Sources Sought Technical Specifications.pdf
PDF 577 KB Posted
- Attached to
- NPDES Lab (Oils/Grease/Chemicals/Water) Testing Federal contract opportunity
- Solicitation number
- W912EF21RSS14
View the file
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
1. Laboratory Certification
The laboratory shall have a current laboratory accreditation including either the State of Oregon Environmental Laboratory Accreditation Program or the Washington State Department of Ecology Environmental Laboratory Accreditation Program for the analytes required in the task. If the laboratory cannot conduct a specific analytical process they may contract with another laboratory to perform the analysis. That laboratory must meet this requirement.
2. Task
The laboratory will conduct the following analysis for each outfall at McNary:
Parameter Method/Range Sensitivity
(MDL)
PQL Precision Accuracy Calibration Method
Oil and Grease
EPA 1664A 0.5 1 18% 79-114% Balance calibration
Parameter Container Type Volume Required Preservation Holding Time
Oil and grease AG 1liter None, Cool to ≤4 ̊C 28 days
AG= amber glass
Parameter Approximate Quantity
Cost Total
Oil and grease 2 Weekly 24 Quarterly
Total: 24
The laboratory will conduct the following analysis for each outfall at Ice Harbor:
PQL Precision Accuracy Calibration Method
TSS EPA 160.2 NA 1 NA NA Balance calibration
Oil and
TSS HDPE 1liter None, Cool to ≤4 ̊C 7 days
HDPE = high density polyethylene, AG= amber glass
Cost Total
TSS 1 Quarterly 1 Quarterly
Oil and grease 21 Weekly 252 Quarterly
Total: 253
The laboratory will conduct the following analysis for each outfall at Lower Monumental:
PQL Precision Accuracy Calibration Method
TSS EPA 160.2 NA 1 NA NA Balance calibration
Oil and
TSS HDPE 1liter None, Cool to ≤4 ̊C 7 days
Cost Total
TSS 1 Quarter 1 Quarterly
Oil and grease 12 Weekly 144 Quarterly
Total: 145
The laboratory will conduct the following analysis for each outfall at Little Goose:
PQL Precision Accuracy Calibration Method
COD EPA 410.4 NA 5 20% 80-120% Laboratory blank
TSS EPA 160.2 NA 1 NA NA Balance calibration
Oil and
COD HDPE 500 mL H2SO4, Cool to ≤4 ̊C
28 days
TSS HDPE 1liter None, Cool to ≤4 ̊C 7 days
Cost Total
COD 2 Quarter 2 Quarterly
TSS 1 Quarter 1 Quarterly
Oil and grease 15 Weekly 180 Quarterly
Total: 183
The laboratory will conduct the following analysis for each outfall at Lower Granite:
PQL Precision Accuracy Calibration Method
Oil and
AG= amber glass, Cost Total
Oil and grease 12 Weekly 144 Quarterly
Total: 144
3. Supplies
The analytical laboratory will provide enough sample containers to conduct the approximate quantity of samples detailed in the table above. Enough coolers and freezer packs shall be prepared for pickup to sufficiently pack all sets of samples along with chain of custody forms. There will be enough room to add additional ice packs in the coolers to maintain temperature storage requirements. The laboratory will provide specific packaging and handling instructions for the containers and each sample container will labeled for the test it will be used for. Any preservatives will be described. Coolers will be labeled for proper DOT shipment.
File details come from the government source that posted it. Updated .