This document is a Specification for a Plant Growth Chamber, solicited by the USDA Agricultural Research Service (ARS) Cereal Disease Lab in Saint Paul, Minnesota. The specification details requirements for a highly sophisticated plant growth chamber with precise environmental control capabilities, including specific dimensional constraints (exterior dimensions not exceeding 79" height x 106" width x 35.5" depth), comprehensive controller features with touchscreen interface, and advanced environmental parameter controls. Key technical requirements include temperature range of 4-35°C, LED lighting optimized for plant growth with at least 1450 µmoles/m²/s intensity, dehumidification capabilities down to 45% relative humidity, and integrated systems for automatic watering, airflow management, and data monitoring.

The procurement involves delivering and installing a single Plant Growth Chamber to the USDA ARS Cereal Disease Lab, with delivery scheduled 130 days after award. The specification mandates a minimum two-year manufacturer warranty and requires delivery, installation, and training. The delivery point of contact is Karen Broz, and the chamber will be delivered to the garage area where a forklift is available. The chamber must have specific features like two reach-in doors with locks, an observation window, interior electrical receptacle, temperature limit safety controls, remote alarm contacts, and comprehensive technical and operator manuals.

View the file

Other files

File Type Posted
Questions and Answers for 1232SA25Q0948.docx DOCX document
Terms and Conditions.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

SPECIFICATION

1.0 Specifications Description:

Plant Growth Chamber

Quantity: 1

Unit Type: Each

Minimum Requirements:

• Exterior dimensions not to exceed: height 79 inches, depth 35.5 inches (must be able to fit through a 35.75 inch doorway), width 106 inches.

• Interior dimensions: one tier with at least 94 cubic ft, and at least 16 ft2 work area. Tier dimensions: at least 72 inches wide, 31.5 inches deep.

• Interior specifications: walls with durable white reflective coating. Stainless steel floor with integrated floor drain, anodized aluminum work surface.

• Maximum growing height: at least 56 inches, preferably greater.

• Water-cooled refrigeration condensing unit.

• Electrical: direct wired 120/208/1/60 (4-wire, plus ground)

• Adjustable legs/ casters for leveling unit.

• Controller requirements: Graphical user interface with touchscreen. Precise control of at least 7 environmental parameters simultaneously. Display of chamber programs and parameters with charting/ graphing of data. Alarm notifications. Must be able to control chamber temperature, lighting, dehumidification, automatic watering, internal electrical outlet. Must have ability to store and recall programs. Diurnal and multi-step program options.

• Temperature requirements: temperature range at least 4 – 35°C with lights off and 10 – 35 °C with lights on, ±0.5 °C within the work area as measured on a horizontal plane with lights on, programmable and integrated with/ controlled by the unit’s control system.

• Lighting requirements: LEDs optimized for plant growth with enhanced/ added red and far-red intensity of at least 1450 µmoles/m2/s 6 inches from LEDs, dimmable to at least 10-100% of output, programmable and integrated with/ controlled by the unit’s control system. Full spectrum light meter for measuring light output.

• Height-adjustable (vertically adjustable) lighting lamp bank.

• Dehumidification requirements: dehumidification down to at least 45% relative humidity ±10% between 15 – 30 °C with lights on or off, with electronic relative humidity sensor, programmable and integrated with/ controlled by the unit’s control system.

• Airflow movement upwards through entire work surface through holes in channels, and with adjustable fresh air inlet(s) and exhaust(s) ports.

• Two reach-in doors with locks that allow access to the entire chamber with tight seal to door frame.

• Observation window with light-tight cover in at least one of the reach-in doors.

• Interior electrical receptacle, at least 1 amp, programmable and integrated with/ controlled by the unit’s control system

• At least two access ports with air-tight plugs for equipment.

• Temperature limit safety controls with audible alarms and visible indicators that shut down chamber power when alarm limit is reached, and restart chamber when temperature is within normal range.

• Remote alarm contacts for building automation system monitoring.

• Automatic watering system: programmable and integrated with/ controlled by the unit’s control system.

• Technical and operator manuals to be included.

• Must include delivery, installation, and training.

2.0 Place of Delivery

USDA ARS Cereal Disease Lab

ATTN: Karen Broz

1551 Lindig St.

Saint Paul, MN 55108

Deliver to Inside the building and install.

3.0 Delivery Schedule

Item QTY Unit Place of Delivery Delivery Date Plant Growth Chamber 1 Each Garage area, we have access to a forklift

130 Days After Award

4.0 Delivery Point of Contact

Karen Broz, karen.broz@usda.gov (612)626-2065

5.0 Additional Requirements

A Manufacturer Warranty of at least 2 years is required for the Plant Growth Chamber.