This document is a Specification for a Plant Growth Chamber, solicited by the USDA Agricultural Research Service (ARS) Cereal Disease Lab in Saint Paul, Minnesota. The specification details requirements for a highly sophisticated plant growth chamber with precise environmental control capabilities, including specific dimensional constraints (exterior dimensions not exceeding 79" height x 106" width x 35.5" depth), comprehensive controller features with touchscreen interface, and advanced environmental parameter controls. Key technical requirements include temperature range of 4-35°C, LED lighting optimized for plant growth with at least 1450 µmoles/m²/s intensity, dehumidification capabilities down to 45% relative humidity, and integrated systems for automatic watering, airflow management, and data monitoring.
The procurement involves delivering and installing a single Plant Growth Chamber to the USDA ARS Cereal Disease Lab, with delivery scheduled 130 days after award. The specification mandates a minimum two-year manufacturer warranty and requires delivery, installation, and training. The delivery point of contact is Karen Broz, and the chamber will be delivered to the garage area where a forklift is available. The chamber must have specific features like two reach-in doors with locks, an observation window, interior electrical receptacle, temperature limit safety controls, remote alarm contacts, and comprehensive technical and operator manuals.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Questions and Answers for 1232SA25Q0948.docx | DOCX document | |
| Terms and Conditions.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
SPECIFICATION
1.0 Specifications Description:
Plant Growth Chamber
Quantity: 1
Unit Type: Each
Minimum Requirements:
• Exterior dimensions not to exceed: height 79 inches, depth 35.5 inches (must be able to fit through a 35.75 inch doorway), width 106 inches.
• Interior dimensions: one tier with at least 94 cubic ft, and at least 16 ft2 work area. Tier dimensions: at least 72 inches wide, 31.5 inches deep.
• Interior specifications: walls with durable white reflective coating. Stainless steel floor with integrated floor drain, anodized aluminum work surface.
• Maximum growing height: at least 56 inches, preferably greater.
• Water-cooled refrigeration condensing unit.
• Electrical: direct wired 120/208/1/60 (4-wire, plus ground)
• Adjustable legs/ casters for leveling unit.
• Controller requirements: Graphical user interface with touchscreen. Precise control of at least 7 environmental parameters simultaneously. Display of chamber programs and parameters with charting/ graphing of data. Alarm notifications. Must be able to control chamber temperature, lighting, dehumidification, automatic watering, internal electrical outlet. Must have ability to store and recall programs. Diurnal and multi-step program options.
• Temperature requirements: temperature range at least 4 – 35°C with lights off and 10 – 35 °C with lights on, ±0.5 °C within the work area as measured on a horizontal plane with lights on, programmable and integrated with/ controlled by the unit’s control system.
• Lighting requirements: LEDs optimized for plant growth with enhanced/ added red and far-red intensity of at least 1450 µmoles/m2/s 6 inches from LEDs, dimmable to at least 10-100% of output, programmable and integrated with/ controlled by the unit’s control system. Full spectrum light meter for measuring light output.
• Height-adjustable (vertically adjustable) lighting lamp bank.
• Dehumidification requirements: dehumidification down to at least 45% relative humidity ±10% between 15 – 30 °C with lights on or off, with electronic relative humidity sensor, programmable and integrated with/ controlled by the unit’s control system.
• Airflow movement upwards through entire work surface through holes in channels, and with adjustable fresh air inlet(s) and exhaust(s) ports.
• Two reach-in doors with locks that allow access to the entire chamber with tight seal to door frame.
• Observation window with light-tight cover in at least one of the reach-in doors.
• Interior electrical receptacle, at least 1 amp, programmable and integrated with/ controlled by the unit’s control system
• At least two access ports with air-tight plugs for equipment.
• Temperature limit safety controls with audible alarms and visible indicators that shut down chamber power when alarm limit is reached, and restart chamber when temperature is within normal range.
• Remote alarm contacts for building automation system monitoring.
• Automatic watering system: programmable and integrated with/ controlled by the unit’s control system.
• Technical and operator manuals to be included.
• Must include delivery, installation, and training.
2.0 Place of Delivery
USDA ARS Cereal Disease Lab
ATTN: Karen Broz
1551 Lindig St.
Saint Paul, MN 55108
Deliver to Inside the building and install.
3.0 Delivery Schedule
Item QTY Unit Place of Delivery Delivery Date Plant Growth Chamber 1 Each Garage area, we have access to a forklift
130 Days After Award
4.0 Delivery Point of Contact
Karen Broz, karen.broz@usda.gov (612)626-2065
5.0 Additional Requirements
A Manufacturer Warranty of at least 2 years is required for the Plant Growth Chamber.
mailto:karen.broz@usda.gov