This Statement of Objectives (SOO) details technical requirements for the United States Air Force (USAF) Calibration and Instrumentation Laboratory (CIL) to procure NIST-traceable radioactive reference standards. The document specifies four distinct product codes with precise technical specifications:
- Product Code 8401-GF-RASA (148.00 kBq): A composite of six 5in x 20in RASA filters containing multiple radioactive isotopes, including Cd-109, Co-57, and Am-241, deposited on a 0.5mg/cm2 Mylar insert. 2. Product Code 8401-GF-FP (148.00 kBq): A 16in diameter U1 filter with similar isotopes, folded into a 1/16 geometry. 3. Product Codes 8405-EG-SD with two variants (185.00 kBq and 37.00 kBq): 10mL polyester matrix standards in Wheaton HDPE containers with additional isotopes like Mn-54 and Zn-65. All standards require +20%/-20% activity tolerance, NIST traceability, and certified documentation. The delivery deadline is not later than September 30, 2025, and remanufactured or refurbished products will not be accepted.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 25QB169 JA Radioactive Standards.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Statement of Objective (SOO)/Technical Requirements Package (TRP)
USAF CIL NIST-Traceable radioactive standards
17 July 2025
1. SOO/TRP for Equipment
USAF CIL NIST-Traceable radioactive standards
2. Purpose
The purpose of this effort is to procure certified NIST-traceable radioactive reference standards prepared in specific geometries/matrices supplied by Radiochemistry Laboratory's (CIL). Specific geometries are detailed in section 3.
3. Technical Requirements
3.1 List of required standards and geometries/matrices
1. Product Code: 8401-GF-RASA – 1.00 Each
148.00 kBq ( 4.00 uCi ) Activity Tolerance: +20% / -20% Cd-109, Co-57, Ce-139, Hg-203, Sn-113, Cs-137, Y-88, Co-60, Am-241 Composite of six 5in x 20in RASA filters in heat sealed plastic bag, dimensions of active area 4.25in x 16in, active material to be deposited on 5in x 20in 0.5mg/cm2 Mylar insert in between the 3rd and 4th RASA filters (filters to be inserted back into original white plastic covering prior to being deflated and heat sealed in a 4mil poly sleeve) Calibrated, NIST Traceable
2. Product Code: 8401-GF-FP – 1.00 Each
148.00 kBq ( 4.00 uCi ) Activity Tolerance: +20% / -20% Cd-109, Co-57, Ce-139, Hg-203, Sn-113, Cs-137, Y-88, Co-60, Am-241 16in dia U1 filter in poly bag, 15.5in AD, 0.5mg/cm2 Mylar cover, folded into 1/16 fold geometry (folded filter to be pressed flat, bag to be deflated prior to heat sealing in a 4mil poly sleeve) Calibrated, NIST Traceable
3. PRODUCT CODE 8405-EG-SD – 1.00 Each
185.00 kBq ( 5.00 uCi ) Activity Tolerance: +20% / -20% Cd-109, Co-57, Ce-139, Hg-203, Sn-113, Cs-137, Y-88, Co-60, Am-241, Mn-54, Zn-65 1.17g/cc +/-3% polyester matrix, NIST traceable Fill Volume: 10mL Wheaton HDPE LSV w/ silicone adhesive to prevent solid fill from dislodging Calibrated, NIST Traceable
4. PRODUCT CODE: 8405-EG-SD – 1.00 Each
37.00 kBq ( 1.00 uCi ) Activity Tolerance: +20% / -20% Cd-109, Co-57, Ce-139, Hg-203, Sn-113, Cs-137, Y-88, Co-60, Am-241, Mn-54, Zn-65 1.17g/cc +/-3% polyester matrix, NIST traceable Fill Volume: 10mL Wheaton HDPE LSV w/ silicone adhesive to prevent solid fill from dislodging Calibrated, NIST Traceable
Warranty and Support characteristics
• Nuclear standards shall be accompanied by certified NIST-traceable documentation.
4. Technical Contractual Requirements
4.1 Remanufactured/Refurbished Products
Remanufactured/Refurbished products shall not be accepted.
5. Delivery NLT 30 Sep 2025