This Statement of Work (SOW) details a comprehensive maintenance and service agreement for medical imaging equipment at Lawton Indian Hospital. The contract covers two GE Definium 656 x-ray rooms, two GE Optima 240 portable x-ray machines, and six imaging detectors for a base one-year term with four optional extension years. Service requirements include OEM planned maintenance, routine system calibrations, quality assurance evaluations, unscheduled repair visits with a 6-hour on-site response time, parts replacement, software updates, and technical support with a 30-minute phone response time.
Key performance metrics include a 97% uptime guarantee, normal business hours coverage (M-F, 8AM-5PM CST), and strict reporting requirements. The service provider must submit service reports within five business days, notify the Biomedical Engineering Department before performing services, and sign in/out through facility security. The solicitation is specifically seeking Indian Small Business Economic Enterprises (ISBEEs) under the Buy Indian Act, with responses due by 3:00 PM CST on May 30, 2025, to be submitted via email to the Contract Specialist at the Indian Health Service.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
STATEMENT OF WORK
GE HEALTHCARE Definium 656 and Optima 240 Maintenance and Service Agreement
1. Offeror shall provide a 1 Year comprehensive coverage for maintenance and service of two GE Defiunim 656 rooms, two GE Optima 240 portable x-ray and six imaging detectors with 4 option years at a fixed annual service cost that includes all of the following:
a. OEM Planned Maintenance Service per OEM specifications
b. OEM Routine System Calibration Tests
c. OEM Quality Assurance Evaluations
d. Table repairs
e. Unscheduled repair visits to correct equipment malfunctions, which includes OEM replacement parts, travel, and labor. Unscheduled repairs shall have a 6 hour on-site response time. Tube coverage, detector coverage and battery replacement included.
f. Grid replacement
g. Repair Parts are included, next day delivery by 10:30 am.
h. Software and quality updates to enhance the performance of the equipment as recommended and issued by GE Healthcare
i. TiP Answer Line
j. Offeror shall provide Technical Support to include unlimited calls for customer and service support. Phone support shall have a 30 minute response time.
k. ILINQ Response time 30 MIN
l. 97% Uptime Guarantee
m. Normal business hours coverage M-F 8AM – 5PM (CST) excluding Federal Holidays
2. Offeror shall submit a service report, and checklist if applicable, to the Lawton Service Unit Biomedical Engineering Department within five business days after services rendered.
3. Offeror shall notify the Lawton Indian Hospital Biomedical Engineering Department prior to performing any services.
4. Offeror shall sign in/out at the Security Office upon enter/exiting facility for service.
5. Offeror shall remit and original and two (2) copies of its invoice to:
IHS OKLAHOMA AREA
ATTN: FINANCIAL MANAGEMENT
701 Market Drive
OKLAHOMA CITY, OKLAHOMA 73114
6. Invoice shall include: Amount for services rendered, address where payment is to be sent, Invoice Number, Purchase Order Number, and the date when services where rendered.
7. Acceptable Quality Levels (AQL) & Performance Surveillance: Non-conformance is not acceptable and may result in termination of the contract.
| Performance Objectives |
| Acceptable Quality Level |
| Performance Measure |
| Surveillance |
| OEM Planned Maintenance to OEM Standard |
| Not more than one (1) instance per performance period |
| Annually |
| Nuvolo |
Availability and Response for customer service concerns
| Not more than one (1) instance per performance period |
| 30 minutes on phone response and 6 hour on-site response |
| Nuvolo |
| Timely submittal of service report |
| Not more than one (1) instance per performance period |
| Annually and upon repair within 5 business days to the Biomedical Engineering Department |
| Nuvolo |
| Notification to Bio-Medical department |
| Not more than one (1) instance per performance period |
| Notify Biomedical Engineering department that FSE is on site and preparing to perform maintenance |
| Security log |
| Timely Submittal of Invoices |
| Not more than one (1) instance per performance period |
| Quarterly and upon request |
| UFMS reports |
| Up time Guarantee |
| Not more than one (1) instance per performance period |
| 100-(base time-down time)/100) |
| Equation will be implemented and performed upon equipment failure |
System ID’s:
1. LIHOPTIMA240
2. LIHOPTIMA240B
3. LIHOPTIMA240DET
4. LIHOPTIMA240DETB
5. 580354XR2656
6. 580354XR1565
7. 580354XR2656FP1
8. 580354XR2656FP2
9. 580354XR1656FP1
10. 580354XR1656FP2