SOW and Specifications - Spectroradiometer v2.pdf

PDF 119 KB Posted

Attached to
ABB MR304SC Spectroradiometer Federal contract opportunity
Solicitation number
80TECH25Q0033
Issued by
National Aeronautics and Space Administration

About this file

This Statement of Work details requirements for purchasing a Fourier Transform Infrared (FTIR) spectroradiometer system for NASA Glenn Research Center's Code KA sensing suite. The system must include a spectroradiometer with dual input/output optical ports, Stirling-cooled MCT and InSb IR detectors with specific wavelength responses (4-13.5µm and 1-5.5µm respectively), and various optical components including an input collimator, motor-operated aperture wheel, and 4° FOV telescope.

The system must meet detailed performance specifications including RMS noise-equivalent spectral radiance requirements, scan rates of 10-107 scans/sec at different resolutions, and environmental operating conditions (0-45°C, up to 90% humidity, 0-2km altitude). Physical requirements specify maximum dimensions of 40cm x 40cm x 40cm and weight limit of 35kg for the main unit. The system must include control software with capabilities for simultaneous measurements, real-time recording, and post-acquisition analysis. Delivery is expected within 26 weeks ARO to NASA Glenn Research Center in Cleveland, OH.

View the file

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

STATEMENT OF WORK

Purchasing a Spectroradiometer

1. Objective/Requirements

The requested spectroradiometer will be an addition to the sensing suite available in Code KA.

2. Specifications

2.1. System Configuration

2.1.1. The fourier transform infrared spectroscopy (FTIR) System shall include a spectroradiometer with two input and two output optical ports:

2.1.1.1. One input port shall be configured with a fixed reference cancellation source

2.1.1.2. The second input port shall be configured to accept a collimator for attaching a telescope

2.1.1.3. The two output ports shall be configured to accept IR detectors

2.1.2. The FTIR System shall incorporate a Stirling-cooled photovoltaic mercury-cadmium-telluride (MCT) IR detector having the following features:

2.1.2.1. 4 to 13.5 µm wavelength response

2.1.2.2. 16-bit analog-digital converter (ADC)

2.1.2.3. Computer-controlled gain selections of 1, 2, 4, 8, 16, 32, 64, 128, 256

2.1.3. The FTIR System shall incorporate a Stirling-cooled indium-antimonide (InSb) IR detector having the following features:

2.1.3.1. 1 to 5.5 µm wavelength response

2.1.3.2. 16-bit ADC.

2.1.3.3. Computer-controlled gain selections of 1, 2, 4, 8, 16, 32, 64

2.1.4. The FTIR System shall include the following optical components:

2.1.4.1. Input collimator

2.1.4.2. Motor-operated aperture wheel

2.1.4.3. 4° field-of-view (FOV) telescope with manual focus and 2 m minimum focusing distance

2.1.4.4. Ocular eye-piece interchangeable with, or supplemented by, a charge-coupled detector (CCD) color camera featuring NTSC video output for collimator boresighting

2.1.5. The FTIR System shall include software for operating the spectroradiometer from a remote computer with the following capabilities:

2.1.5.1. Obtain measurements from both IR detectors simultaneously

2.1.5.2. Continuously record and save up to 30 minutes of spectroradiometer data to disk in real-time

2.1.5.3. Trigger data recording using an external signal provided to the computer from a remote device (via serial, parallel or other electronic interface)

2.1.5.4. Monitor and diagnose the operating status of the spectroradiometer

2.1.5.5. Provide control of the spectroradiometer’s remotely-configurable settings

2.1.5.6. Perform post data acquisition analysis and processing

2.1.5.7. Perform radiometric calibration of the spectroradiometer in terms of

Radiance, Irradiance and Intensity

2.1.5.8. Software shall include a perpetual license with an offline installation

2.1.6. The FTIR System shall include a temperature control system internal to the spectroradiometer.

2.1.7. The FTIR System shall include a room-temperature cancellation reference source with the spectroradiometer.

2.1.8. The FTIR System shall integrate the electronic controls with the spectroradiometer assembly.

2.1.9. The FTIR System shall allow the spectroradiometer to be mounted and secured to a tripod platform using threaded fasteners.

2.1.10. The FTIR System shall include protective carrying cases to store and transport the spectroradiometer equipment.

2.2. Measurement Performance

2.2.1. The FTIR System shall have a root-mean-square (RMS) noise-equivalent spectral radiance (NESR) at 16 cm-1 resolution in the following wavelength bands:

2.2.1.1. NESR (RMS) below 2.5 X 10-9 W/cm2/sr/cm-1 at peak response between

4 and 13.5 µm.

2.2.1.2. NESR (RMS) below 2.5 X 10-10 W/cm2/sr/cm-1 at peak response between

1 and 5.5 µm.

2.2.2. The FTIR System shall have an optical throughput of 8.1 X10-3 cm2-sr.

2.2.3. The FTIR System shall exhibit no more than 7.5% non-uniformity across 85% of the measured FOV.

2.2.4. The FTIR System shall provide 6 computer-selectable unapodized wavenumber resolutions of 1, 2, 4, 8, 16 and 32 cm-1 at all wavelengths.

2.2.5. The FTIR System shall have the following scan rates:

2.2.5.1. 10 scans/sec at 1 cm-1 resolution.

2.2.5.2. 17 scans/sec at 2 cm-1 resolution.

2.2.5.3. 34 scans/sec at 4 cm-1 resolution.

2.2.5.4. 54 scans/sec at 8 cm-1 resolution.

2.2.5.5. 82 scans/sec at 16 cm-1 resolution.

2.2.5.6. 107 scans/sec at 32 cm-1 resolution.

2.2.6. The FTIR System shall be spectrally stable to 0.01 cm-1.

2.3. Electrical, Physical

2.3.1. The FTIR System equipment shall accept 115VAC/60 Hz, single-phase maximum electrical power (15 amps, maximum).

2.3.2. The FTIR System equipment shall adhere to the following weight and size restrictions:

2.3.2.1. Spectroradiometer weight: 35 kg (excluding power supply, collimator, telescope attachments)

2.3.2.2. Spectroradiometer size envelope: 40 cm X 40 cm X 40 cm (excluding power supply, collimator, telescope attachments)

2.3.3. The FTIR System equipment shall remain fully-functional and mechanically intact when subjected to the following environmental conditions:

2.3.3.1. Operating temperature: 0 to +45°C

2.3.3.2. Storage temperature: -30 to +55°C

2.3.3.3. Relative humidity: less than 90% (non-condensing atmosphere)

2.3.3.4. Operating altitude: 0 to 2 km

2.3.4. The FTIR System equipment shall operate in any orientation with respect to the ground plane under a 1 g static load.

2.3.5. The FTIR System shall remain fully-functional and mechanically intact following vibration load testing according to MIL-STD 810F, method 514.5:

2.3.5.1. Acceleration spectral density of 0.015 g2/Hz from 5 to 40 Hz, decreasing monotonically to 0.0015 g2/Hz at 500 Hz (along typical mounting direction)

2.3.5.2. Acceleration magnitude of 1 g RMS (along typical mounting direction),

0.63 g RMS (along remaining 2 transverse directions)

2.3.6. The FTIR System shall remain fully-functional and mechanically intact following shock load testing according to MIL-STD 810F, method 516.5:

2.3.6.1. 6 g half-sine acceleration amplitude over 10 millisecond period

2.3.6.2. 5 shock loads per direction (15 shock loads, total)

3. Period of Performance Delivery is expected with 26 weeks ARO

4. Deliver to Location NASA Glenn Research Center 21000 Brookpark Rd

MS100-5

Cleveland, OH 44135