Choose Environment] Other Power Considerations:

[List Any Other Power Considerations.] ☐ Utility Connections:

Air drop ☐ Other Considerations:

[List Other Considerations to Ensure an Accurate Quote.] Warranty ☐ N/A

Base Manufacturer Warranty is acceptable

☒ Warranty Length: 1 Year What is included?

☒ Labor ☒ Parts ☒ Travel ☒ Cost and Liability for Returns ☒ Other [Enter Other Requirements] ☒ 1 Option Period ☒ 2 Option Periods ☒ 3 Option Periods ☒ 4 Option Periods (Maximum)

Calibration ☐ N/A

[Describe Calibration Requirements Per Item] Calibration Plan Length: 1-Year What is Included?

☒ Labor ☒ Spare Parts ☒ Consumables

☐ Travel ☐ Cost and Liability for Shipment ☐ Software Updates ☐ Manuals

☐ Other [Enter Other Requirements] ☐ Requirements:

[Enter Requirements] (i.e.: On-Site Within 72 hours, etc.)

☒ 1 Option Period

Length: [i.e. 1 – Year / 6 - Months] ☒ 2 Option Periods ☒ 3 Option Periods ☒ 4 Option Periods (Maximum)

Training ☒ N/A

☐ On-Site

Is Travel Required: NO

☐ Off-Site Location: [Enter Location(s)] ☐ Number of Trainees: 3 Scope of Training ☐ Operations ☐ Maintenance ☐ Troubleshooting ☐ Safety ☐ Other Alignment and calibration of stage and interferometers ☐ Training materials to be provided:

☐ Other

NOTE: When training is required, TPOC shall be coordinator.

Inspection and Acceptance

☒ On-Site

☐ N/A [Inspection Requirements Per Item] ☐ Government Acceptance Expected to Not to Exceed 7 Business Days ☒ Government Acceptance Expected to Exceed 7 Business Days 30 Days ☐ Inspection and Acceptance Plan:

[Explanation] ☐ Other:

[Enter Other Requirements]

Maintenance ☐ N/A

[Describe Maintenance Requirements Per Item] Service Plan Length: 1-Year

What is Included?

☐ Labor ☐ Spare Parts ☐ Consumables ☐ Travel

☐ Cost and Liability for Returns ☐ Software Updates ☐ Manuals l

☐ Other [Enter Other Requirements] ☐ Requirements:

[Enter Requirements] (i.e.: On-Site Within 72 hours, etc.)

☒ 1 Option Period (Base + One) ☒ 2 Option Periods (Base + Two) ☒ 3 Option Periods (Base + Three) ☒ 4 Option Periods (Base + Four) Government Furnished Material / Equipment ☒ N/A

[List Any Government Furnished Material / Equipment Per Item] ☐ Samples for Testing

☐ Government Equipment which will leave Government Site (TPOC must coordinate with Division Property Custodian)

Description: [Enter Description] Serial Number: [Enter Serial Number]

NIST Property Number: [Enter NIST Property Number] ☐ Other:

[Enter Other Details] Travel ☒ N/A

[Describe any Travel Requirements Per Item] ☐ Explain:

[Enter Justification / Requirement] ☐ Other:

[Enter Other Requirements] NOTE TO VENDOR: To the maximum extent possible, include travel requirement estimates as part of pertinent sections above (Installation, Maintenance, Repair, etc.). Otherwise, Travel will be a separate line item, reimbursable per the Federal Travel Regulation.