Conviron model PGC-FLEX Plant Growth Chamber, including installlation, as per specifications
2 1 ea LED Light intensity approximately 1300 micromoles/m²/s on top tier and 750 micromoles/m²/s on middle tier measured at a distance of 6”. Broadband spectrum for plant growth using Fluence PhysioSpec Indoor LED fixtures and supplemental far red LED. Broad spectrum is controlled by closed loop dimming. Far red is controlled by open loop dimming. Temperature range becomes +10°C to +35°C lights ON. The LI-
190R replaces the traditional Quantum sensor, providing improved performance, high stability and greater sensitivity
SNH Spray Nozzle Humidification: Range: Up to
90% RH lights OFF and 75% RH lights ON, limited by a +25°C maximum dew point (empty chamber).
Additive humidity through use of spray nozzles.
Control: ±3% RH. System uses a dry humidity sensor to directly measure humidity in
%RH. Spray nozzles require 60 psi (4.2 bar) pressure and must be supplied with reverse osmosis water.
Water resistivity must be between 0.01 to 0.02 Meg ohm-cm (MΩ- cm). Maximum water usage to maintain specified levels is 4 litres/hr. If the above water pressure is not available, the CPSNH option is required to supply necessary pressure
Provide misters for all three tiers. Misting infrastructure shall be able to be easily attached/detached to each of the three growth tiers.
Tubing and mister materials shall consider UV degradation. Connect to a solenoid for operation via the existing CMP6060
Please refer to the attached Technical Proposal
TOTAL PRICE
© 2020 by Controlled Environments Limited. All rights reserved. Specifications Subject To Change Without Notice; E & OE
28-May-20 info@conviron
.com www.conviro n.com
Controlled
Environments Inc. #2 -
572 S 5th Street, PO Box
347 Pembina, ND, 58271, USA tel: +1.701.280.9635 toll free: +1.800.363.6451 fax: +1.204.786.7736
TO: NO: CP21047-F04
QUOTATION & TECHNICAL
SPECIFICATIONS CONVIRON MODEL PGC-
FLEX GROWTH CHAMBER
1.0 Control System - Conviron CMP6060:
1.1 General1: The CMP6060 v6.0 control system is a robust industrial grade control system that is designed specifically for plant growth chambers and rooms. ts intuitive color touch screen interface and limitless scalability make it a reliable, flexible and user friendly system.
1.2 Programming: The CMP6060 is capable of storing up to 16 real-time user-entered programs each containing up to 48 lines with one minute resolution. Each program defines a single 24-hour period. Programs are then entered into a "Schedule" (a sequence of up to 8 program entries) to create a single day, multi day, or seasonal schedule. Programs are entered via a high resolution color touch-screen display.
1.3 Process Control Mode: All chamber process transitions occur in STEP mode, with the exception of temperature, which can be set to either STEP or global ramp
(gRAMP) mode.
1.4 Data Logging and Graphing:
1.4.1 Local Data: The data available at the local display screen includes up to seven chamber variables such as temperature, humidity, lighting and CO2.
Parameters are all logged automatically every 30 seconds and can be viewed in the form of Trend graphs. Trend graphs can be viewed, with up to five days of history, directly on the local display screen for both 'set-point' and 'actual' values. Zoom and Pan Functions provide additional visualization tools allowing the User to precisely and accurately view and access the data. A cursor can be used to display the value and the setpoint at the cursor location.
mailto:info@conviron.com http://www.conviron.com/
© 2020 by Controlled Environments Limited. All rights reserved. Specifications Subject To Change Without Notice; E & OE
1.4.2 The data available remotely is more substantial and includes every input and output parameter being monitored and controlled. Data is automatically logged and log rate can be set in multiples of 6 seconds regardless of the number of parameters. This provides unprecedented data resolution for the
User while it also benefits service personnel. Log data is accessed in several ways:
• A network-connected PC (Note: Requires WEBCARD option on GEN
Series products, sold separately)
• An internet-enabled PC (Note: Requires WEBCARD option on GEN
Series products, sold separately)
• Conviron's Central Management™ System (sold separately)
1.4.3 Trend data can also be downloaded through the USB port of the chamber touch screen. Chamber data archived on the touch screen is logged every 30 seconds with up to fifteen days of history.
1 All performance control tolerances including temperature, humidity, CO2 are based on steady state conditions. Environmental conditions will fluctuate during programmed transitions between set points, and return to control tolerances once steady-state conditions resume.
CONVIRON REFERENCE: NO. CP21047-F04
MODEL PGC-FLEX PLANT GROWTH CHAMBER
© 2020 by Controlled Environments Limited. All rights reserved. Specifications Subject To Change Without Notice; E & OE
1.5 Alarms: Audible and color coded visual alarms are provided. User-settable absolute alarms for all controlled processes and User-settable process tracking alarms for temperature are available for the User to determine how tightly they want to monitor the experiment. n addition, there are also "built in" alarms available for chamber component monitoring. Alarm history is logged to daily text files easily retrievable from any Ethernet connected computer (either with or without a Central
Management™) or through the local control screen (last 15 alarm events).
1.6 Security: Password-protected levels are reserved for users, administrators and service personnel. The CMP is capable of storing up to 50 unique names and passwords for user and administrator profiles.
1.7 Start-Up Delay: Chambers can be programmed with varying restart times to reduce start-up loads after a power failure.
1.8 Communications: Ethernet connectivity is delivered with every controller.
Controller, mask and Host P address are easily settable from the Status screen. The controller is shipped communications-ready.
1.9 Auxiliary and Central Alarm Contacts: Alarm contacts for connection to a building management system or (optional) auto dialer.
1.10 The chamber local display can be mirrored on Conviron's Central Management™
System (sold separately) allowing remote support, training and access to all the local chamber functions remotely.
1.11 Lighting Run Times: The controller tracks the amount of time (in hours) each lighting output has been ON. The run time of each output can be viewed and reset using the local display.
2.0 Construction2: (Note: All dimensions are nominal.)
2.1 Exterior Dimensions: 100"W x 35½"D x 101"H (2540mmW x 900mmD x 2565mmH)
(106"H with service cover opened.)
2.2 nterior Dimensions: 97"W x 30½"D (2465mmW x 775mmD)
2.3 Growth Area: Single tier mode: 19ft2 (1.8m2)
Two tier mode: 38ft2 (3.6m2)
2.4 Growth Height: Single tier mode: 59¾" (1520mm)
Two tier mode: 38ft2 (3.6m2)
2.5 Fresh Air: Control is manually adjustable.
2.6 Aluminum channel floor providing uniform upward air flow.
2.7 Cabinet Construction: Aluminum paneling using CFC-free polystyrene insulation.
2.8 Exterior Finish: Blue-green enamel baked on patterned aluminum.
2.9 nterior Finish: Reflective white enamel baked on smooth aluminum on front wall, doors and back wall. End walls have reflective specular aluminum finish.
2.10 Condensing Unit Access: Compressor and refrigeration components located on top
© 2020 by Controlled Environments Limited. All rights reserved. Specifications Subject To Change Without Notice; E & OE front of chamber.
2.11 Doors: Two reach-in doors with keyed magnetic locks.
2.12 Observation Window; Two (2) dual pane windows with light tight cover.
2.13 Shelf: Corrosion resistant wire adjustable on ½" (12mm) centers.
2.14 Control Panel: Center mounted.
2.15 nstrument Ports: Two ports, 2" (50mm) with light tight caps.
2.16 Anchoring: Due to high center of gravity, unit is secured to floor using concrete anchors.
2 Local Regulatory Codes: Equipment design and manufacture may not address all local, state or country-specific regulatory requirements including elements such as seismic anchoring, electrical configuration, or similar. Client to identify and advise of such requirements prior to issuance of a purchase order.
© 2020 by Controlled Environments Limited. All rights reserved. Specifications Subject To Change Without Notice; E & OE
2.17 Unit is pre-wired for three (3) tier operation and can be modified for horizontal airflow.
2.18 Lifting Brackets: Two (2) hand brackets provided to facilitate lifting of machine compartment. One set shall be provided to accommodate multiple chambers. Note: A lifting frame is available (optional) for installation using fork lift. Consult factory.
2.19 Factory assembled, tested and fully crated. Machine compartment is then removed for crating and shipping. Requires re-installation on site.
2.20 Two Tier Operation: Option includes four drain pans which double as irrigation trays for convenient watering of plants, two for the additional tier and two for the bottom
Unifloor® shelf. Two shelves are included for the second tier to support the drain pans. ncludes 8 shelf clips and tubing to drain.
3.0 Lighting:
3.1 ntensity3: Two independent side-by-side canopies make up the second tier and provide a maximum light intensity of 500 micromoles/m2/s each. n the single tier configuration mode the chamber has a maximum light output of approximately 1125 micromoles/m2/s.
3.2 Programming and Control: 5 level programming of lamps in single tier configuration
(3 level in two tier and three tier configuration).
3.3 Lamps: Balanced spectrum for plant growth using T5 fluorescent lamps and halogen incandescent lamps evenly spaced.
3.4 Lamp Fixture: Height adjustable. The removable canopies are cantilevered from the back wall and adjust on 6" (150mm) increments.
3.5 Lamp Heat: Removed by refrigeration system.
3.6 Ballasts: High efficiency electronic.
3.7 Light Meter: Quantum light meter for display and recording of light output, displayed through user interface.
4.0 Temperature Control: (Maximum design ambient temperature is +35°C)
4.1 Range: +4°C to +40°C lights OFF, +10°C to +40°C lights ON. (Full fresh air.)
4.2 Control4: ±0.5°C, at control point.
4.3 Temperature Safety Limits:
Primary: A programmable min and max temperature or a limit tracking alarm that automatically follows the programmed set point.
Secondary: An independent factory-set high and low temperature limit is also provided for increased assurance.
Audible alarms are included based on user-defined limits.
4.4 Aspirator: Located in the growth area for accurate controlling and recording of chamber conditions.
4.5 Fan Speed Control: Variable fan speed control on conditioning unit, from 50% to maximum.
© 2020 by Controlled Environments Limited. All rights reserved. Specifications Subject To Change Without Notice; E & OE
5.0 Refrigeration:
5.1 Provisions for Remote Outdoor Air-Cooled Condenser: Remote location (up to 50'
[15m] combined horizontal and vertical distance) of condenser only (not included).
Compressor, receiver and other refrigeration components remain in cabinet machine compartment.
5.2 Valve: Electronic 3-way modulating valve that regulates the temperature function of the chamber.
5.3 Heat Exchanger Coil(s): Copper-tube construction.
3 Light measurement at 6" (150mm), chamber temperature of 25°C. Light intensities are nominal values measured at the rated chamber supply voltage and (dimmable lighting) with a maximum light output (verified by a L -190R factory reference
Quantum Sensor). With fluorescent lamps, light intensity values may be lower than specified if measured at temperature set points below 25°C.
4 Measured by N ST traceable calibrated temperature sensor without test materials or optional accessories
© 2020 by Controlled Environments Limited. All rights reserved. Specifications Subject To Change Without Notice; E & OE
5.4 Refrigerant: Refrigeration system is charged with CFC-free refrigerant.
5.5 Monitoring:
Refrigeration system operation is monitored by the control system, including visual and audible alarm.
Pressure transducers allow for real-time diagnostics for preventative maintenance and repair.
6.0 Air Flow:
6.1 Vertical: Optimized air flow, provided by centrifugal impellers, is directed to growth area uniformly upward through Unifloor® less than 50ft/min (15.2m/min). Note:
airflow changes from vertical to horizontal with the removal of back wall plugs.
6.2 Fresh Air: ndividual adjustment of positively sealed inlet and outlet from open
20ft3/min (0.57m3/min) to closed.
7.0 Humidity Control: (Optional)
7.1 Range: No control on basic unit.
8.0 Carbon Dioxide Additive Control: (Optional)
8.1 Range: No control on basic unit.
9.0 Utility Requirements5: (rating increases with some options.)
9.1 Electrical Service: (Alternative services available, consult factory) 60Hz: 120/208-3 -60Hz-4 wire plus ground
9.2 Drain: Floor drain must be provided outside footprint of cabinet.
9.3 Provided with a U.S. electrical inspection by an inspection authority accredited by OSHA.
10.0 Warranty and After Sales Service:
10.1 Warranty and after sales service is handled by our dedicated team of service professionals. They provide a 24-hour manned service phone to facilitate any repair emergency. n addition, the team is available, during regular business hours, to assist with any service or warranty repairs, replacement parts or questions. Conviron also offers a fully trained international network of independent third party service providers who can affect repairs wherever your chamber is located.
10.1.1 Two year warranty (extended warranty optional).
10.2 To access this free support, please contact the service group at;
Phone: (204) 786-6451 Fax: (204) 786-7736 Toll Free: 1-800-363-6451
Email info@conviron.com www.conviron.com
11.0 Installation:
The following specifications outline the typical responsibilities of the parties as they pertain to installation of Conviron chambers, rooms and central management
(CM).
11.1 Conviron to be responsible for/provide the following:
11.1.1 Transport the equipment from the off-loading or storage area to the assembly area.
© 2020 by Controlled Environments Limited. All rights reserved. Specifications Subject To Change Without Notice; E & OE
11.1.2 Uncrate equipment and dispose of crating material in customer supplied disposal bins.
11.1.3 Assemble and make work the supplied equipment using qualified personnel.
11.1.4 Start-up and check equipment performance and operation to factory specifications.
5 This unit will tolerate ±10% voltage fluctuation from the rated voltage on the serial plate. A voltage stabilizer must be used if the fluctuation is greater than ±10%. Failure to do so can result in serious damage to the compressor and electronic components and will void warranty. Disconnect switch must be sized by a local qualified electrician.
© 2020 by Controlled Environments Limited. All rights reserved. Specifications Subject To Change Without Notice; E & OE
a) Verification and function testing of building power, all input devices, all controlled devices, all non controlled devices, refrigeration and safety components and if applicable all special control sequences and devices; and
b) A 24-hour test to demonstrate performance of all controlled parameters to their specified limits.
11.1.5 Provide four hours of instruction on use and maintenance of the equipment.
Additional instruction and/or training available by agreement.
11.1.6 nstallation nstruction: Electronically provided to customer upon project completion. Printed copy available (optional), consult factory.
11.2 Customer to be responsible for/provide the following:
11.2.1 Receive and off-load equipment. f there will be a delay between the receipt and building utility connections of the equipment, the customer is responsible for providing adequate and secure storage. Customer must provide forklift type equipment to remove equipment from common carrier highway transport truck/s.
11.2.2 The customer must prepare a clean and unobstructed route, including sufficient clearance through doorways, openings, elevators, etc. between the off-loading or storage area and the location of chamber assembly.
11.2.3 The customer must ensure that installers have adequate on-site storage and staging areas. f no such space is available, Conviron may arrange for storage containers at the customer's expense.
11.2.4 Supply and connection of building services to the equipment in accordance with local codes, restrictions and legislation. These services must be fully functional and sized in accordance with the equipment specifications in a timely manner. The customer shall supply and connect all building services to the equipment, including any cutting, penetrations, or patching required.
Building services may include some or all of the following:
-Electrical service (more than one may be required)
-Condenser water (supply and return)
-Purified water for humidity system
-Compressed air for humidity system
-Fresh air and/or exhaust air connections
-Carbon dioxide supply
-Floor drain connections
-Conduit or cable trays or other special requirements for Ethernet support
(applicable to CM only)
11.2.5 Equipment must be installed on level floors. The customer shall ensure that all floors are level to within ½ inch.
© 2020 by Controlled Environments Limited. All rights reserved. Specifications Subject To Change Without Notice; E & OE
11.2.6 The quotation for installation services assumes that all on-site responsibilities of the customer have been fulfilled upon Conviron commencing the installation such that Conviron can complete the installation in a single mobilization. This includes the provision of all building services, operating as specified. Should site delays result in Conviron having to suspend its installation work, the customer shall reimburse Conviron for all downtime, including the cost of return visits and associated expenses. Expenses shall be charged on a cost plus 15% basis, and time shall be charged at Conviron's standard installation rates.
11.3 General: Conviron understands there is no requirement to provide unionized labor in the performance of its work. As such, Conviron may supply non-union or unionized trade labor at its sole discretion.
Page
© 2020 by Controlled Environments Limited. All rights reserved. Specifications Subject To Change Without Notice; E & OE
11.4 Local Regulatory Codes: No allowances have been made in the equipment design or manufacture to consider local regulatory codes including seismic requirements.
Building code requirements for your area should be investigated prior to equipment installation.
11.5 Conviron provides installation services with the understanding that said services are to install and make work, within a given period of time, the Conviron supplied equipment. n no way does Conviron guarantee, or imply to guarantee, to supply a minimum number of man-hours and/or installation personnel at any time during the process of this installation.
ISO 9001 Registered
CONFIDENTIAL
Controlled Environments Limited, Controlled Environments nc., Conviron Europe Ltd., Conviron Science and Technology (Shanghai) Co. Ltd., Conviron Asia Pacific PTY Ltd.
and/or Argus Control Systems Ltd. (herein referred to collectively as "Conviron") claims full ownership, property, design and marketing rights to the information and material herein disclosed. t is supplied in confidence and any reproduction, fabrication or re-communication of any nature or kind of the contents without the express prior written consent of Conviron is strictly prohibited.
Page
© 2020 by Controlled Environments Limited. All rights reserved. Specifications Subject To Change Without Notice; E & OE
$ 69,800.00
$ 12,550.00
$ 4,156.00
Pricing Information
Pricing for one (x1) only Conviron model PGC-FLEX Plant Growth Chamber, including installlation, as per specifications;
F.O.B. Destination: Salinas, CA
LED Light intensity approximately 1300 micromoles/m2/s on top tier and 750 micromoles/m2/s on middle tier measured at a distance of 6". Broadband spectrum for plant growth using Fluence
PhysioSpec ndoor LED fixtures and supplemental far red LED.
Broad spectrum is controlled by closed loop dimming. Far red is controlled by open loop dimming. Temperature range becomes
+10°C to +35°C lights ON. The L -190R replaces the traditional
Quantum sensor, providing improved performance, high stability and greater sensitivity
SNH Spray Nozzle Humidification: Range: Up to 90% RH lights OFF and 75% RH lights ON, limited by a +25°C maximum dew point (empty chamber). Additive humidity through use of spray nozzles. Control: ±3% RH. System uses a dry humidity sensor to directly measure humidity in %RH. Spray nozzles require 60 psi
(4.2 bar) pressure and must be supplied with reverse osmosis water. Water resistivity must be between 0.01 to 0.02 Meg ohm-cm (MO-cm). Maximum water usage to maintain specified levels is 4 litres/hr. f the above water pressure is not available the
CPSNH option is required to supply necessary pressure.
*M ST Provide misters for all three tiers. Misting infrastructure shall be able to be easily attached/detached to each of the three growth tiers. Tubing and mister materials shall consider UV degradation.
Connect to a solenoid for operation via the existing CMP6060.
Extended Warranty Options:
Conviron offers extended warranties on all its products. Extended warranties must be purchased at the time of the original product purchase to be eligible for the discounted warranty price shown. Extended Warranties may be purchased at a later date for a non-discounted price provided it is within the Standard Warranty period.
For Extended Warranty pricing, please contact Conviron. Further details regarding
Page
© 2020 by Controlled Environments Limited. All rights reserved. Specifications Subject To Change Without Notice; E & OE
Conviron's Standard Warranty and Extended Warranty are available on Conviron's website at www.conviron.com, or by calling 1-800 363-6451 or 1-204-786-6451.
Preventative Maintenance & Service Program - GroGuardian™:
Conviron offers comprehensive preventative maintenance and service programs for its equipment. Program frequency and structure are customized based on your requirements and desired level of service. To learn more about GroGuardian™, please contact us at info@conviron.com
Page
© 2020 by Controlled Environments Limited. All rights reserved. Specifications Subject To Change Without Notice; E & OE
6 Lift Gate Availability: Customer supplied forklift required for lift gate assistance of crated dimensions greater than 6' (1.8m) in length or 2,000lbs (907kg). Lift gate not available with full trailer load delivery.
| Spec 1 |
| Spec 2 |
File details come from the government source that posted it. Updated .