SOW Lasers.docx
DOCX document 45 KB Posted
- Attached to
- 38 Various Govt. -owned Lase Service Maintenance Federal contract opportunity
- Solicitation number
- HT941025Q2013
- Issued by
- Defense Health Agency
About this file
This Statement of Work (SOW) outlines requirements for service and maintenance of 38 government-owned laser units at Naval Medical Center San Diego (NMCSD), with a base period of January 15, 2024 - January 14, 2025 plus four one-year options through January 14, 2029. The Defense Health Agency seeks a contractor to provide all labor, parts, materials, and equipment necessary for service and calibration of various laser systems including carbon dioxide, diode, dye, alexandrite, and Nd:YAG lasers used in ophthalmology, dermatology, and surgical applications.
Key requirements include performing calibrations twice per year (May and November), maintaining 95% equipment uptime, responding to service calls within 24 hours by phone and 2 business days on-site, and providing detailed field service reports within 72 hours of service completion. The contract is a 100% small business set-aside with firm fixed pricing. Contractor must complete Naval Base San Diego access requirements and coordinate with NMCSD Biomedical Repair Department. Quotes are due by 10:00am PDT on December 30, 2024 and must be submitted electronically to dernell.w.wade.civ@health.mil. Questions are due by December 23, 2024.
View the file
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
STATEMENT OF WORK
SERVICE AND MAINTENANCE
1. Scope: The contractor shall provide all services, labor, parts, materials, and equipment necessary for the service and maintenance for the government owned laser units as listed on Appendix A, "LABEL FOR APPENDIX" at Naval Medical Center San Diego (NMCSD).
Contractor shall service and maintain the equipment within the original equipment manufacturers (OEM) specifications, in accordance with the US, State, and Local laws and regulations, US Navy regulations, and Joint Commission requirements.
2. Definitions: As used throughout this work statement, the following terms shall have the meaning set forth below:
2.1. Service and maintenance: (a) any modification, adjustment, or replacement of the equipment that corrects a malfunction and effectively brings the equipment into material conformity with the technical specification for that equipment as designated by the original equipment manufacturer (OEM), (b) a procedure or routine that, when observed in the regular, proper operation of the equipment, avoids the material adverse effect of the applicable nonconformity, (c) any measures performed that assure the continued operation of the equipment or returns the equipment to its designed efficiency and capacity, (d) replacement of a piece of equipment on a one-for-one basis in order to provide a substitute of equal or greater value in lieu of a damaged or defective device for compensation, return, or refund of equipment.
2.2. Contracting Officer (KO): A warranted representative of the federal government who has the express authority to enter, administer, and/or terminate contracts and make related determinations and findings. The KO is the only person authorized to approve changes to any terms, conditions, or requirements of the contract.
2.3. Technical Representative (TR): A Government employee appointed by the KO and responsible for the inspection, evaluation, and acceptance of products and services provided in accordance with the terms and conditions of the contract.
3. Location:
Naval Medical Center San Diego 34800 Bob Wilson Drive San Diego, CA 92134-5000
Biomedical Repair Department Building One, Ground Floor
And/or:
Ophthalmology Department Building Two, Second Floor
Dermatology Clinic Building two, third floor
Otorhinolaryngology Clinic Building two, second floor
4. Requirements:
4.1. General Performance Requirements:
: Biomedical Equipment Repairs test measurement and diagnostic equipment’s (TMDE) annual inspection, Calibration. Consist of 167 pieces of equipment that is vital to keeping Naval Medical Center San Diego as they are used to service High risk and Low risk medical equipment and allowing to follow the Joint Commission and ensure the patient safety.
4.1.1. The Contractor shall provide equipment maintenance services to include all periodic planned maintenance and unscheduled repairs in accordance with OEM specifications and in compliance with Beacon Point standards, to include repairs necessitated by operator error.
4.1.2. The Contractor may use third party service vendors not limited to the Original Equipment Manufacturer (OEM), however, these vendors must be approved in advance by the KO and TR. The Contractor must submit a list of any sub-contractors they intend to use as third-party service vendors to the KO and TR for approval. Third party service vendors shall not perform any work under this contract prior to such approval.
4.1.3. Ensure that only fully qualified field engineers and/or technicians who have gone through OEM or equivalent third-party service schools for the equipment shall be employed in the performance of all work performed under this contract. Service personnel shall have sufficient experience to demonstrate proficiency in maintenance of the product. Upon request, Contractor shall provide written assurance of the competency their personnel as well as a list of credentials for approved field service personnel. The KO and/or TR may authenticate the training requirements and request training certificates or credentials from the service vendors at any time.
4.1.4. Test equipment: Test equipment calibration shall be traceable to National Institutes of Standards and Technology (NIST traceable). Any efforts conducted without equipment within the calibration intervals is not performance in accordance with this contract. Upon request by TR, the Contractor and/or service vendors shall provide a copy of the current calibration certification of all test equipment to be used, or which has been used, on NMCSD equipment.
4.1.5. Field Service Reports: Contractor shall furnish a written or electronic vendor supplied Field Service Report (FSR) to the Duty Staff at NMCSD Biomed. All FSR's shall be submitted within 72 hours of completion of service and shall be sent to the following email address:
usn.nmcsd-contractadministrator@health.mil
The contractor or their representative shall complete the FSR to include, at a minimum, the following information:
1. Contractor Name
2. Technician's printed name, telephone number, and signature
3. Date and Time of Arrival
4. ECN (to be provided by Navy) and Serial Number of Equipment
5. Time Expended Repairing/Service: Labor Hours, Rate, Materials
6. Summary of Work Performed and Accepted by End-User, with Government Representative's Printed name and Signature) Completed Field Service Reports are required prior to acceptance of any invoice. Failure to submit a Field Service Report to NMCSD Biomed can result in delay of payment or rejection of invoice from Government. It is highly recommended that the contractor attach a copy of the Field Service Report to their Wide Area Workflow Invoice in addition to the required submission to NMCSD Biomed in order to expedite invoicing process.
4.2. NMCSD Preventive Maintenance Requirements:
4.2.1. Preventive maintenance shall be performed in a manner which will minimize corrective maintenance and will include but is not limited to: calibration, adjustment, lubrication, updating software for security, safety, or system functionality issues, testing required to maintain a safe environment of care, electrical safety testing, output testing, replacement of batteries and other shelf life components on a scheduled basis (which becomes corrective maintenance if the component fails due to lack of preventive maintenance).
4.2.4. Calibrations shall be performed at a minimum of 2 time per year, with these scheduled calibrations (May and NOV) to be coordinated with NMCSD BIOMED after contract award.
4.2.5. Calibration Completion: Contractor must ensure that 95% or more of all periodic maintenance is performed within 30 days of due date, and that 100% of periodic maintenance is performed within 60 days of the due date.
4.2.6. Equipment in Use: If, for reasons beyond the Contractor’s control, equipment is not made available for maintenance for the entire 30 days before the maintenance due date, and 60 days following due date, the Contractor shall immediately notify the COR.
4.2.7. Equipment Unable to Locate (UTL): If the Contractor is unable to locate an item within the maintenance window specified, the Contractor shall immediately notify the TR. The TR may determine that the property has been disposed of or otherwise inactive or unavailable for use and may adjust the equipment covered under this Statement of Work accordingly.
4.2.8. 100% Calibration: Under the circumstances of 4.2.6. and/or 4.2.7., the TR and Contractor shall arrange a means of complying with the requirements for preventive maintenance at no additional cost to the government. Equipment determined missing, lost, or stolen shall be exempt from maintenance.
4.2.9. Uptime Rate: The Contractor shall maintain an overall uptime rate of 95% or better for all equipment covered by this effort. Individual uptime rate for specific equipment, if required, will be as specified on Appendix A.
4.2.10. Uptime Calculation: Uptime is calculated on a rolling 6-month basis. Until the Contractor has at least 6 months of performance history with NMCSD, the uptime shall be calculated cumulatively for the entire duration of the contract.
4.3. NMCSD Corrective Maintenance Requirements:
4.3.1. Not required
4.3.3. Equipment improvements/modifications shall be made only upon BIOMED written approval and direction.
This includes software updates to the device. All updates/improvements/modifications performed shall be notated on the field service report or equivalent.
4.3.4. Notify BIOMED immediately upon receipt of OEM safety recall notices or upon receipt of recall notices pertaining to any parts or materials utilized by contractor on the equipment.
4.3.5. Ensure that the original design and functional capabilities shall not be changed, modified, or altered without express written authorization from BIOMED.
4.3.6. Provide suitable OEM recommended repair equipment/tools required for the satisfactory execution of all repairs made.
4.3.7. Furnish OEM approved lubricants and lubricate wear points within the equipment in accordance with OEM guidelines.
4.3.8. Replacement parts shall be covered by the contractor for the duration of the contract. Extend too the Government all commercial warranties on replacement parts. All replacement parts or supplies utilized shall be new parts or OEM refurbished parts. OEM refurbished parts shall only be accepted upon written request from the contractor to NMCSD Biomed and shall only be utilized/installed upon written approval from NMCSD Biomed to the contractor.
4.4. End-User Requirements:
4.4.1. Service shall be performed on all devices included under contract, regardless of the nature of the cause warranting service/maintenance. Contractor shall not refuse any device under this contract submitted for services.
Contractor will perform an official inspection of the device, make an estimate for the services/parts required and the time required for the repairs and an official determination of the cause of the issue, and will provide the required parts and labor to restore the equipment to function within the OEM specifications.
4.4.2. Services shall be unlimited for the devices under the contract for the term of the agreement. There shall be no cap or limit placed upon the number of repairs/replacements on any individual unit, or on the facility.
4.4.3. Contractor shall perform Calibration on the (289) Biomed TMDE units at least 2 times per year, with a minimum of 6 months between preventive maintenance inspections. Contractor shall perform preventive maintenance as per the schedule on Appendix B.
4.4.4. Contractor shall keep department personnel informed as to the status/condition of equipment during PM and the repair status/condition when equipment is down or out for repair. Points of contact for the department are:
Biomedical Repair Contracts team at 619-532-8017 or usn.nmcsd-contractadministrator@health.mil.
CATASTROPHIC DAMAGE CLAUSE
4.4.5. If equipment is received and is irreparably damaged, the department requires a written report as to the extensiveness of the damage and the suspected cause or type of maltreatment necessary to cause such damage (e.g., dropped, crushed mechanically, overloaded, etc.). The field service report may serve in this regard if comprehensive.
4.5. Response Time and Labor:
4.5.1. Contractor shall use commercially reasonable efforts to respond by telephone to any report of a malfunction requiring repair within (24) hours of notification from NMCSD Biomed, Monday through Friday, 0800-1600, Pacific Standard Time. The Contractor shall also provide telephone consulting services, twenty-four (24) hours per day, seven (7) days a week. This consulting service shall be provided to avoid Delays in contract service coverage during after-hours and weekend/holiday periods as applicable. Under no circumstances shall the Contractor assume an inquiry, via telephone or any other means, is authorization to provide services not specifically covered by the terms of the contract.
4.5.2. Contractor shall use commercially reasonable efforts to provide on-site support within two (2) business days from notification by NMCSD Biomed, unless otherwise noted in Section 4.4.
4.5.3. Work required in the performance of this contract shall be performed between the hours of 0800 and 1700, Pacific Standard Time. All compensation for labor from the hours of 0800 to 1700 Pacific Standard Time is provided for in this contract and included in the contract price. Contractor may be permitted to arrange for the performance outside of these hours via coordination with NMCSD Biomed at no additional-cost-to-the-government.
4.5.4. Any work that would require overtime must be approved, in writing, in advance by the KO.
5. Travel and Related Expenses: Contractor shall be responsible for its service-related trip expenses, including round trip travel, mileage, and overnight living expenses.
6. Shipment: The contractor shall not exceed two (2) business days for shipment of all equipment coming from and/or going to NMCSD Biomed minus exceptions or conditions as previously noted in this Statement of Work. All shipping expenses/costs are provided for in this contract and included in the contract price.
7. Liability. N/A
8. Utilities, Facilities, and Equipment Access:
8.1. The contractor may use Government utilities (e.g., electrical power, compressed air, water) that are available and required for any service performed under this contract while at Government facilities. The facilities engineer however, must approve of any contractor equipment which would connect to these utilities to ensure compatibility with NMCSD utilities.
8.2. Any outages required to perform services must be approved by NMCSD Facilities Management prior to initiation of work and shall be performed per Facilities' instruction. Contractor may request Facilities assistance in performing these outages if required.
8.3. Ensure that all use, storage, transport of any materials regulated by OSHA as "hazardous", required for the performance of this contract, is in in compliance with all federal, state, and local regulations. Contractor shall have available MSDS's for any such materials on-site.
8.4. The Contractor shall protect all facilities from damage caused by the Contractor and shall repair, at no additional cost to the Government, all damages caused by the Contractor's activities. During performance of services, Contractor shall keep the site safe, neat, clean and orderly. Contractor shall be responsible for all means, methods, techniques, sequences, and procedures of the services
8.5. Access to Equipment: Contractor shall be provided reasonable access to all equipment that is to be serviced and utility outlets required to do the service. The contractor shall be free to start/stop all primary equipment incidentals to the operation of the maintained equipment after permission is received from on-duty personnel responsible for such equipment.
9. Security and Safety Requirements:
9.1. The Contractor is required to submit/complete Naval Base San Diego Base access requirements for access to NMCSD. Contractor shall contact Naval Base San Diego Security Department at 619-556-1653/5336 or at https://www.cnic.navy.mil/regions/cnrsw/installations/navbase_san_diego/om/force_protection/pass_and_decal.html in order to obtain the necessary clearances to perform this contract. All costs for completing the requirements for access are the responsibility of the contractor.
While waiting for authorized access, contractor shall, at a minimum, contact NMCSD Biomed not later than 1400 on the day prior to a scheduled visit to submit the required information to complete the NMCSD Temporary Security Pass. In case of emergent repairs or service, the contractor shall wait at the NMCSD Main Gate for escort from Biomed personnel.
9.2. Identification, Parking, and Smoking Regulations at NMCSD: Field service personnel shall wear visible photo identification at all times while on the premises of NMCSD. It is the responsibility of the field service personnel to park in the appropriate designated parking areas. Information on parking is available from NMCSD Security personnel. Smoking is prohibited inside any buildings of NMCSD and is restricted to designated smoking locations only. Possession of weapons, alcohol, or pharmaceuticals that are not over the counter and not prescribed to the individual/family is prohibited on the premises. Enclosed containers, including tool kits, are subject to search.
Violations of regulations may result in citation answerable in the United States (Federal) District Court. Violation of any policy which results in lost time for Contractor or his sub-contractors is the sole responsibility of the Contractor, and shall not be charged to the government. Contractor employees shall demonstrate appropriate behavior and dress at all times.
9.3. Contractor Check-In/Check-Out: Contractor is required to report to Material Management Department, Biomedical Repair Division at Building One (1), Ground Floor, Room GD-18H1 at NMCSD prior to and upon completion of any services or repair work performed.
9.4. General Safety Requirements: The Contractor shall exercise safety precautions as the TR may determine to be reasonably necessary to protect the lives and health of occupants of buildings in which work is being performed.
NMCSD will notify the Contractor of any non-compliance with the foregoing conditions and the action(s) to be taken. Service vendors shall, after receipt of such notice, immediately correct the conditions to which attention has been directed. If field service engineer refuses to comply promptly, the TR may issue an order stopping all or any part of the work. Under such circumstances, the KO and hold the service vendor in contract default. Nothing in this section will be construed so as to limit the Contractor or his sub-contractors responsibilities to comply with any existing law, nor to switch jurisdictions.
9.5. Government Right of Refusal: NMCSD reserves the right to reject any of the contractor's personnel and refuse them permission to enter the premises or work on NMCSD equipment. This right may be exercised at any time. The reasons for rejections may include: lack of training and experience, but may also include but not be limited to:
behavior, attire, security concerns, immigration status.
10. Compliance with Applicable Laws: Contractor shall be knowledgeable of and comply with all applicable Federal, State, and local laws, regulations, and requirements regarding the conduct of his business. The Contractors personnel shall comply with the DoD Information Assurance Policies applicable. Contractors are required to maintain a Drug Free Workplace, as well as to require any sub-contractor to do the same. Contract shall require wage determinations and comply with applicable Federal and State worker's compensation and occupational disease statutes.
11. Modifications to Contract: Only the KO has the authority to obligate the government, and then only in writing.
No modification or instruction from any party shall be construed to authorize the Contractor to take any action for which they expect compensation without written authorization from the KO.
12. Invoicing: All invoicing shall be done in accordance with this statement of work and in accordance with contract. Failure to do so may delay or interfere with payment for services rendered.
13. NMCSD Points of Contact: For all matters concerning performance of this contract, the primary point of contact shall be:
Biomedical Repair Division, Material Management Department Building One, Ground Floor, Room GD-18H1 Naval Medical Center San Diego 34800 Bob Wilson Drive San Diego, CA 92134
TEL: 619-532-8017
Fax: 619-532-8013 Email: usn.nmcsd-contractadministrator@health.mil For administrative matters, the contractor may contact the contracting officer as designated in blocks 7.a, 7.b, 8, and 9 as found on the Standard Form 1449 issued for the contract, or may contact:
Material Management Department Naval Medical Center San Diego 34800 Bob Wilson Drive San Diego, CA 92134
TEL: 619-532-8110
FAX: 619-532-5596
| Ecn |
| Mfr. Serial No. |
| Nameplate Model |
| Nomenclature |
| Manufacturer |
| 115041 |
| 114 |
| KT-1005890 |
| LASER DELIVERY SYSTEM, WAVEGUIDE |
| LUMENIS INC |
| 152309 |
| 991490359910 |
| 9914-00-9035 |
| LASER, ALEXANDRITE, DERMATOLOGIC |
| CANDELA CORP |
| 166262 |
| RGXR00 |
| UG-1000324 |
| LASER, ARGON, DERMATOLOGIC |
| LUMENIS INC |
| 136019 |
| 1000 |
| GA-0000530 |
| LASER, CARBON DIOXIDE |
| LUMENIS INC |
| 37199 |
| 010-55938 |
| UP ENCORE |
| LASER, CARBON DIOXIDE, SURGICAL/DERMATOLOGIC |
| LUMENIS INC |
| 101404 |
| 66 |
| UP ENCORE |
| LASER, CARBON DIOXIDE, SURGICAL/DERMATOLOGIC |
| LUMENIS INC |
| 117288 |
| 583 |
| GA-000050 |
| LASER, CARBON DIOXIDE, SURGICAL/DERMATOLOGIC |
| LUMENIS INC |
| 162759 |
| 225 |
| GA-2000000 |
| LASER, CARBON DIOXIDE, SURGICAL/DERMATOLOGIC |
| LUMENIS INC |
| 163081 |
| 1333 |
| GA-0000530 |
| LASER, CARBON DIOXIDE, SURGICAL/DERMATOLOGIC |
| LUMENIS INC |
| 24034 |
| GL14355C30 |
| GLX |
| LASER, DIODE, OPHTHALMIC |
| IRIDEX CORP |
| 102011 |
| GL16791C30 |
| 31118A |
| LASER, DIODE, OPHTHALMIC |
| IRIDEX CORP |
| 118333 |
| 10456 |
| NOVUS VARIA |
| LASER, DIODE, OPHTHALMIC |
| LUMENIS INC |
| 118450 |
| 2559 |
| OME2000 |
| LASER, DIODE, OPHTHALMIC |
| ENDO OPTIKS |
| 163363 |
| B4891 |
| OME200SMAHRVA |
| LASER, DIODE, OPHTHALMIC |
| BEAVER-VISITEC INTERNATIONAL INC |
| 163364 |
| B4890 |
| OME200SMAHRVA |
| LASER, DIODE, OPHTHALMIC |
| BEAVER-VISITEC INTERNATIONAL INC |
| 163365 |
| B4855 |
| OME200SMAHRVA |
| LASER, DIODE, OPHTHALMIC |
| BEAVER-VISITEC INTERNATIONAL INC |
| 163366 |
| B4872 |
| OME200SMAHRVA |
| LASER, DIODE, OPHTHALMIC |
| BEAVER-VISITEC INTERNATIONAL INC |
| 111276 |
| 9914-0300-1248 |
| V BEAM PERFECTA |
| LASER, DYE, DERMATOLOGIC |
| CANDELA CORP |
| 133503 |
| 9914-0300-3206 |
| 9914-08-0300 |
| LASER, DYE, DERMATOLOGIC |
| CANDELA CORP |
| 114710 |
| 14117-76468 |
| 1500-038-20 |
| LASER, ER:YAG, DERMATOLOGIC |
| SCITON INC |
| 130924 |
| 9440 |
| 0030-4864 |
| LASER, EXCIMER |
| ABBOTT MEDICAL OPTICS |
| 151589 |
| 1424 |
| 1100-0001-01 |
| LASER, EXCIMER |
| RA MEDICAL SYSTEMS |
| 140917 |
| 1016-3-409 |
| EX500 |
| LASER, EXCIMER, OPHTHALMIC |
| ALCON INC |
| 140918 |
| 1025-1-1012 |
| FS200 |
| LASER, FEMTOSECOND PULSE |
| ALCON INC |
| 158281 |
| 1217609 |
| 1345-518 |
| LASER, FEMTOSECOND PULSE |
| CARL ZEISS |
| 136125 |
| 151 |
| GA-0006800 |
| LASER, HO:YAG, SURGICAL |
| LUMENIS INC |
| 166662 |
| 10076 |
| GA-0001700DIG |
| LASER, ND:YAG |
| LUMENIS INC |
| 32453 |
| MC62077 |
| MEDLITE C6 |
| LASER, ND:YAG, DERMATOLOGIC |
| HOYA CONBIO |
| 98038 |
| F0972 |
| MC-SYS-SR1500-US |
| LASER, ND:YAG, DERMATOLOGIC |
| RELIANT TECHNOLOGIES |
| 103578 |
| 123 |
| LUMINES ONE |
| LASER, ND:YAG, DERMATOLOGIC |
| LUMENIS INC |
| 111278 |
| 9914-2025-0451 |
| 9914-00-2025 |
| LASER, ND:YAG, DERMATOLOGIC |
| CANDELA CORP |
| 164105 |
| 26710 |
| M22 |
| LASER, ND:YAG, DERMATOLOGIC |
| LUMENIS INC |
| 97820 |
| A1739 |
| 0010-8118 |
| LASER, ND:YAG, OPHTHALMIC |
| IRIDEX CORP |
| 101139 |
| 002-515720 |
| SELECTA II 110V |
| LASER, ND:YAG, OPHTHALMIC |
| LUMENIS INC |
| 130923 |
| 0814-80001 |
| J20007D |
| LASER, ND:YAG, OPHTHALMIC |
| ABBOTT MEDICAL OPTICS |
| 137360 |
| 1155991 |
| 266002-1162-133 |
| LASER, ND:YAG, OPHTHALMIC |
| CARL ZEISS MEDITEC GMBH |
| 24035 |
| 844119 |
| 690 |
| LASER, ND:YAG, SURGICAL |
| CARL ZEISS INC, ZEISS HUMPHREY SYSTEMS |
| 136333 |
| 20511 |
| UG-0005100 |
| LASER, ND:YAG, SURGICAL |
| LUMENIS INC |